DEPARTMENT OF TECHNICAL EDUCATION

ANDHRA PRADESH
Name : A. Vinay Kumar
Designation : HEIE
Branch : AEIE
Institution : Q.Q GPT, Hyderabad.
Subject : Process Instrumentation – II
Semester : IV
Code No : AEI-405
Major Topic : Pressure measurement
Duration : 50 Mins.
Sub Topic : Pressure Calibration
Teaching Aids : Photos, animations, diagrams
AEI405.20 1
Recap

In last class we have learnt about the


• Pressure multiplexer

AEI405.20 2
Objectives

Upon completion of this topic you will be able to


know about

Pressure calibration
• Description
• Principle
• Operation

AEI405.20 3
Principle
It follows the principle
F=p A
Where,
F = Equivalent force of the piston weight
combination in N.
(Total mass x Acceleration due to gravity)
p = Pressure
A = equivalent area of the piston-cylinder
combination

AEI405.20 4
Principle…

• By changing the weights appropriately the sensor can be


calibrated.
• Its range is 50 to 50 x 106 pa

AEI405.20 5
Description

• Hydraulic or pneumatic operated dead weight tester


using piston gauge is the basic standard calibration
device
• It consists of two accurately machined cylinders
• Honed to micron tolerance
• Inserted into two closed and known cross sectional area

• They are coupled together to a reservoir as shown in


fig.1.

AEI405.20 6
Deadweight tester for pressure calibration

Calibrated Standard Weights

Sensor under
calibration

Piston
( Precision
bore) Screwed plunger

Fluid
Fig.1

AEI405.20 7
Diagram

AEI405.20 8
Description…

• One of the cylinders is fitted with a close fitting precision


piston

• It will have a top platform

• Where accurately known weights can be loaded

• Another cylinder is connected with the transducer under


test

AEI405.20 9
Operation
• The fluid pressure is then gradually applied

• Until the force is large enough to just lift the precision


weight combination

• Piston floats freely

• Piston gauge with arranged weights are in equilibrium


with the pressure developed in the cylinder

AEI405.20 10
Summary

We have discussed about


Pressure calibration
• Description

• Operation

• Range

AEI405.20 11
Quiz

1. The range of pressure calibration using dead


weight tester is

• 50-60 Pa
• 60 -70 K Pa
• 50 - 50 x 106 Pa
• None

AEI405.20 12
Frequently Asked Questions

1) Explain the operation of Pressure calibration


using dead weight tester?

AEI405.20 13