Eleganza Smart Service manual LINET spol. s r.o. elevice 5 274 01 Slan, esk Republika Service Department: E-mail: servis@linet.cz Tel.: +420 312 576 258 Fax: +420 312 522 668 Author: LINET, s.r.o Version: 01 Date of printing: 03/2006 File name: servismanual_Eleganza_smart.doc Copyright LINET, s.r.o., 2005 Translation LINET, 2005 All rights reserved. All trademarks and brands are the property of the appropriate owners. The manufacturer reserves the right to changes in the contents of this manual that relate to the products technical regulations. It is for this reason that the contents of this manual may indicate differences from the current manufacture of the product.
Contents
1
1.1
2
2.1 2.2
Page 3 from 11
Picture Index:
Picture nr. 1: Battery charging indication PB30.XX .............................................................................5 Picture nr. 2 :ACP panel used with power box PB 29.XX .......................................................................5 Picture nr. 3: Power box PB 30.XX ........................................................................................................6 Picture nr. 4: Power box PB 29.XX .........................................................................................................6 Picture nr. 5: Function lock LEDs ...........................................................................................................7 Picture nr. 6: Active GO button ...............................................................................................................7 Picture nr. 7: Connection of motors and controls to power box PB 30. ..................................................9 Picture nr. 8: Connection of motors and controls to power box PB 29 ..................................................9 Picture nr. 9: CPR setting .....................................................................................................................10
Page 4 from 11
1 Fault diagnosis
1.1
The battery LED on the ACP panel illuminates as the bed is being connected to the power supply to indicate that mains power is present. The battery LED flashes during charging of the battery. If the battery LED does not illuminate, the battery is fully-charged. This only applies to power box PB 30.XX. Picture nr. 1: Battery charging indication PB30.XX
The ACP panel used with power box PB 29.XX does not include the battery-charging LED or Trendelenburg function.
Page 5 from 11
Only when a working ACP panel is connected to the power box are we able to use all electric functions and controls. If the ACP panel is not connected to the power box, the power box beeps intermittently.
Page 6 from 11
What should be done if the bed does not function at all? It is possible, using the function locks, to lock individual functions temporarily. If the function lock LEDs are illuminated, it is not possible to position the locked part of the bed. Picture nr. 5: Function lock LEDs
It is always necessary, before using any of the beds functions, to activate the keyboard by pushing the GO button.
Page 7 from 11
YES
Check the main power supply connection Activate the ACP panel by pressing the GO button
NO
Connect the bed to the main power supply Check whether the main power cable is damaged. If it is ok, check the fuses in the power box,which is located under the middle section of the patient surface. The bed does not work after connecting the new ACP panel or it is not possible to position the bed or any of its parts using another beds controls.
If it is not possible to position the bed using the ACP panel or any other controls. (It is necessary to push the GO button). Connect a new/working ACP panel. After connecting a working ACP panel the fault disappears. It is possible to position the bed from all controls. The ACP was faulty.
The bed still does not function using the ACP panel. It is necessary to change the power box.
If it is possible to position the bed using the ACP panel, there is a faulty control. By reconnecting the individual controls, find which control is faulty and replace.
Page 8 from 11
P5 P6 P7 P1 P2 P3 P4 7 6 5 4 3 2 1
P5 P6
Lko Eleganza Smart me a PB 29. 6bt vybaveno 5 4 zdrojem 3 PB 2 30 (na 1 obrzku) P1 P2 zdroj PB 30.00 nem monost zapojen dalch ovlada oznaen P1-P8
P7 P3 P4
7 6 5 4 3 2 1
Battery (only power box PB 30.XX) Leg-rest motor Backrest motor Head-end column Foot-end column Hand control Supervisor panel ACP
PB 30.02, PB29.02 power box control outlets (Optional controls and undercarriage light) P1 Siderail control P2 Siderail control P3 Foot switch P4 Foot switch
Page 9 from 11
2.1
YES
One of the motors does not position from all controls (foot switch, siderail, ACP). Check the function locks on the ACP panel are not activated. One of the motors still does not work. Eliminate the fault by changing the connection between the motors in the power box. The columns are connectors 3 and 4. The leg-rest and backrest motors are 5 and 6 After swapping the connectors, the motor which didnt work before is still not working. Replace the faulty motor and connect the motors to the power box in the original way.
NO
One of the motors does not function from only one control replace faulty control.
After swapping the connectors, the motor which worked before stops working. Replace the power box.
Attention! If the backrest motor is working but the backrest is not rising, the CPR wire may be too tight. It is necessary to loosen it using the setting nut. All service operations (motors, controls, spare parts replacement), can be found on the service video CD, which is part of this manual.
Page 10 from 11
2.2