Professional Documents
Culture Documents
Millivoltmeters
(Anode Voltage Distribution Measurement Voltmeter)
Revision No. : 00
Effective Date :
TABLE OF CONTENTS
PAGE
1.0 SCOPE
4.0 PREPARATION
5.0 PROCEDURE
8.0 APPENDICES
DATE:
Calibration Procedure
Millivoltmeters
(Anode Voltage Distribution Measurement Voltmeter)
Revision No. : 00
Effective Date :
1.0 - SCOPE
This Procedure describes how to calibrate Millivoltmeters used in ANODE VOLTAGE DISTRIBUTION
MEASUREMENT.
Note:
The millivoltmeter being calibrated is referred to herein as the UUT (Unit Under Test) and STD is
referred to Master or Reference Standard Instrument.
Specifications is carried out in accordance with Manufacturer Manual.
MINIMUM-USE-SPECIFICATION
MANUFACTURER/ MODEL
DC Voltage Source
25ppm or better
Fluke 5500A
N/A
N/A
ACCESSORIES
4.0 PREPARATION
4.1 - Visually inspect the UUT for obvious damage that may affect accuracy and/or operation and
note any observations. Clean the UUT as necessary to avoid cross contamination to yourself or
other instruments.
4.2 - If the UUT is a New Instrument, assign a unique Calibration Control # to the instrument.
4.3 - If the UUT already has an assigned Calibration Control #, inspect the label for integrity. Replace
damaged control labels as needed.
4.4 - Allow STD and UUT to stabilize at (18-22) C for no less than 1 hour.
Calibration Procedure
Millivoltmeters
(Anode Voltage Distribution Measurement Voltmeter)
Revision No. : 00
Effective Date :
5.0 PROCEDURE
5.1 Connect a low thermal test leads across the positive and negative input terminals of the STD,
then connect the other end of the thermal test leads to UUT as shown in Figure 1 below.
Figure 1 Connection
5.2 Set the STD to the voltage level specified in Table 1 below.
Function/Range
Nominal
Tolerance
0 to 199.9mV
0.0mV
0.1% + 1digit
0 to 199.9mV
40.0mV
0.1% + 1digit
0 to 199.9mV
80.0mV
0.1% + 1digit
0 to 199.9mV
120.0mV
0.1% + 1digit
0 to 199.9mV
160.0mV
0.1% + 1digit
0 to 199.9mV
199.5mV
0.1% + 1digit
5.3 Using Table 1, obtains an output from the STD equal to the values in the Nominal column, and
verifies the UUT reads within the limits specified.
5.4 If any step fails, perform the adjustment procedure as per manufacturer recommendation.
5.5 Disconnect the UUT from the STD input terminals. This completes the performance verification
test.
Calibration Procedure
Millivoltmeters
(Anode Voltage Distribution Measurement Voltmeter)
Revision No. : 00
Effective Date :
Revised By:
00
Initial Release
-----
Reviewed and
Approved By:
Revision Date
8.0 APPENDICES
List of Manufacturers Manuals and Guides Used
- Manufacturer specification.
- The Expression of Uncertainty and Confidence in Measurement M3003.
DC Voltage Calibration