804 Ti Stand

Manual

8.804.8001EN

06.2010 ek

Teachware
Metrohm AG
CH-9101 Herisau
teachware@metrohm.com

This documentation is protected by copyright. All rights reserved.


Although all the information given in this documentation has been
checked with great care, errors cannot be entirely excluded. Should you
notice any mistakes please send us your comments using the address
given above.
Documentation in additional languages can be found on
http://products.metrohm.com under Literature/Technical documentation.

Table of contents

Table of contents
1 Introduction
1.1

Instrument description ......................................................... 1

1.2
About the documentation ................................................... 1
1.2.1
Symbols and conventions ........................................................ 1
1.3
Safety instructions ................................................................ 2
1.3.1
General notes on safety ........................................................... 2
1.3.2
Electrical safety ........................................................................ 2
1.3.3
Working with liquids ................................................................ 3
1.3.4
Flammable solvents and chemicals ........................................... 3
1.3.5
Recycling and disposal ............................................................. 4

2 Device overview

3 Installation

3.1
Setting up the instrument .................................................... 6
3.1.1
Packaging ................................................................................ 6
3.1.2
Checks .................................................................................... 6
3.1.3
Location .................................................................................. 6
3.2

Mounting the 804 Ti Stand ................................................. 7

3.3

Connecting the 804 Ti Stand ............................................. 10

4 Technical specifications

12

4.1

Rotational speed 804 Ti Stand .......................................... 12

4.2

Power supply ...................................................................... 12

4.3

Safety specifications ........................................................... 12

4.4

Electromagnetic compatibility (EMC) ................................ 12

4.5

Ambient temperature ......................................................... 13

4.6

Dimensions .......................................................................... 13

5 Conformity and warranty

14

5.1

Declaration of Conformity ................................................. 14

5.2

Quality Management Principles ........................................ 15

5.3

Warranty (guarantee) ......................................................... 16

6 Accessories

17

6.1
Scope of delivery ................................................................ 17
6.1.1
804 Ti Stand without stand 2.804.0010 ................................. 17
6.1.2
804 Ti Stand with stand 2.801.0040 ...................................... 17

804 Ti Stand

III

Table of contents

6.2
Optional accessories ........................................................... 19
6.2.1
804 Ti Stand 2.804.0010, 2.804.0040 ................................... 19

IV

804 Ti Stand

Table of figures

Table of figures
Figure 1
Figure 2
Figure 3
Figure 4

804 Ti Stand

Front 804 Ti Stand ............................................................................. 5


Rear 804 Ti Stand .............................................................................. 5
Connecting the 804 Ti Stand ........................................................... 11
Connecting the Dosino to the 804 Ti Stand ..................................... 11

1 Introduction

1 Introduction
1.1

Instrument description
The 804 Ti Stand is a titration stand and controller for the propeller stirrer
802 Stirrer.
The titration stand is directly connected to an analysis device via an MSB
connection cable.

1.2

About the documentation


Caution
Please read through this documentation carefully before putting the
instrument into operation. The documentation contains information
and warnings which the user must follow in order to ensure safe operation of the instrument.

1.2.1

Symbols and conventions


The following symbols and styles are used in this documentation:
Cross-reference to figure legend
The first number refers to the figure number, the
second to the instrument part in the figure.
Instruction step
Carry out these steps in the sequence shown.
Warning
This symbol draws attention to a possible life hazard
or risk of injury.
Warning
This symbol draws attention to a possible hazard due
to electrical current.
Warning
This symbol draws attention to a possible hazard due
to heat or hot instrument parts.

804 Ti Stand

1.3 Safety instructions

Warning
This symbol draws attention to a possible biological
hazard.
Caution
This symbol draws attention to a possible damage of
instruments or instrument parts.
Note
This symbol marks additional information and tips.

1.3

Safety instructions

1.3.1

General notes on safety


Warning
This instrument may only be operated in accordance with the specifications in this documentation.
This instrument has left the factory in a flawless state in terms of technical
safety. To maintain this state and ensure non-hazardous operation of the
instrument, the following instructions must be observed carefully.

1.3.2

Electrical safety
The electrical safety when working with the instrument is ensured as part
of the international standard IEC 61010.
Warning
Only personnel qualified by Metrohm are authorized to carry out service
work on electronic components.
Warning
Never open the housing of the instrument. The instrument could be
damaged by this. There is also a risk of serious injury if live components
are touched.
There are no parts inside the housing which can be serviced or replaced
by the user.

804 Ti Stand

1 Introduction

Mains voltage
Warning
An incorrect mains voltage can damage the instrument.
Only operate this instrument with a mains voltage specified for it (see
rear panel of the instrument).
Protection against electrostatic charges
Warning
Electronic components are sensitive to electrostatic charges and can be
destroyed by discharges.
Always pull the mains cable out of the mains connection socket before
connecting or disconnecting electrical appliances on the rear panel of
the instrument.

1.3.3

Working with liquids


Caution
Periodically check all system connections for leaks. Observe the relevant
regulations in respect to working with flammable and/or toxic fluids
and their disposal.

1.3.4

Flammable solvents and chemicals


Warning
All relevant safety measures are to be observed when working with
flammable solvents and chemicals.

804 Ti Stand

Set up the instrument in a well-ventilated location (e.g. laboratory


flue).
Keep all sources of flame far from the workplace.
Clean up spilled fluids and solids immediately.
Follow the safety instructions of the chemical manufacturer.

1.3 Safety instructions

1.3.5

Recycling and disposal


This product is covered by European Directive 2002/96/EC, WEEE Waste
from Electrical and Electronic Equipment.
The correct disposal of your old equipment will help to prevent negative
effects on the environment and public health.
More details about the disposal of your old equipment can be obtained
from your local authorities, from waste disposal companies or from your
local dealer.

804 Ti Stand

2 Device overview

2 Device overview

Figure 1
1

Front 804 Ti Stand

LED function display


Lights up, when the stirrer is switched on.

MSB

2
3.804.300x

Figure 2
1

Rear 804 Ti Stand

MSB connector
Metrohm Serial Bus.

Clamping screw
For fixing the positioning height or the
swing position of the magnetic stirrer.

Propeller stirrer connector


For connecting the 802 Stirrer

For connecting additional devices


3

MSB connection cable


For connecting to a control device.

804 Ti Stand

3.1 Setting up the instrument

3 Installation
3.1

Setting up the instrument

3.1.1

Packaging
The instrument is supplied in highly protective special packaging together
with the separately packed accessories. Keep this packaging, as only this
ensures safe transportation of the instrument.

3.1.2

Checks
Immediately after receipt, check whether the shipment has arrived complete and without damage by comparing it with the delivery note.

3.1.3

Location
The instrument has been developed for operation indoors and may not be
used in explosive environments.
Place the instrument in a location of the laboratory which is suitable for
operation, free of vibrations, protected from corrosive atmosphere, and
contamination by chemicals.
The instrument should be protected against excessive temperature fluctuations and direct sunlight.

804 Ti Stand

3.2

3 Installation

Mounting the 804 Ti Stand


Mounting the stand

6.2001.060

6.2016.070

2
Mount the stand as follows:
1 Screw the 6.2001.060 stand plate with the accompanying four
screws tightly to the bottom of the instrument.
2 Screw the 6.2016.070 stand plate with the hexagon screw (hexagon
key supplied) firmly to the stand plate.

804 Ti Stand

3.2 Mounting the 804 Ti Stand

Fastening the titration stand

Fasten the titration stand as follows:


1 Attach the titration stand on the support rod and push it downwards.
2 With the clamping screw, fix the positioning height or the swing
position of the titration stand.

804 Ti Stand

3 Installation

Attaching the clamping ring and the electrode holder

6.2021.020
6.2013.010

Fasten the clamping ring and the electrode holder as follows:


1 Push the 6.2013.010 clamping ring onto the support rod.
2 Attach the 6.2021.020 electrode holder on the support rod.
3 Keep the red button pressed down and push the electrode holder
downwards to the required height.
4 Fix the position of the electrode holder with the aid of the clamping
ring.

804 Ti Stand

3.3 Connecting the 804 Ti Stand

Mounting the propeller stirrer

Mounting the propeller stirrer as follows:


1 Insert the propeller stirrer 802 Stirrer without the stirring propeller
from above into the center opening of the electrode holder.
2 Plug the stirring propeller from below to the propeller stirrer.

3.3

Connecting the 804 Ti Stand


Connect the titration stand as follows:
1 Exit the control software.
2 Connect the connection cable of the titration stand to one of the
sockets marked with MSB on the rear of the control device.

10

804 Ti Stand

3 Installation

Figure 3

Connecting the 804 Ti Stand

3 Connecting an additional device (e.g. Dosimat or Dosino) to the MSB


connector of the titration stand.

MSB

3.804.300X

Figure 4

Connecting the Dosino to the 804 Ti Stand

4 Start the control software.

804 Ti Stand

11

4.1 Rotational speed 804 Ti Stand

4 Technical specifications
4.1

Rotational speed 804 Ti Stand

Maximum rotational speed

1700...1900 U/min

Setting of the
rotational speed

15 steps

Increase of rotational speed per


step

115...125 U/min

4.2

Power supply

Voltage

+12 V, -12 V, +5 V

Power consumption

4W

Fuse

Electronic overload protection

4.3

Safety specifications

Design and testing

4.4

EN/IEC 61010-1:2001
UL 61010-1: 2004
CSA-C22.2 No. 61010-1: 2004
Degree of protection IP40
Protection class III

Electromagnetic compatibility (EMC)

Emission

Standards fulfilled:
EN/IEC 61326-1:2006
EN/IEC 61000-6-3: 2006
EN 55011 / CISPR 11: 2007

Immunity

Standards fulfilled:
EN/IEC 61326-1:2006
EN/IEC 61000-6-2: 2005
EN/IEC 61000-4-2: 2001
EN/IEC 61000-4-3: 2006
EN/IEC 61000-4-4: 2004
EN/IEC 61000-4-5: 2001

12

804 Ti Stand

4 Technical specifications

4.5

EN/IEC 61000-4-6: 2001


EN/IEC 61000-4-11: 2004
EN/IEC 61000-4-14: 2004
NAMUR:2004

Ambient temperature

Nominal function
range

+5+45 C
(at a maximum of 85% humidity)

Storage

20+60 C

Transport

40+60 C

4.6

Dimensions

Width

90 mm

Height (without
stand)

70 mm

Height (with
stand)

430 mm

Depth

140 mm

Weight

330 g

804 Ti Stand

13

5.1 Declaration of Conformity

5 Conformity and warranty


5.1

Declaration of Conformity
This is to certify the conformity to the standard specifications for electrical
appliances and accessories, as well as to the standard specifications for
security and to system validation issued by the manufacturing company.

Name of commodity

804 Ti Stand
Titration stand and controller for 802 Stirrer.
This instrument has been built and has undergone final type testing
according to the standards:

Electromagnetic
compatibility

Safety specifications

Emission:

EN/IEC 61326-1: 2006, EN/IEC 61000-6-3: 2006,


EN 55011 / CISPR 11: 2007

Immunity:

EN/IEC 61326-1: 2006, EN/IEC 61000-6-2: 2005,


EN/IEC 61000-4-2: 2001,
EN/IEC 61000-4-3: 2006,
EN/IEC 61000-4-4: 2004,
EN/IEC 61000-4-5: 2001,
EN/IEC 61000-4-6: 2001,
EN/IEC 61000-4-11: 2004,
EN/IEC 61000-4-14: 2004, NAMUR: 2004

EN/IEC 61010-1: 2001, UL 61010-1: 2004,


CSA-C22.2 No. 61010-1: 2004, degree of protection IP40, protection
class III
This instrument meets the requirements of the CE mark as contained in
the EU directives 2006/95/EC (LVD), 2004/108/EC (EMC). It fulfils the following specifications:

Manufacturer

EN 61326-1

Electrical equipment for measurement, control


and laboratory use EMC requirements

EN 61010-1

Safety requirements for electrical equipment for


measurement, control and laboratory use

Metrohm Ltd., CH-9101 Herisau/Switzerland


Metrohm Ltd. is holder of the SQS certificate ISO 9001:2000 Quality management system for development, production and sales of instruments
and accessories for ion analysis.

14

804 Ti Stand

5 Conformity and warranty

Herisau, 6 January, 2009

5.2

D. Strohm

A. Dellenbach

Vice President, Head of R&D

Head of Quality Management

Quality Management Principles


Metrohm Ltd. holds the ISO 9001:2000 Certificate, registration number
10872-02, issued by SQS (Swiss Association for Quality and Management
Systems). Internal and external audits are carried out periodically to assure
that the standards defined by Metrohms QM Manual are maintained.
The steps involved in the design, manufacture and servicing of instruments
are fully documented and the resulting reports are archived for ten years.
The development of software for PCs and instruments is also duly documented and the documents and source codes are archived. Both remain
the possession of Metrohm. A non-disclosure agreement may be asked to
be provided by those requiring access to them.
The implementation of the ISO 9001:2000 quality management system is
described in Metrohms QM Manual, which comprises detailed instructions on the following fields of activity:
Instrument development
The organization of the instrument design, its planning and the intermediate controls are fully documented and traceable. Laboratory testing
accompanies all phases of instrument development.
Software development
Software development occurs in terms of the software life cycle. Tests are
performed to detect programming errors and to assess the programs
functionality in a laboratory environment.
Components
All components used in the Metrohm instruments have to satisfy the quality standards that are defined and implemented for our products. Suppliers of components are audited by Metrohm as the need arises.
Manufacture
The measures put into practice in the production of our instruments guarantee a constant quality standard. Production planning and manufacturing
procedures, maintenance of production means and testing of components, intermediate and finished products are prescribed.

804 Ti Stand

15

5.3 Warranty (guarantee)

Customer support and service


Customer support involves all phases of instrument acquisition and use by
the customer, i.e. consulting to define the adequate equipment for the
analytical problem at hand, delivery of the equipment, user manuals, training, after-sales service and processing of customer complaints. The
Metrohm service organization is equipped to support customers in implementing standards such as GLP, GMP, ISO 900X, in performing Operational Qualification and Performance Verification of the system components or in carrying out the System Validation for the quantitative determination of a substance in a given matrix.

5.3

Warranty (guarantee)
Metrohm guarantees that the deliveries and services it provides are free
from material, design or manufacturing errors. The warranty period is 36
months from the day of delivery; for day and night operation it is 18
months. The warranty remains valid on condition that the service is provided by an authorized Metrohm service organization.
Glass breakage is excluded from the warranty for electrodes and other
glassware. The warranty for the accuracy corresponds to the technical
specifications given in this manual. For components from third parties that
make up a considerable part of our instrument, the manufacturer's warranty provisions apply. Warranty claims cannot be pursued if the Customer
has not complied with the obligations to make payment on time.
During the warranty period Metrohm undertakes, at its own choice, to
either repair at its own premises, free of charge, any instruments that can
be shown to be faulty or to replace them. Transport costs are to the Customer's account.
Faults arising from circumstances that are not the responsibility of
Metrohm, such as improper storage or improper use, etc. are expressly
excluded from the warranty.

16

804 Ti Stand

6 Accessories

6 Accessories
6.1

Scope of delivery
Note
Subject to change without notice.

6.1.1
Qty.
1

804 Ti Stand without stand 2.804.0010


Order no.

Description

1.804.0010

804 Ti Stand

Titration stand with permanently attached cable for MSB (Metrohm


Serial Bus).
1

6.1.2
Qty.
1

8.804.8001ML Manual 804 Ti Stand

804 Ti Stand with stand 2.801.0040


Order no.

Description

1.804.0010

804 Ti Stand

Titration stand with permanently attached cable for MSB (Metrohm


Serial Bus).
1

6.2001.060

Stand plate without support rod

For mounting the 801, 804, 803 stirrers on a Titrando, Titrino plus,
Dosimat plus and on the 856 and 867 measuring modules.

804 Ti Stand

17

6.1 Scope of delivery

Qty.
1

Order no.

Description

6.2013.010

Clamping ring

For support rods with a diameter of 10 mm


Material:
Metal
Width (mm):
20
Height (mm):
16

6.2016.070

Support rod / 400 mm

Material:
Outer diameter (mm):
Length (mm):

6.2021.020

Steel, stainless 18/8


10
400

Electrode holder

Electrode holder for 4 electrodes and 2 buret tips


Material:
Plastic

18

6.2621.070
Length (mm):

Hexagon key 5 mm
80

804 Ti Stand


Qty.

6 Accessories

Order no.

Description

6.2621.130
Length (mm):

Hexagon key 2 mm
50

8.804.8001ML Manual 804 Ti Stand

6.2

Optional accessories

6.2.1

804 Ti Stand 2.804.0010, 2.804.0040

Order no.

Description

6.2001.090

Base plate two-sided

Double stand plate for fixing two 801, 803, 804 stirrers and a Titrando

6.2151.010

Cable MSB plug / socket

Extension cable for MSB connections and for Touch Control - Titrando. Mini DIN
socket - Mini DIN plug
Length (m):
2

804 Ti Stand