SPESIFIKASI TEKNIS

PENGADAAN ALAT LABORATORIUM UTAMA I


BALAI BESAR PENGAWAS OBAT DAN MAKANAN DI MEDAN
TA. 2016

NO

NAMA ALAT

HPLC Auto Sampler with UV-VIS and ELSD


Detector

JUMLAH
1

SPESIFIKASI MINIMUM

unit Consist of :
1 Pump
2 ELSD detector
3 UV-VIS detector
4 Nitrogen generator
5 Column heater
6 Data system
7 Accessories, Consumable
I Separation module
1 Solvent management
- Number of solvent

: One to four

- Flow rate setting range

: Minimum 0,050 - 5,000 ml/min, in 0,001 ml/min increaments

- Flow rate accuracy

: 1% or 10 ml/min, whichever is greater (0,200 to 5,000 ml/min)

- Flow rate precision

: Maximum 0,075% RSD or Maximum 0,02 min SD

- Plunger seal wash

: Programable

2 Sample management

(Autosampler)

- Number of sample vials

: Minimum 100 vials

- Injection volume range

: 0,1 - 100 ml

- Injection accuracy

: 2%

- Injection precision

: < 0,5% RSD

- Injection linearity

: > 0,999 coefficient of deviation (1 to 100 ml)

3 Seal Wash

: programable flow rate

4 Degasser
5 Bottle Tray
6 Needle Wash
II ELSD detector
- Nebulizer High-flow rate

: 300 to 3000 ml/min, (tas flow at 1,80 L/min at 25 psi)

- Nebulizer heater power

: 0% to 100%

- Nebulizer cooler

: on/off

- Drift tube

: ambient +5 to 100 C; 0,1C increments

- Gas supply

: Nitrogen, to be supplied at least 65 psi

- Optics

: Heated optics bench (constant 50C)

- Light source

: Tungsten halogen polychromatic front mounted, pre-aligned, user installable.


Warranty : 2000 hours or 1 year (whichever comes first)

- Detector

: Photomultiplier tube

- Scattering angle

: 60 degrees

III UV-VIS detector

- Dual wavelength
- Wavelength

: 190 - 700 nm

- Bandwidth

: Maximum 5 nm

- Wavelength accuracy

: 1 nm

- Wavelength repeatability

: 0,1 nm

- Light source

: Deuterium (D2) lamp; warranty 2000 hours

IV Nitrogen Generator

V Column heater
- Temperature-control method

: Forced air-circulation

- Temperature setting range

: Column environment controllable from ambient plus 5C up to 65C

VI Data System

(PC-based Data Station)

1 Computer

: Minimum : Intel Core i5-4590, 4GB DDR3, 500GB HDD, VGA, Intel HD
Graphics 4600, NIC, Win7 Pro 64bit - Monitor LED 18,5"

2 Printer

: black and colour

3 Software
VII Accessories + consumable
1 Guard holder universal

: 1 pcs

2 Column C18, 5 mm, 4,6x100 mm

: 2 pcs

3 Column C18, 5 mm, 4,6x150 mm

: 2 pcs

4 Guard Column C18, 5 mm, 4,6x20 mm

: 2 pack @ 2 pcs

5 Sampel vial with cap and preslit PTFE/Silicone septa : 1 pack @ 100 pcs

Include :
1 Instalasi
2 Training
3 IQ, OQ, PQ
4 Garansi

: minimum 1 tahun

5 Manual book

HPLC Auto Sampler with PDA Detector

unit Consist of :
1 Pump
2 PDA detector
3 Column heater
4 Data system
5 Accessories, Consumable

I Separation module
1 Solvent management
- Number of solvent

: One to four

- Flow rate setting range

: Minimum 0,050 - 5,000 ml/min, in 0,001 ml/min increaments

- Flow rate accuracy

: 1% or 10 ml/min, whichever is greater (0,200 to 5,000 ml/min)

- Flow rate precision

: Maximum 0,075% RSD or Maximum 0,02 min SD

- Plunger seal wash


2 Sample management

: Programable
(Autosampler)

- Number of sample vials

: Minimum 100 vials

- Injection volume range

: 0,1 - 100 ml

- Injection accuracy

: 2%

- Injection precision

: < 0,5% RSD

- Injection linearity

: > 0,999 coefficient of deviation (1 to 100 ml)

3 Seal Wash

: programable flow rate

4 Degasser
5 Bottle Tray
6 Needle Wash

II PDA detector
1 Operating specification
- Wavelength range

: 190 - 800 nm

- Wavelength accuracy

: 1 nm

- Wavelength repeatability

: 0,1 nm

- Photodiodes

: 512

2 Optical component specification


- Light source

: Deuterium arc lamp; warranty 2000 hours or 1 year warranty (whichever come
first)

- Pressure limit

: 1000 psi (standard analytical cell)

3 Electrical specification

- Power requirements

: 100 to 240 VAC

- Power consumption

: 195 VA (nominal)

III Column heater


- Temperature-control method

: Forced air-circulation

- Temperature setting range

: Column environment controllable from ambient plus 5C up to 65C

IV Data System

(PC-based Data Station)

1 Computer

: Minimum : Intel Core i5-4590, 4GB DDR3, 500GB HDD, VGA, Intel HD
Graphics 4600, NIC, Win7 Pro 64bit - Monitor LED 18,5"

2 Printer

: black and colour

3 Software

V Accessories + consumable
1 Column C18, 5 mm, 4,6x250 mm

: 2 pcs

2 Sampel vial with cap and preslit PTFE/Silicone septa : 1 pack @ 100 pcs

Include :
1 Instalasi
2 Training
3 IQ, OQ, PQ
4 Garansi
5 Manual book

: minimum 1 tahun

Microwave Digest System

unit

- Microwave cavity

: Rugged, High Grade 316 Solid-Steel

- Power Output

: 1800 watts

- High-Resolution, full color touchscreen with speakers


- 8-Gigabyte onboard controller
- Magnetron Protection
- Reactiguard sensor
- Integrated high capacity exhaust system
- Sensors :
- Vessel Recognition
- Vessel Counting
- Loading capacity

: Minimum 12 vessel per loading

- Vessel size

: 100 ml

- Transparent window
- Completed with USB ports, ethernet ports, RS-232 port
- Onboard help and training videos
- Integrated printer
- Temperature and Pressure Control :
- Single vessel reference pressure sensor
- DuoTemp Temperature Control System

Consist of :

: Combination of floor mounted, contacless, all vessels, Dual IR sensors and


direct fiber optic temperature sensor of the reference vessel

1 3 Set : Vessel set 100 ml, contain

: 11 vessel assemblies, 1 control vessel assembly, 1 turntable assembly and


torque station

2 Control vessel assembly

: 1 buah

3 Temperature control probe

: 1 buah

4 Stabilizer 5 kva

: 1 buah

Include :
1 Instalasi
2 Uji Fungsi
3 Training
4 Garansi

: minimum 1 tahun

5 Manual book

pH/ion meter

unit

- Universal instrument for measurements of pH, mV/ORP and ions


- pH range

: Minimum : -2,000 to 20,000

- pH resolution

: 0,001

- pH relative accuracy

: Maximum : 0,002

- mV-range

: Minimum : -2000,0 to 2000,0

- mV-resolution

: 0,1

- mV-relative accuracy

: 0,2

- Concentration range

: Minimum : 1,00E-9 to 9,99E+9

- Concentration accuracy

: Maximum : 0,5%

Consist of :
- Electrode for Fluoride

: 2 unit

- Reference electrolyte for Fluoride

: 2 unit (Expired Date minimum 1 tahun)

- ISA solid state ISE

: 2 unit (Expired Date minimum 1 tahun)

- Fluoride ISE standard solution 1000 mg/L, 500 ml

: 2 unit (Expired Date minimum 1 tahun)

- Fluoride ISE standard solution 100 mg/L, 500 ml

: 2 unit (Expired Date minimum 1 tahun)

- Fluoride ISE standard solution 10 mg/L, 500 ml

: 2 unit (Expired Date minimum 1 tahun)

Total ionic strength adjustment buffer for Fluoride


determination, 250 ml

: 2 unit (Expired Date minimum 1 tahun)

Include :
1 Instalasi
2 Training
3 IQ, OQ, PQ
4 Garansi

: minimum 1 tahun

5 Manual book

UPS 10 kva

unit

- Capacity

10000 VA / min 6500 W

- Battery (capacity)

min 9 Ah

Include :
1 Instalasi
2 Uji fungsi
3 Manual book

Fume Hood

unit

4 Garansi

: minimum 1 tahun

- External dimension

: W (1550 - 1900) x D (900 - 1000) mm

- Internal dimension

: W (1500 - 1700) mm

- Worktop

: Phenolic Resin, Drip-cup cut out on left

- Sash material

: Tempered Laminated and Framed Sash Glass

- Maximum sash opening

: 740 5 mm

- Sash configuration

: vertical

Consist of
- Exhaust collar

: 1 unit

- Base cabinet

: 1 unit

- Airflow alarm kit


- Water remote controlled fixture plumbed to the top
- Gas remote controlled fixture plumbed to the top

- Single electrical outlet with splash proof cover

: 4 unit

Include :
1 Instalasi
2 Pemasangan Ducting Standard
3 Uji Fungsi
4 Training
5 Garansi

: minimum 1 tahun

6 Manual book

AAS Flame and HVG

unit Consist of :
I Basic
- Analysis methods

: Flame Atomizer

- Optics

: Optical Double Beam

- Wavelength range

: 185 to 900 nm

- Bandwidth

: 2,0 nm (automatic 4 slit selection)

- Detector

: Photomultiplier tube.

II Flame
- Burner head

: Pure Titanium or Corrosion Resistant Material

- Number of lamps

: minimal 4

- Background correction
III Vapor generator

: Minimal BGC-D2 (D2 lamp method)


: Continuous flow system in determination of Hg and other hydride forming
elements such as As, Sn in ppb concentration.

IV Autosampler
- Repeatability

: max 1%RSD

- Sampling Function

: Dilution function, reagent addition function

- Sample capacities

: Minimum 60 samples

V Piping line system (stainless steel) for Gas Distribution Systems, Three Line (for Acetylene, Nitrous oxide and Air)
VI Data System

1 Computer

: Minimum Intel (R) Core i3 (TM) Processor Mini Tower Black, Chipset Intel
H61 Express Ram 4GB DDR3 HDD 500GB Serial ATA/100 (7200RPM), USB
Optical mousse, USB keyboard, DVDRW integrated video, Intel HD Graphics
2000 (1 HDMI & 1 VGA), Ethernet 10/100/1000 Base, COM RS 232, LED
Monitor 18,5".

2 Printer

: black and colour

3 Software
Complete with :
1 High Temperature Burner
2 Oil-less Air Compressor
3 Mist Separator Kit
4 Acetylene gas with cylinder and pressure regulator
5 Nitrous gas with cylinder and pressure regulator
6 Argon gas with cylinder and pressure regulator

7 Canopy / Fume Hood


8 Flexible ducting
9 Blower
10 Hollow Cathode Lamps

: minimum 5 lamps (As, Hg, Pb, Cd, Cu)

Include :
1 Instalasi
2 Training
3 IQ, OQ, PQ
4 Garansi

: minimum 1 tahun

5 Manual book

Medan, 04 April 2016


Balai Besar POM di Medan
Pejabat Pembuat Komitmen,

Jufri Sibarani, S.Si., Apt.


NIP. 19730511 200003 1 001