Professional Documents
Culture Documents
http://www.yumpu.com/zh/browse/user/charliechong
http://issuu.com/charlieccchong
http://independent.academia.edu/CharlieChong1
http://greekhouseoffonts.com/
Signal conditioning
Signal detection
Signal processing
Source location
Advanced signal processing
Acoustic emission test systems
Accessory materials
Factors affecting test equipment
selection
3 Techniques
Equipment calibration and set up for
test
Establishing loading procedures
Precautions against noise
Special test procedures
Data displays
Fion
Zhang
DNcn=Fion
Zhang,
o=Technical,
ou=Academic,
email=fion_zhang
@qq.com, c=CN
2016.07.24
00:27:39 +08'00'
Charlie Chong/ Fion Zhang
Designation: E 1316 05
Standard Terminology for
Nondestructive Testing
count, ring-downsee count, acoustic emission, the preferred term. (ringdown, the word down could be confusing to indicate the counts of subsiding
amplitude after maxima)
couplanta material used at the structure-to-sensor interface to improve the
transmission of acoustic energy across the interface during acoustic emission
monitoring.
cumulative (acoustic emission) amplitude distribution F(V) see distribution,
amplitude, cumulative.
cumulative (acoustic emission) threshold crossing distribution Ft (V)see
distribution, threshold crossing, cumulative.
(1)
where:
A0 = 1 V at the sensor (before amplification), and
A1 = peak voltage of the measured acoustic emission signal (also before
amplification).
Acoustic Emission Reference Scale:
dBAE Value
0
20
40
60
80
100
Voltage at Sensor
1V
10 V
100 V
1 mV
10 mV
100 mV
Felicity ratiothe ratio of the stress at which the Felicity effect occurs to the
previously applied maximum stress. (E 1067, E 1118)
NOTE 5The fixed sensitivity level will usually be the same as was
used for the previous loading or test. (E 1118)
instrumentation dead timesee dead time, instrumentation.
first hit locationa zone location method defined by which a channel
among a group of channels first detects the signal.
floating thresholdany threshold with amplitude established by a time
average measure of the input signal. (E 750)
hitthe detection and measurement of an AE signal on a channel.
interval, arrival time (tij)the time interval between the detected arrivals of
an acoustic emission wave at the ith and jth sensors of a sensor array.
1. Scope
1.1 This document provides guidelines for mounting piezoelectric
acoustic emission (AE) sensors.
1.2 UnitsThe values stated in either SI units or inchpound units are to be
regarded separately as standard. The values stated in each system may not
be exact equivalents; therefore, each system shall be used independently of
the other. Combining values from the two systems may result in
nonconformance with the standard.
1.3 This standard does not purport to address all of the safety concerns, if
any, associated with its use. It is the responsibility of the user of this standard
to establish appropriate safety and health practices and determine the
applicability of regulatory limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
E976 Guide for Determining the Reproducibility of Acoustic Emission Sensor
Response E1316 Terminology for Nondestructive Examinations
3. Terminology
3.1 Definitions of Terms Specific to This Standard:
3.1.1 bonding agenta couplant that physically attaches the sensor to the
structure.
3.1.2 couplanta material used at the structure-to-sensor interface to
improve the transfer of acoustic energy across the interface.
3.1.3 mountingfixturea device that holds the sensor in place on the
structure to be monitored.
3.1.4 sensora detection device that transforms the particle motion produced
by an elastic wave into an electrical signal.
3.1.5 waveguide, acoustica device that couples acoustic energy from a
structure to a remotely mounted sensor. For example, a solid wire or rod,
coupled to a sensor at one end and to the structure at the other.
3.2 Definitions:
3.2.1 For definitions of additional terms relating to acoustic emission, refer to
Terminology E1316. (Read)
5. Mounting Methods
5.1 The purpose of the mounting method is to hold the sensor in a fixed
position on a structure and to ensure that the acoustic coupling between the
sensor and the structure is both adequate and constant. Mounting methods
will generally fall into one of the following categories:
5.1.1 Compression MountsThe compression mount holds the sensor in
intimate contact with the surface of the structure through the use of force.
This force is generally supplied by springs, torqued-screw threads, magnets,
tape, or elastic bands. The use of a couplant is strongly advised with a
compression mount to maximize the transmission of acoustic energy through
the sensor-structure interface.
5.1.2 BondingThe sensor may be attached directly to the structure with a
suitable adhesive. In this method, the adhesive acts as the couplant. The
adhesive must be compatible with the structure, the sensor, the environment,
and the examination procedure.
Note: Use of couplant is not a mounting method.
Charlie Chong/ Fion Zhang
6. Mounting Requirements
6.1 Sensor SelectionThe correct sensors should be chosen to optimally
accomplish the acoustic-emission examination objective. Sensor parameters
to be considered are as follows: size, sensitivity, frequency response,
surface-motion response, and environmental and material compatibility.
When a multichannel acoustic-emission examination is being conducted, a
subset of sensors with characteristics similar to each other should be selected.
See Guide E976 for methods of comparing sensor characteristics.
6.3.2 For sensors that are primarily sensitive to particle motion perpendicular
to their face, the viscosity of the couplant is not an important factor. Most
liquids or greases will work as a couplant if they wet the surfaces of both the
structure and the sensor.
For those few sensors which are sensitive primarily to motion in the plane of
their face, very high-viscosity couplant or a rigid bond is recommended. (lamb
wave?)
6.3.3 The thickness of the couplant may alter the effective sensitivity of the
sensor. The thinnest practical layer of continuous couplant is usually the best.
Care should be taken that there are no entrapped voids in the couplant.
Unevenness, such as a taper from one side of the sensor to the other, can
also reduce sensitivity or produce an unwanted directionality in the sensor
response.
6.4.2 The mounting fixture should provide support for the signal cable to
prevent the cable from stressing the sensor or the electrical connectors. In the
absence of a mounting fixture, some form of cable support should be
provided. Care should be taken to ensure that the cable can neither vibrate
nor be moved easily. False signals may be generated by the cable striking
the structure and by triboelectric effects produced by cable movement.
6.4.3 Where necessary, protection from the environment should be provided
for the sensor or sensor and mounting fixture.
6.4.4 The mounting fixture should not affect the integrity of the structure being
monitored.
6.4.4.1 Permanently installed mounting fixtures must be constructed of a
material compatible with the structure. Possible electrolytic effects or other
forms of corrosion must be considered when designing the mounting fixture.
6.4.4.2 Alterations of the local environment by the mount, such as removal of
the insulation, must be carefully evaluated and corrected if necessary.
6.4.5 The mounting fixture should be designed to have a minimal effect on the
response characteristics of the sensor.
Wave Guide
http://www.mdpi.com/1424-8220/15/5/11805/htm
7. Verification of Response
7.1 After the sensor(s) are mounted on a structure, adequate response
should be verified by injecting acoustic signals into the structure and
examining the detected signal either on an oscilloscope or with the AE
system to be used in the examination. If there is any doubt as to the sensor
response, the sensor should be remounted.
7.1.1 The test signal may be injected by an external source such as the Hsupencil source, or a gas jet (helium or other suitable gas), or by applying an
electrical pulse to another sensor mounted on the structure. For a description
of these methods see Guide E976.
7.2 Periodic VerificationOn an extended acoustic emission examination, it
may be desirable to verify the response of the sensors during the examination.
Verification should be performed whenever circumstances indicate the
possibility of a change in the coupling efficiency.
7.3 Post VerificationAt the end of an acoustic emission examination, it is
good practice to verify that all sensors are still working and that there have
been no dramatic changes in coupling efficiencies.
Charlie Chong/ Fion Zhang
8. Report
8.1 Any report of the mounting practice should include details of the sensor
mounting fixture(s), surface preparation method, and the couplant that was
used.
9. Keywords
9.1 acoustic emission; acoustic emission sensors; Acoustic emission
transducers; AE; bonding agent; couplant; Mounting fixture; Waveguide
>| dsssa
1. Scope
1.1 This practice is recommended for use in testing and measuring operating
characteristics of acoustic emission electronic components or units. (See
Appendix X1 for a description of components and units.) It is not intended that
this practice be used for routine checks of acoustic emission instrumentation,
but rather for periodic evaluation or in the event of a malfunction. The sensor
is not addressed in this document other than suggesting methods for
tandardizing system gains (equalizing them channel to channel) when
sensors are present.
1.2 Where the manufacturer provides testing and measuring details in an
operating and maintenance manual, the manufacturers methods should be
used in conjunction with the methods described in this practice.
1.3 The methods (techniques) used for testing and measuring the
components or units of acoustic emission instrumentation, and the results of
such testing and measuring should be documented. Documentation should
consist of photographs, screenshots, charts or graphs, calculations, and
tabulations where applicable.
Charlie Chong/ Fion Zhang
1.4 AE systems that use computers to control the collection, storage, display,
and data analysis, might include waveform collection as well as a wide
selection of measurement parameters (features) relating to the AE signal. The
manufacturer provides a specification for each system that specifies the
operating range and conditions for the system. All calibration and acceptance
testing of computer-based AE systems must use the manufacturers
specification as a guide. This practice does not cover testing of the computer
or computer peripherals.
1.5 The values stated in SI units are to be regarded as standard. No other
units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns, if
any, associated with its use. It is the responsibility of the user of this standard
to establish appropriate safety and health practices and determine the
applicability of regulatory limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
E1316 Terminology for Nondestructive Examinations
2.2 ANSI Standard:
ANSI/IEEE 100-1984 Dictionary of Electrical and Electronic Terms
2.3 Other Documents:
Manufacturers Operating and Maintenance Manuals pertinent to the specific
instrumentation or component
3. Terminology
3.1 DefinitionsFor definitions of additional terms relating to acoustic
emission, refer to Terminology E1316.
4. Summary of Practice
4.1 Tests and measurements should be performed to determine the
instrumentation bandwidth, frequency response, gain, noise level, threshold
level, dynamic range, signal overload point, dead time, and counter accuracy.
4.2 Where acoustic emission test results depend upon the reproduced
accuracy of the (1) temporal, (2) spatial, or (3) spectral histories, additional
measurements of instrumentation parameters should be performed to
determine the specific limits of instrumentation performance. Examples of
such measurements may include amplifier slew rate, gate window width and
position, and spectral analysis.
4.3 Tests and measurements should be performed to determine the loss in
effective sensor sensitivity resulting from the capacitive loading of the cable
between the preamplifier and the sensor. The cable and preamplifier should
be the same as that used for the acoustic emission tests without substitution.
(See also Appendix Appendix X2.)
http://home.mira.net/~marcop/ciocahalf.htm
6. Apparatus
6.1 The basic test instruments required for measuring the operating
characteristics of acoustic emission instrumentation
include:
6.1.1 Variable Sine Wave Generator or Oscillator,
6.1.2 True RMS Voltmeter,
6.1.3 Oscilloscope,
6.1.4 Variable Attenuator, graduated in decibels, and
6.1.5 Tone Burst Generator.
http://www.parts-express.com/rolls-mo2020-test-oscillator-variable-frequency-sine-square-wave-generator--245-1068
http://www.sm5cbw.se/ake/av8000/av8073.htm
http://www.sm5cbw.se/ake/av8000/av8073.htm
Spectrum Analyzer,
Spectrum Analyzer,
http://www.edelpro.com/generators/generators.html
7. Measurement Procedure
7.1 Frequency Response and Bandwidth Measurements:
7.1.1 The instrumentation, shown in Fig. 1, includes the preamplifier with
amplification and signal filters, possibly connected to the AE system which
might have additional signal filters, amplification, and interconnecting cables.
All measurements and tests should be documented. If the preamplifier is to
be tested without the AE system connected, it should be terminated with the
normal working load as shown on the bottom right of Fig. 1.
7.1.2 An acceptable frequency response should be flat between cutoff
frequencies within 3 dB of the reference frequency. The reference frequency
is the geometric mean of the nominal bandwidth of the instrumentation.
Keypoint:
An acceptable frequency response should be flat between cutoff frequencies
within 3 dB of the reference frequency.
= (fL fH)
where:
fM = mean frequency,
fL = nominal lower cutoff, and
fH = nominal upper cutoff.
7.1.3 The bandwidth should include all contiguous frequencies with amplitude
variations as specified by the manufacturer. Instruments that include signal
processing of amplitude as a function of frequency should have bandwidth
amplitude variations as specified by the manufacturer.
7.1.4 With the instrumentation connected to the oscillator and attenuator, see
Fig. 1 and the sine wave oscillator set well within the instrumentations
specified dynamic range, the frequency response should be measured
between frequency limits specified in 7.1.2. The oscillator is maintained at a
fixed amplitude and the frequency is swept through the frequency limits. The
preamplifier or AE system voltage output is monitored with an RMS voltmeter.
Values of amplitude are recorded for each of several frequencies within and
beyond the nominal cutoff frequencies. The recorded values should be
plotted. The amplitude scale may be converted to decibels. The frequency
scale may be plotted either linearly or logarithmically. Appendix X2 provides
further discussion of wave shaping components.
https://en.wikipedia.org/wiki/Band-pass_filter
White Noise
The White Noise cell generates a signal that contains equal energy per
frequency division (hertz or similar increment).
7.3.2.3 For an amplifier that has a threshold comparator as its output device,
the dynamic range is the ratio of maximum threshold level to input noise level
at the comparator. Excess amplitude range in the amplifier contributes to
overload immunity but not to the dynamic range. The following measurement
will give the effective dynamic range:
DRe = 20 log10(MaxTh/MinTh)
where:
DRe = the effective dynamic range of the system,
MaxTh = the highest settable threshold value that just passes the largest
undistorted peak signal input, and
MinTh = the threshold value that passes less than 1 count/s with no input
signal.
This dynamic range is the difference between the largest and the
smallest AE input that can be reliably detected by the system.
7.3.4 The signal overload level is measured by replacing the sensor with the
sine wave oscillator as shown in Fig. 1. The frequency is set to the mid-band
frequency of the instrumentation. The oscillator amplitude is fixed at 1 V peak
to peak monitored at Vosc. The attenuator is adjusted to increase the signal
level to the preamplifier until the instrumentation output (Vout) is 0.5 dB less
than the computed output.
7.3.5 Should the peak amplitude of acoustic mission activity exceed the
dynamic range, several deleterious effects may be produced; these include
clipping, saturation, and overload recovery time- elated phenomena. (See
Appendix X2 for a discussion of overload recovery.) The instrumentation gain
should be adjusted to limit these effects to an absolute minimum in order to
increase the reliability of the data.
7.4.3 Where the dead time in question is related to AE hit processing such as
measurement of source location, energy, duration, or amplitude, the
measurement is best accomplished by using an electronic AE simulator as
follows:
7.4.3.1 Select an AE hit parameter to evaluate the dead time.
7.4.3.2 Set the electronic AE simulator frequency, rise, decay, duration, and
repetition rate such that the observed AE hit rate in the selected parameter
equals the repetition rate of the simulator.
7.4.3.3 Increase the repetition rate of the simulator until the observed AE hit
rate falls below the simulator rate.
7.4.3.4 Record this value as maximum AE hit rate (processing speed) for the
selected parameter.
Floating or Automatic Threshold cannot be measured with a continuouswave generator because the automatic threshold level is usually derived from
the rectified and averaged input signal.
http://www.nvo.com/aecpipes/nss-folder/literature/AE%20Thresholding.pdf
http://www.nvo.com/aecpipes/nss-folder/literature/AE%20Thresholding.pdf
http://www.nvo.com/aecpipes/nss-folder/literature/AE%20Thresholding.pdf
7.7.3.4 Record the displayed location coordinates for this initial simulated
input. Compute and input a new delta T set for a nearby point. Record the
difference between input position and displayed position. Continue this
incremental movement of the simulated AE source away from the sensor
array center until the output position differs from the input position by 10 % or
the source location range specified by the AE system manufacturer is
exceeded. Evaluate any error with respect to the AE system manufacturers
specification for source location linearity. 7.7.3.5 The source location test
procedure should be repeated for two additional rays extending in different
directions from the array center. 7.7.3.6 The source location procedure should
be repeated for each multichannel array of the system.
8. Keywords
8.1 acoustic emission; AE; dead-time; gain; preamplifier; sensitivity; sensor;
signal processor; threshold
APPENDIXES
(Nonmandatory
Information)
X1. DESCRIPTION OF
AE INSTRUMENT
APPENDIXES
(Nonmandatory Information)
X1. DESCRIPTION OF AE INSTRUMENT
COMPONENTS
X1.1 Acoustic Emission Instrumentation - Acoustic emission electronic
components or units should include the sensor(s), preamplifier(s), filter(s),
power amplifier(s), line drive amplifier(s), threshold and counting
instrumentation, and signal cables. The sensitivity calibration and transfer
characteristics of sensors are excluded from this standard.
X1.2 Acoustic Emission Sensor:
X1.2.1 An acoustic emission sensor is an electro-acoustic transducer that
converts stress wave energy into electrical energy.
X1.2.2 A transformer or amplifier, or acoustic waveguide, if combined with the
sensor in such a way that the readily accessible terminals include these
components should be considered part of the sensor and the term sensor
should apply to the combination.
EMAT Transducer
http://ultrakraft.ru/en/products/testing-of-plate
Keywords:
Inductive sensors may require relatively low impedance loads.
Capacitive sensors generally require high impedance loads.
The low impedance loads depend upon current (or power) drive
The high impedance loads depend upon voltage (or charge) drive.
Because the most commonly used sensor is a piezoelectric device, the
preamplifier input impedance is moderately high.
http://www.soundwel.cn/enproducts/54.html
http://www.soundwel.cn/enproducts/54.html
X1.5 Filter:
X1.5.1 A filter separates signals on the basis of frequency. It introduces
relatively small loss to waves in one or more frequency bands and relatively
large loss to waves of other frequencies.
X1.5.2 Filters may be active or passive. Active filters require electrical power.
Passive filters require no electrical power.
X1.5.3 The most frequently used filter is the bandpass filter. A bandpass filter
is a filter that has a single transmission band extending from a lower cutoff
frequency greater than zero to a finite upper cutoff frequency. The gain at the
cutoff frequencies should be 3 dB less than the passband geometric mean
(reference) frequency as defined in 7.1.2. The slope of the filter characteristic
outside the passband is very important for rejection of extraneous signals.
Slopes of 30 dB per octave are typical for AE instruments.
X1.5.4 The filter should not limit the specified signal overload point of the
preceding component or unit.
X1.5.5 AC-coupled amplifiers and preamplifiers limit the bandwidth by circuit
design. Typical bandwidths may extend from a low of 1 KHz to a high of 2
MHz.
X2.4.4 The calibration voltage is chosen to be any convenient value near that
expected from an acoustic emission event of interest, taking into account the
dynamic range expected from the data.
X2.4.5 The frequency of the calibration voltage should be selected to be well
below the resonances of the AE sensors, which are presumed to be of the
resonant (undamped) type, but within the band pass of the AE system. This
will prevent the individual resonances, which may be different from channel to
channel, from influencing the gain adjustments.
X2.4.6 If the oscillator is calibrated and terminated, a known signal can be
applied and the effective gain of the system with cables and sensor can
bemeasured with an indicator on the channel output such as an RMS
voltmeter.
The tone burst should have a repetition time in excess of the instrumentation
recovery time such that instrumentation recovery should occur in less time
than the next tone burst would occur. The oscilloscope should record the
signal at Vout such that the residual feed-through from the tone burst
generator may be observed following the tone burst and instrumentation
overload recovery. The instrumentation overload recovery time is the time
from the end of the tone burst to the time at which the residual has returned to
its quiescent value (usually 1 % of the tone burst amplitude). A waveform
synthesizer may be substituted for the tone burst generator, but provision
should be made to allow measurement of residual amplitude between bursts.
X4.3.2 Noise may be introduced to the amplifier from noise impulses on the
power mains through inadequately filtered power supplies. The noise is often
intermittent and includes random impulse or burst signals of short duration
and very fast rise time compared with acoustic emission signals. A series of
several low pass (on fast raise time signals?) , power main filters will often
suppress this noise to an acceptable level.
X4.4 Electronic Component NoiseThere are several sources of noise in
electronic circuits. In practice, the noise figure of an amplifier is usually
determined by the first, or input, stage of the amplifier. This is because noise
introduced by other succeeding circuits of the instrumentation will undergo
less amplification, and, thus, will be relatively unimportant in the
instrumentation as long as the amplification of the first stage is moderate.
20160724 0020hrs
http://slideplayer.com/slide/1579129/
Good Luck!
Charlie Chong/ Fion Zhang