You are on page 1of 48

Outline

Ultrasound

I. Basic Ultrasound
II. Ultrasound in Medicine
III. Devices Standard
IV. Ultrasound Instrument
V. Performance and Safety Testing

Bats!
Medical uses of ultrasound

Bats navigate using ultrasound

Bats: Navigating with ultrasound


Bats make high-pitched chirps which are too high for
humans to hear. This is called ultrasound
Like normal sound, ultrasound echoes off objects
The bat hears the echoes and works out what caused
them
Dolphins also navigate with ultrasound
Submarines use a similar method called sonar

We can also use ultrasound to look inside the


body

Bats: Navigating with ultrasound


If a bat hears an echo 0.01 second after it makes a chirp,
how far away is the object?

Ultrasound imaging

Clue 1: the speed of sound in air is 330 ms-1


Clue 2: The speed of sound equals the distance travelled
divided by the time taken
Answer: distance = speed x time
Put the numbers in:
distance = 330 x 0.01 = 3.3 m
But this is the distance from the bat to the object and back
again, so the distance to the object is 1.65 m.

Ultrasound imaging: What does it look


like?

Ultrasound imaging: How does it work?

An ultrasound element acts like a bat.


Emit ultrasound and detect echoes
Map out boundary of object

Ultrasound imaging: How does it work?

Ultrasound imaging: development of a


pregnancy
24 weeks

Now put many elements together to make a probe and create


an image

8 weeks gestation (out of a 40 week pregnancy)


18 weeks

Ultrasound imaging: more surface


rendering

Ultrasound imaging: foetus feet

This is a 2D ultrasound scan


through the foot of a foetus. You can
see some of the bones of the foot.

We can process the image in a


computer to find the outline of the foot.
This is called surface rendering. Here,
the foot has been surface rendered

Ultrasound imaging: imaging the heart

Ultrasound imaging: kissing!

atrium
heart valves

ventricle

Ultrasoundaudible sound (1520,000


Hz) vs. therapeutic ultrasound (800 KHz
3.0 MHz)

Form of inaudible, high frequency acoustic


energy which produces vibrations within
tissues
Produces both thermal and non-thermal
effects
Within medicine proper, utilized for
diagnosis, destruction of tissue, and
treatment
ATs only utilize therapeutic ultrasound

Ultrasound is generated by the


vibration of a crystal in an alternating
electrical currentreferred to as the
reverse piezoelectric effect

Basic Principles of
Ultrasound
Theory and Practice

Examples of oscillation
ball on
a spring

pendulum

rotating
earth

Pulse
The ball starts to oscillate as soon as it is pushed

Movement of the ball over time

Oscillation

Frequency

The actual displacement a is


termed as:

Time
From the duration of one
oscillation T the frequency f
(number of oscillations per
second) is calculated:

One full
oscillation
T

Time

90

180

270

360
Phase

Gelombang
ialah satu gangguan yang membawa
tenaga melalui medium.
Dua ciri umum bagi semua gelombang
iaitu
gelombang adalah satu travelling
disturbance.
gelombang membawa tenaga dari satu
tempat ke satu tempat.

Sinusoidal pressure source

28

"Gelombang melintang

Sound propagation

"Gelombang melintang" (Transverse


wave) ialah gelombang yang
menyebabkan zarah-zarah medium
bergetar berserenjang dengan arah
gerakan gelombang.

Transverse wave
Direction of oscillation

Direction of propagation

29

"Gelombang membujur

Sound propagation

"Gelombang membujur" (Longitudinal


wave) pula menyebabkan getaran zarahzarah medium selari dengan arah gerakan
gelombang.

Longitudinal wave Direction of propagation


Direction of
oscillation

31

What is Ultrasound

Wave propagation
Longitudinal waves propagate in all kind of materials.
Transverse waves only propagate in solid bodies.
Due to the different type of oscillation, transverse
waves
travel at lower speeds.
Sound velocity mainly depends on the density and Emodulus of the material.
Air
Water
Steel, long
Steel, trans

330 m/s
1480 m/s
5920 m/s
3250 m/s

Ultrasound is sound with a frequency greater


than the upper limit of human hearing, this limit
being approximately 20 kilohertz (20,000 hertz).
It is generated by applying an alternate current
to a piezoelectric crystal. This crystal contracts &
expands at the same frequency at which current
changes polarity. The sound field generated by
this crystal in turn makes the molecules in the
sound filed vibrate and oscillate.
34

What is Ultrasound

Spectrum of sound
Frequency range
Hz

Description

Example

0 - 20

Infrasound

Earth quake

20 - 20.000
> 20.000

Audible
sound
Ultrasound

Speech, music
Bat, Quartz
crystal

The crystal commonly used in US units is synthetic


plumbium zirconium titanate (PZT)
The quality of the crystal is what makes your US
expensive
Crystal quality depends on the following:
Beam nonuniformity ratio: ranges from 2 to 6 the
smaller the better.
Effective Radiating Area: as close to sound head area as
possible
Therapeutic ultrasound has a frequency range of 0.7 and 5.0
MHz

Most clinic will have 1 MHz and 3 MHz sound head.


36

Sound wave physics

Atomic structures

Solid and liquid consists of molecules held together by elastic forces that behave like
rubber bands connecting each molecule to each of its nearest neighbors.
If one molecule is set in vibration, then it will cause its immediate neighbors to
vibrate, and it turn their neighbors, and so on until the vibration has propagated
throughout the entire material. This is a wave.
A sound wave is sound energy that is transmitted from one molecule to the next.
A sound wave cannot travel by itself. It needs a medium for transmission (solid,
liquid, gas)
Energy contained within a soundbeam is decreased as it travels through tissue.
Energy is lost to :
Reflection or scattering of the soundbeam when it strikes a
reflecting surface
Absorption energy lost by the sound wave as it overcomes internal
friction that exists in tissue while traveling through it.
Higher the frequency, the more rapidly the molecules are forced to move against this
friction. The more they move, the more energy is consumed (absorbed); the
soundbeam will therefore have less sound energy available to propagated further
through the tissue.

liquid

gas

low density
weak bonding
forces

solid

medium density high density


medium bonding strong bonding
forces
forces
crystallographic
structure

37

Understanding wave propagation:

T
Ball =
atom

Spring = elastic bonding


force

distance travelled

10

Sound wave physics


During one oscillation T the wave
front propagates by the distance
:

The velocity of the wave travel depends on the closeness of the molecules of the
medium. The closer the molecules, the quicker they collide with each other and
sooner they respond to disturbance, the faster they loose energy in a short distance.

So a 3MHz sound head will affect more superficial tissues while a 1 MHz sound head
will affect deeper tissues.

Example :
Sound travels through air easily and can go far (yelling out in the back yard). Sound
can go further because there is little energy loss by absorption. Air molecules are
easily compressed.

Energy finial = Energy initial (Ereflected + Eabsorbed)

Distance travelled

From this we derive:

Wave equation

or

Sound does not travel easily through a brick wall that is denser (someone yells from
the outside of the house and you cant hear him from the inside). Brick molecules are
a lot of closer together, harder to compress. Brick therefore absorbs more of the
sound energy going through it.

42

Frequency
< 20 Hz
Infrasound

20 Hz 20 kHz
Sound

0.001 15 Seismolo 200


Hz
gy
8kHz
0.01 20
Hz

Nature
Sound

Human
Speech

Electromagnetic Wave
> 20 kHz
Ultrasound
20 kHz
60 kHz

Sonar

70 kHz
500 kHz

Sonoche
mestry

500 kHz NDT


3 MHz
1 20
MHz

USG

100 Mhz
3 GHz

Micros
copy

EEG: 0.5 100 Hz

43

1.1. Wave Properties

EMG: 7 20 Hz

44

1.1. Wave Properties

11

Definition

Definition
Ultrasound Velocity (m/s)

V, the velocity is the speed at which sound waves travel


through a particular medium. Velocity is equal to the
frequency x wavelength. The velocity of US through
human soft tissue is 1540 meters per second. Velocity
varies with medium.
Z, acoustic impedance is dependent on the density of
the material in which sound is propagated through. The
greater the impedance the more dense the material.
I, the intensity of the ultrasound beam describes the
amount of energy flowing through a unit cross-sectional
area each second. The intensity is proportional to the
square of the pressure amplitude, particle-displacement
amplitude or particle velocity amplitude.

Acoustic Impedance (kg/m2s = Rayls)

Intensity (joules/cm2 = watts/m2)

45

1.1. Wave Properties

Penyelesaian

pengiraan panjang gelombang

Contoh
Kirakan panjang gelombang bagi isyarat
ultrabunyi 10 MHz di dalam kanta mata

Panjang gelombang di dalam kanta


Halaju gelombang, = frekuensi, f panjang gelombang,
f = 10 MHz isyarat ultrasonik
= 1620 m/s (daripada jadual)

Jadual: Kelajuan ultrabunyi dan acoustic impedance di dalam bahan umum.Data


daripada Wells (1969); Goss, Johnston, Dunn (1978); and Bamber (1986).
Bahan

46

1.1. Wave Properties

Kelajuan (m/s)

Acoustic impedance (g/cm2s)

Aqueous humour

1500

1.50 105

Kanta mata

1620

1.84 105

Kulit

1600

Tulang

3500

7.80 105
47

48

12

Acoustic Properties
Material

v (m/s)

kg/m3)

Z (MRayls)

Chloroform

987

1490

1.47

Glycerin

1904

1260

2.34

Honey

2030

1420

2.89

Water at 20C

1480

1000

1.483

Water at 60C

1550

1000

1.55

Bone

4080

1912

7.8

Blood

1570

1040

1.64

Tissue

1540

1006

1.55

Fat

1450

952

1.38

Muscle

1567

1040

1.63

Air

331

1.2

0.0004

PZT

3791

7650

29

Interaction of Ultrasound with Tissue

Reflection- pantulan
Refraction- pembiasan
Scattering- selerak
Diffraction- pembelauan
Interference- gangguan
Absorption- serapan

49

1.1. Wave Properties

50

I. Basic Ultrasound

Interaction of Ultrasound with Tissue

Reflection and Transmission

Reflection

As soon as a sound wave comes to a change in material


characteristics ,e.g. the surface of a workpiece, or an
internal inclusion, wave propagation will change too:

51

I. Basic Ultrasound

13

Interaction of Ultrasound with Tissue

Behaviour at an interface
Medium 1

Medium 2

Incoming wave

Transmitted wave

The basis of diagnostic ultrasound imaging is the reflection of the


transmitted ultrasound signal by an interface between media of
different acoustic impedances.
If the interface separates two media with the same acoustic
impedance, the ultrasound is not reflected, but continues to
propagate (to be transmitted).
When the interface separates media with different acoustic
impedances, part of the ultrasound is reflected and the remaining is
transmitted.
The amount of reflected ultra-sound energy is proportional to the
difference in acoustic impedance of the two media (the greater the
acoustic mismatch, the greater the amount of sound reflected).
Optimal return (maximal reflection) occurs when the ultra-sound
wave strikes the interface perpendicularly (the angle of incidence is
90).
At any other angle of incidence, less energy is reflected back. The
angle of reflection is equal to the angle of incidence.
The strength of the recorded (reflected) echo is a function of both
the acoustic mismatch at the interface and the angle of incidence.

Reflected wave
Interface

54

I. Basic Ultrasound

Scattering

Echo Type

The scattering occurs


because the interfaces
are smaller than the
diameter of the sound
beam or the interfaces
are less than one
wavelength in size.
The wave is reflected in
all directions. These non
specular reflection have a
strong frequency
dependence.
55

1.2. Ultrasound Interaction

56

1.2. Ultrasound Interaction

14

Echo Type

Absorption
Absorption is the only process whereby sound
energy is dissipated in a medium.
Absorption is the process whereby ultrasound
energy is transformed into other energy form,
primarily heat. (Principle of therapeutic
ultrasound)
The absorption of the ultrasound beam is related
to the frequency of the beam, the viscosity of
medium and the relaxation time of medium.

Specular - echoes originating from relatively


large, regularly shaped objects with smooth
surfaces. These echoes are relatively intense
and angle dependent. (i.e. IVS, valves)
Scattered - echoes originating from relatively
small, weakly reflective, irregularly shaped
objects are less angle dependant and less
intense. (ie. blood cells)

A= ampitude of the beam at a distance x.


Alpha is amplitude absorption coefficient
57

1.2. Ultrasound Interaction

58

1.2. Ultrasound Interaction

Intensity

A-Half Value Layer (HVL)

The intensity of the ultrasound beam is the


amount of energy flowing through a unit crosssectional area each second.
The intensity is the rate at which the energy is
being transmitted by the wave over a small area.
The units are joules/s.cm2 or watt/cm2
The intensity is proportional to the square of
pressure amplitude.
I = Intensity at the point of interest.
Myu= the intensity absorption coefficient.

HVL is the amount of material that reduces the


intensity to one-half its original value (-3dB).
Material
Air
Water
Blood
Bone
Fat
Tissue

1 MHz
0.25
1360
17
0.2
5
4.3

59

1.2. Ultrasound Interaction

2 MHz
0.06
340
8.5
0.1
2.5
2.1

5 MHz
0.01
54
3
0.04
1
0.86

10 MHz
14
2
0.5
0.43
60

1.2. Ultrasound Interaction

15

How can we use US in Medical


Application

Bagaimanakah gelombang ultrasonik boleh

Ultrasound, as the name implies, is sound at a frequency


too high for the human ear to perceive. The instrument
that produces these sound waves is called a
transducer. The doctor can position the transducer on a
portion of the body and send ultrasound waves into the
tissues below.
The sound waves bounce back off internal structures
and return to the transducer where they are collected.
A computer takes these returning sound waves and
creates a picture of the internal structures on a video
screen.

Denyutan bunyi berfrekuensi tinggi dihasilkan oleh


pemancar dan ditujukan ke dalam badan.
Seperti mana di dalam sonar, pantulan akan berlaku.
Ianya akan berlaku setiap masa denyutan bertemu
sempadan di antara dua tisu yang mempunyai
ketumpatan berbeza atau sempadan di antara tisu dan
cecair bersebelahan.
Dengan pengimbasan gelombang ultrasonik
keseluruhan badan dan mengesan gema (echo) yang
dihasilkan daripada pelbagai lokasi dalaman, maka imej
atau sonogram anatomi di dalam badan manusia
diperolehi

digunakan untuk tujuan diagnosis

61

Acoustic impedan

Acoustic impedance of muscle


Acoustic impedan (g/cm2s) = speed(cm/s)
density (g/cm3)

Formula
Acoustic impedan (g/cm2s) = speed(cm/s)
density (g/cm3)
Kelajuan bunyi di dalam

62

Ketumpatan

otot

= 1580 m s-1

otot

= 1.076 103 kg m-3

tulang

= 3500 m s-1

tulang

= 2.23 103 kg m-3

63

64

16

Speed of ultrasound and acoustic impedance in some common Materials. Data from
Wells(1969); Gros, Johnston, Dunn (1978); and Bamber (1986). The acoustic impedance
cannot be calculated where the density of the material is not known.

Acoustic impedance of bone


Acoustic impedan

(g/cm2s)

Material

= speed(cm/s)
density (g/cm3)

65

Nisbah pantulan keamatan (reflected


intensity) kepada incident intensity
Apabila gelombang ultrasound bertemu dengan
sempadan yang memisahkan dua media yang
berlainan di dalam badan manusia,
Nisbah pantulan keamatan (reflected intensity) kepada
incident intensity diberikan seperti di bawah

Acoustic impedance (g/cm2s)

Speed m/s

Amniotic fluid

1510

X 105

Aqueous humour

1500

1.50

X 105

Air (NTP)

330

0.0004

X 105

40

Blood

1570

1.61

X 105

161000

Bone

3500

7.80

X 105

780000

Brain

1540

1.58

X 105

158000

Cartilage

1660

X 105

Castor oil

1500

1.43

X 105

CSF

1510

X 105

Fat

1450

1.38

X 105

138000

Kidney

1560

1.62

X 105

162000

Lens of eye

1620

1.84

X 105

184000

150000

143000

Liver

1550

1.65

X 10

165000

Muscle

1580

1.70

X 105

170000

Perspex

2680

3.20

X 105

320000

Polythene

2000

1.84

X 105

184000

Skin

1600

X 105

Soft tissue (average)

1540

1.63

X 105

Tendon

1750

X 105

Tooth

3600

X 105

163000

Vitreous humour

1520

1.52

X 105

152000

Water (20oC)

1480

1.48

X 105

148000

66

Gelombang ultrasound bertemu dengan


sempadan yang memisahkan dua media
yang berlainan di dalam badan manusia

Sumber
Ultrasound

Pemindaharuh

Otot

Lemak

Antaramuka 1

Otot

Antaramuka 3

Antaramuka 2

67

68

17

Penyelesaian
Medium 1

Medium 2

Medium 3
Acoustic impedanace (g/cm2s)
Otot = 1.70 105
Lemak = 1.38 105
Udara = 0.00004 105
Pada antaramuka 1

100 %

Ultrasound
Probe
Tx & Rx

1
2
3
Interface 1

Interface 3

% reflection = 1.08%

% reflection = 99.88%

Interface 2
% reflection = 1.08%
1

+ 2

Hanya 100 % - 1.08 % = 98.92 % dipancarkan ke dalam lemak dan terkena pada antaramuka 2.
Manakal 1.08 % akan dipantulkan ke arah pemindaharuh penerima.

= Total % of Reflection
69

Pada antaramuka 2

70

Pada antaramuka 3

1.08 % daripada 98.92% iaitu 1.07 % yang terkena antaramuka 2 akan dipantulkan ke antaramuka
1. 1.07 %, ini daripada keamatan asal akan bergerak kembali menerusi antaramuka 1 di mana
1.08% akan dipantulkan kembali semula ke arah antaramuka 2.
98.92 % daripada 1.07 % iaitu 1.06 % akan menerusi antaramuka 1,
dan kembali ke Pemindaharuh Penerima yang asalnya daripada antaramuka 2.

71

99.88 % 97.85 % iaitu 97.73 %(daripada keamatan asal yang dipancarkan, dipantulkan kembali daripada
antaramuka 3 dan perlu melalui antaramuka 2 dan 1. (hampir 100% pantulan).
Apabila sampai ke antaramuka 2, 98.92% daripada 97.73 % iaitu 96.67 % akan menerusi antaramuka 2 dan
terus ke antaramuka 1.
Apabila mengenai antaramuka, 1.08 % daripada 96.67% iaitu 1.04 % akan dipantulkan kembali ke antaramuka
2. Manakala 98.92 % daripada 96.67% iaitu 95.62 %.
Jumlah peratus pantulan yang diterima pada pemindaharuh penerima adalah = 1.08 % + 1.06 % + 95.62 % =
97.76 %
(dengan mengabaikan pantulan berulang-ulang)
72

18

Gelombang ultrasound bertemu dengan


sempadan yang memisahkan dua media
yang berlainan di dalam badan manusia

Sumber
Ultrasound

Pemindaharuh

Lemak

Otot

Antaramuka 1

Solution

Tulang

Antaramuka 3

Antaramuka 2

73

Gelombang ultrasound bertemu dengan


sempadan yang memisahkan dua media
yang berlainan di dalam badan manusia

Sumber
Ultrasound

Pemindaharuh

Tulang

Otot

Antaramuka 1

74

Solution

Lemak

Antaramuka 3

Antaramuka 2

75

76

19

Sound energy & effect on soft


tissue

Sound energy & effect on soft


tissue

So how does sound wave behave when it travels through human


tissues?
From the air medium, it must enter the skin/fat which has significantly
higher density. There is 100% reflection of the sound wave at the airskin interface. If we put a coupling medium such as a gel to create a
sound head-gel-skin interface, reflection is reduced to only 0.1%; the
rest of the sound energy will be transmitted through the skin barrier. As
noted in the absorption coefficients table, sound energy travels through
much of the soft tissue without much absorption until it reaches tissues
with high collagen (bundles of protein fibres) content, namely bone,
periosteum, ligaments, capsules, fascia, tendons, and tissue interface
(bursa).
Ultrasound at high intensity near bony areas can be detrimental to the
periosteum (dense layer of connective tissue around a bone) because
of high energy accumulation and heating effect on the soft tissue as
sound wave hits the bone (transverse or shear wave).

Ultrasound energy is absorbed mostly in


tissue with high collagen (bundles of protein
fibres) content (bone, periosteum, ligaments,
capsules, fascia, tendons, and tissue
interface i.e. bursa & synovium).

77

What happens to the tissues that


absorb sound energy?

78

Why US is not good for bone


imaging
Ultrasound has selective
interfacial effects at the bone
surface resulting from bone
having a high absorption
coefficient, a high relative
acoustic impedance, and an
ability to propagate shear
waves.
Unlike x-rays, ultrasound will
not go through bone or
air. Therefore, ultrasound is
not a good way to diagnose
bone or lung disease. Its real
value comes from the
examination of internal organs.

Sound energy is non-ionizing radiation and


therefore its use does not impose the hazard,
such as cancer and chromosome breakage,
attributed to ionizing radiation.
Sound energy has two physiological effects:
1. Enhance inflammatory response and tissue repair.
2. Heat soft tissue

79

80

20

Ultrasound Transducer
Pieozoelectric

Ultrasound Transducer

81

1.3. Ultrasound Transducer

82

I. Basic
1.3.
Ultrasound
Ultrasound
Transducer

Piezoelectric Effect

Piezoelectric Effect

+
Battery

Piezoelectrical
Crystal (Quartz)

The crystal gets thicker, due to a distortion of the crystal


lattice

21

Piezoelectric Effect

Piezoelectric Effect
Sound wave
with
frequency f

U(f)

+
The effect inverses with polarity change

Piezoelectric Effect

An alternating voltage generates crystal oscillations at the


frequency f

Reception of ultrasonic waves


A sound wave hitting a piezoelectric crystal, induces
crystal vibration which then causes electrical voltages
at the crystal surfaces.

Short pulse
( < 1 s )

Electrical
energy

Piezoelectrical
crystal

Ultrasonic wave

A short voltage pulse generates an oscillation at the crystals


resonant
frequency f0

22

Ultrasound Transducer
Fungsi pemindaharuh ultrasonik
A transducer is a device that can convert
one form of energy into another.
Ultrasonic transducers are used to convert
an electric signal into ultrasonic energy
that can be transmitted into tissues, and to
convert ultrasonic energy reflected back
from the tissues into an electric signal

Ultrasound Transducer
The piezoelectric (pressure-electric) effect
enable transducer to transmit and receive the
ultrasound signal (pressure wave).
The crystal has a natural vibrational frequency
that is related to the distance between two
surfaces.
For example 1.5 MHz transducer corresponding
to wavelength of 2.7 mm in the crystal material
has a crystal thickness of 1.35 mm. A higher
frequency transducer that produces a shorter
wavelength has an even thinner crystal.

89

I. Basic
1.3.
Ultrasound
Ultrasound
Transducer

90

I. Basic
1.3.
Ultrasound
Ultrasound
Transducer

Transducer Construction

Transducer Construction

The construction of the transducer can be viewed as a crystal of piezoelectric


material with electrodes on each side of the crystal. The entire crystal is housed in
an acoustically and electrically insulated casing. The backing material should
absorb all the energy, and the matching layer should transfer the output maximally.
91

1.3. Ultrasound Transducer

92

1.3. Ultrasound Transducer

23

The general composition of an


ultrasonic transducer

Transducer Design

The most important component is a thin (approximately 0.5mm) piezoelectric crystal element located near the face of the
transducer.
The front and back faces of the crystal are coated with a thin
conducting film to ensure good contact with the two electrodes
that will supply the electric field used to strain the crystal.
The term strain refers to deformity of the crystal caused
when a voltage is applied to the crystal.
The surfaces of the crystal are plated with gold or silver
electrodes.
The outside electrode is grounded to protect the patient from
electrical shock, and its outside surface is coated with a
watertight electrical insulator.
The inside electrode abuts against a thick backing block that
absorbs sound waves transmitted back into the transducer.
The housing is usually a strong plastic.
An acoustic insulator of rubber or cork prevents the sound
from passing into the housing.
A large variety of sizes and shapes of transducers are
available to perform specific functions, but they all have this
general design
94

1.3. Ultrasound Transducer

Advantages of ultrasound transducer which is


made with Piezoelectric Ceramics material
A great advantage of piezoelectric
Ceramics is that they can be formed
into different shapes.
Piezoelectric crystals can be
designed to vibrate in either the
thickness or radial mode
Medical crystals are designed to
vibrate in the thickness mode.
They still vibrate to a lesser extent in
the radial mode, however, so the
receiving amplifier is gated to tune
out all frequencies except those from
the thickness mode

Ketebalan & frekuensi salunan


Pemindaharuh ultrabunyi direkabentuk supaya
mempunyai kepekaan yang tinggi terhadap
frekuensi tabii (natural frequency).
Ketebalan hablur piezoelektrik (piezoelectric
crystal) menentukan frekuensi tabiinya atau
dipanggil frekuensi salunan.
Selalunya hablur direkabentuk supaya
ketebalannya adalah bersamaan dengan
separuh daripada nilai panjang gelombang
ultrabunyi yang perlu dihasilkan oleh
pemindaharuh

95

1.3. Ultrasound Transducer

96

1.3. Ultrasound Transducer

24

Contoh

Penyelesaian

Ketebalan hablur PZT-4 = 0.001 m dan


halaju bunyi di dalam PZT-4 = 4000 m/sec
Kirakan frekuensi salunan pemindaharuh
ini

Diberi,
V (PZT-4) = 4000 m/sec
Ketebalan PZT-4 = 0.001 m
Oleh kerana, hablur akan menyalun (resonate) pada frekuensi (f)
bersamaan dua kali ganda ketebalan hablur.
Maka = 2f = 2 0.001 m = 0.002 m.
Dengan menggunakan formula berikut;
= f.
f (frekuensi) = /
= 4000 m/sec / 0.002 m
= 2,000,000/sec
= 2 MHz

97

1.3. Ultrasound Transducer

Sekiranya ketebalan hablur


berkurang atau bertambah

98

1.3. Ultrasound Transducer

Frequency Response

Ketebalan PZT-4 iaitu 0.001 m (1 mm), frekuensi


salunannya adalah 2 MHz
Ketebalan PZT-4 iaitu 0.002 m (2 mm), frekuensi
salunannya adalah 1 MHz
Daripada contoh ini dapat ditunjukkan bahawa
sekiranya ketebalan hablur yang digunakan berkurang
atau makin nipis, maka frekuensi salunan yang
dihasilkan adalah meningkat. Ini menunjukkan
bahawa sekiranya kita mengkehendaki hablur
menyalun pada frekuensi megahertz yang tinggi maka
ketebalan hablur yang diperlukan akan menjadi terlalu
nipis

Quality factor
Q = fo/f
Q = Estored/Elost

99

1.3. Ultrasound Transducer

100

I. Basic
1.3.
Ultrasound
Ultrasound
Transducer

25

Frequency Response

Ultrasound Probes

Quality factor
Q = fo/f
Q = Estored/Elost

The Q value relates the damping characteristics to the pulse length and can
be viewed as the energy stored per cycle divided by energy lost per cycle.
A high Q transducer means that most of energy is stored and very little energy is
lost each cycle. Low Q for Diagnostic and High Q for therapeutic. Shorter pulse
length better axial resolution. Frequency increased, pulse length decrease,
improved axial resolution. Q indirectly proportional to BW. High Q has very narrow
bandwidth. Low Q transducer has broad bandwidth and better receiver of the
transducer. High Q better transmitter.

Sector Probe Convex Probe Linear Probe

101

I. Basic
1.3.
Ultrasound
Ultrasound
Transducer

102

1.3. Ultrasound Transducer

Scanning Beam

Ultrasound in Medicine

Heart

Skin, Orthopedic

Abdomen

103

1.3. Ultrasound Transducer

104

II. Ultrasound in Medicine

26

II. Ultrasound in Medicine

Modes
A Mode - amplitude mode. Where the signals are
displayed as spikes that are dependent on the amplitude
of the returning sound energy.
B Mode - brightness mode. Where the signals are
displayed as various points whose brightness depends
on the amplitude of the returning sound energy.
M Mode - motion mode. The application of B-mode and a
strip chart recorder allows visualization of the structures
as a function of depth and time.
2D Mode - 2 dimensional mode name is usually reserved
to indicate b-mode imaging of the heart. The spatially
oriented B-mode where structures are seen as a function
of depth and width. The beam is rapidly swept back and
forth to create a cross section of the imaged structures.

2.1. Ultrasound Diagnostic


A Mode
B Mode
M Mode
D Mode
3D / 4D Mode

2.2. Ultrasound Therapeutic


2.3. Biological Effects
105

II. Ultrasound in Medicine

A-mode Ultrasound

106

2.1. Ultrasound Diagnostic

A-mode

Amplitude of reflected signal vs. time

27

2.1. Ultrasound Diagnostic

2.1. Ultrasound Diagnostic


A-Mode Ultrasound
On an oscilloscope, this simple clap-echo
system would look like this.
The initial spike from the clap would be
followed some time later by the echo. The
earliest "A-Mode" ultrasound machines
worked in this way. You could know how far
the echo has traveled, and how loud the echo
was when it got back to you.
There were (and are) several problems with
this simple system:

A-Mode

You don't know the exact direction it came


from.
You don't know for sure what the echo
bounced off of.
You don't know what the object generating the
echo looks like.
109

II. Ultrasound in Medicine

110

II. Ultrasound in Medicine

B-Mode

B-mode Ultrasound

111

2.1. Ultrasound Diagnostic

28

B-mode Example

Cardiac Ultrasound

Standard Sites for Echocardiograms

29

Conventional Cardiac 2D Ultrasound

Short-axis Interrogation

B-mode Image of Heart

B-mode

30

Mitral stenosis

M-Mode

M-mode Ultrasound

M Mode - motion mode. The application of


B-mode and a strip chart recorder allows
visualization of the structures as a function
of depth and time.

123

II. Ultrasound
2.1.
Ultrasound
inDiagnostic
Medicine

31

M-mode

M-Mode

126

II. Ultrasound
2.1.
Ultrasound
inDiagnostic
Medicine

Doppler Mode

Doppler Mode

DOPPLER PRINCIPLES
Christian Johann Doppler described the effect of motion of sound
sources and its effect on the frequency of the sound to the observer.
In medical applications we find that the frequency of the reflected
signal is modified by the velocity and direction of blood flow. If blood
cells are moving towards the transducer, they increase the
frequency of the returning signal. As cells move away from the
transducer, the frequency of the returning signal decreases.
Doppler techniques are dependent on the transducers used. The
transducer operating in continuous wave mode utilizes one half of
the element(s) and are continuously sending sound energy while the
other half is continuously receiving the reflected signals.
If the transducer is being used in a pulsed wave mode, the whole
transducer is used to send and then receive the returning signals.
The display of Doppler velocity data is the Doppler frequency shifts
versus time. Included in the display are the Doppler settings such as
frequency, calibration, range, and timing markers.
127

II. Ultrasound
2.1.
Ultrasound
inDiagnostic
Medicine

128

II. Ultrasound
2.1.
Ultrasound
inDiagnostic
Medicine

32

kesan Doppler (Doppler Effect)

Doppler Principle

Kesan Doppler adalah pengurang di


dalam frekuensi yang diterima oleh
pemerhati daripada sumber bunyi apabila
ianya bergerak melepasi dan berundur
(bergerak menjauhi) daripada pemerhati

129

2.1. Ultrasound Diagnostic

Doppler

Doppler measurements

33

Doppler angle

Normal flow

Diseased flow

Blood Flow Measurements

34

3-D Mode

4-D Mode

137

II. Ultrasound
2.1.
Ultrasound
inDiagnostic
Medicine

138

II. Ultrasound
2.1.
Ultrasound
inDiagnostic
Medicine

2.2. Ultrasound Therapeutic


Basic Principle:
Ultrasound Waves
Compression & Rarefaction
Molecular Vibration in
Tissue Thermal Effect
(Heating) & Non Thermal
Effect (Micro massage)

Ultrasound Therapeutic

Frequency = 3 MHz,
Wavelength in Tissue = 0.5
mm

139

II. Ultrasound in Medicine

140

II. Ultrasound in Medicine

35

Ultrasound Absorption in Tissue

Ultrasound Absorption in Tissue


Although cartilage and bone are at the upper end of this scale,
the problems associated with wave reflection mean that the
majority of US energy striking the surface of either of these
tissues is likely to be reflected. The best absorbing tissues in
terms of clinical practice are those with high collagen content
LIGAMENT, TENDON, FASCIA, JOINT CAPSULE, SCAR
TISSUE (Watson 2000, ter Haar 99, Nussbaum 1998, Frizzel &
Dunn 1982)
The application of therapeutic US to tissues with a low energy
absorption capacity is less likely to be effective than the
application of the energy into a more highly absorbing material.
Recent evidence of the ineffectiveness of such an intervention
can be found in Wilkin et al (2004) whilst application in tissue
that is a better absorber will, as expected, result in a more
effective intervention (e.g. Sparrow et al 2005, Leung et al
2004).

the tissues with the higher protein content will absorb


US to a greater extent
the tissues with high water content and low protein
content absorb little of the US energy (e.g. blood and
fat)
141

142

2.2. Ultrasound Therapeutic

2.2. Ultrasound Therapeutic

2.3. Biological Effects


No acute harmful effects have been
reported after diagnostic ultrasound
examinations.
Biological effects observation: high
intensity level and long exposure time
To assess biological effects, the dose of
ultrasound must be known.
Dose: quantity of energy absorbed per unit
mass of absorbing medium = intensity in
the medium

Biological Effects

143

144

II. Ultrasound in Medicine

36

Interaction of Ultrasound with Matter


Three mechanism: Thermal, Cavitation, Mechanical
Thermal mechanism: Ultrasound energy is absorbed and
converted into heat. The increased temperature may cause
irreversible tissue damage. The rate of temperature rise depends on
TA (Temporal Average) intensity, absorption coefficient of the
medium and heat transport process (thermal conductivity and blood
flow). The absorption coefficient increases with frequency.

Interaction of Ultrasound with Matter


Mechanical mechanism: The ultrasonic wave is propagated through
the medium by interactions between neighboring particles. This
particles undergo considerable changes in velocity and acceleration.
An object of different density than that of the surrounding medium will
experience a torque in ultrasonic field. At high intensities, the wave
shape becomes distorted, resulting in micro streaming. These high
velocity gradients can fragment the macromolecules located in these
region.

Cavitation mechanism: As the ultrasound wave propagates


through the medium, regions of compression and rarefaction are
created. The localized regions in the medium are subjected to
increases and decreases in pressure in alternating fashion. These
pressure changes can cause gas bubbles to form and/or grow and
exhibit dynamic behavior.
145
2.3. Biological Effects

146

2.3. Biological Effects

FDA Regulation 1976

FDA Regulation 1993


The new regulations combine an overall limit of
I-SPTA of 720 mW/cm2 for all equipment with a
system of output displays to allow users to employ
effective and judicious levels of ultrasound
appropriate to the examination undertaken.
The new regulations allow an eight-fold increase
in ultrasound intensity to be used in fetal
examinations.
The output display is based on two indices, the
mechanical index (MI) and the thermal index (TI).

For spatial peak time-averaged intensity (ISPTA) (the measure most associated with
temperature rise), the maximum levels
were:
Ophthalmic

17 mW/cm2

Fetal and other

94 mW/cm2

Cardiac

430 mW/cm2

Peripheral vessel

720 mW/cm2
147

148

37

Medical Ultrasound
Advantages of
acoustic energy:
can be directed in a
beam
obeys the laws of
reflection and
refraction
reflected off object
borders
no known unwanted
health effects

Why Ultrasound in Cardiology?

Disadvantages of
acoustic energy:
propagates poorly
through a gaseous
medium
reflected off of borders
of small objects
quickly dissipates (as
heat)

Portable, relatively cheap


Non-ionizing
During the echocardiogram, it is possible for the
cardiologist to:
Watch the hearts motion in 2D real-time
Ascertain if the valves are opening and closing properly,
and view any abnormalities
Determine the size of the heart chambers and major
vessels
Measure the thickness of the heart walls
Calculate standard metrics of health/disease
e.g., Volume, EF, SV, CO

Dynamic evaluation of abnormalities

III. Devices Standard


3.1. Ultrasound Medical Diagnostic
3.2. Ultrasound Medical Therapeutic

Devices Standard

151

152

III. Devices Standard

38

IEC 60601

IEC 60601

IEC 60601-1: Medical electrical equipment - Part 1: General


requirements for safety
IEC 60601-1 ed 2.2: Medical electrical equipment- Part 1:
General requirements for safety
IEC 60601-1-1 ed 2.0: Medical electrical equipment- Part 1:
General requirements for safety-1. Collateral standard:
Safety requirements for medical electrical systems
IEC 60601-1-2 ed 1.0: Medical electrical equipment- Part 1:
General requirements for safety-2. Collateral Standard:
Electromagnetic compatibility - Requirements and tests
IEC 60601-1-4 ed 1.1: Medical electrical equipment-Part 1:
General requirements for safety-4. Collateral standard:
Programmable electrical medical systems
IEC 60601-2-37 ed 1.1: Medical electrical equipment- Part
2-37: Particular requirements for the safety of ultrasonic
medical diagnostic and monitoring equipment
153
III. Devices Standard

This standard establishes general safety requirements for


all aspects of medical devices from light boxes to beds to
high-end diagnostic equipment like ultrasound or magnetic
resonance (MR). Included are test requirements,
documentation, protection from electrical hazards,
protection from mechanical hazards, protection against
excessive or unwanted radiation, protection against
temperature, fire prevention, ingress of liquids, disinfection,
biocompatibility etc.
IEC 60601-1 is presently at edition 2.2 and was last
published in 1997. This standard is presently in the final
stages of a major overhaul to edition 3 due out in mid-2005.
As the release date gets closer, a separate column will
discuss the key differences of this future revision.
154

III. Devices Standard

Potential sources of harm of


ultrasound diagnostic equipment

Marking and Display

Noise on a waveform, artifacts, distortion in an image


or error of a displayed numerical value which cannot
be attributed to a physiological effect and which may
alter the diagnosis
The display of inaccurate numerical values
associated with the diagnosis to be performed
The display of inaccurate safety related indications
The production of unintended or excessive
ultrasound output
The production of unintended or excessive transducer
assembly surface temperature
The production of unintended or uncontrolled motion
of transducer assemblies intended for intra-corporeal
use

Marking on the outside of equipment for:


Compliance with EMC requirement
Output adjustable
Physiological effect, symbol and warning
statements

Display of controls and instruments


A display of Thermal and Mechanical Index
A display of surface temperature of
ultrasonic transducer
A display of ultrasound output level

155

3.1. Ultrasound Medical Diagnostic

156

3.1. Ultrasound Medical Diagnostic

39

Marking

Standard in Therapeutic Devices

Identical to Diagnostic Devices and


replace with:
IEC 60601-2-5: Medical electrical
equipment - Part 2-5: Particular
requirement for safety of ultrasonic
physiotherapy equipment

The generator shall be provided following marking:


Acoustic working frequency in MHz or kHz
Waveform type
Description of output waveform (Duration, Pulse
Repetition Period, Duty factor)
The generator shall carry a nameplate, with unique
serial number
The treatment head shall be marked with its rated
output power in watts, the effective radiating area in
square centimeters, the beam non uniformity ratio,
beam type, a designation of the specific generator (if
applicable) and a unique serial number.

157

III. Devices Standard

158

3.1. Ultrasound Medical Therapeutic

IV. Ultrasound Instrument

4.1. Ultrasound Machine

4.1. Ultrasound Machine


4.2. Block Diagram

159

IV. Ultrasound Instrument

160

IV. Ultrasound Instrument

40

4.2. Block Diagram

161

IV. Ultrasound Instrument

Performance Parameters

V. Performance & Safety Testing

5.1. Performance Testing


Parameters
Phantom

5.2. Safety Testing


Electrical Safety
MI/TI
Acoustic Power

5.3. Test Record

Resolution: Axial & Lateral


Dead Zone
Beam width
Focal Zone
Geometry Accuracy
TGC Characteristic
Uniformity
Sensitivity
Penetration
Low contrast detection

163

V. Performance & Safety Testing

164

5.1 Performance Testing

41

Test Objects and Phantom

AIUM STANDARD TEST OBJECT


Diagram of the AIUM
100 mm x 100 mm
standard Test Object
The rods are 0.75 mm
diameter stainless steel
wires placed at specific
locations with an
accuracy of 0.25 mm.
The material may be
water at 37oC, alcoholwater solution or NaClWater Solution

Test object is composed of material in


which the velocity of ultrasound is the
same as in tissue, but other properties of
the material with respect to sound
propagation may be very different from
that of tissue.
Phantom is tissue equivalent test object.

165

5.1. Performance Testing

166

5.1 Performance Testing

Tissue-Equivalent Phantoms

Phantom

The tissue-equivalent phantom are composed of


animal hide gelatin impregnated with graphite
powder or polysaccharide gel (agar)
impregnated with graphite powder.
These materials mimic soft tissue in terms of the
speed of sound, attenuation, density and scatter.
Nylon fibers (0.3 mm in diameter) are placed at
various location to evaluate the same
parameters as discussed for test object.
167

5.1 Performance Testing

168

5.1 Performance Testing

42

Dead Zone

Geometric Accuracy

The dead zone corresponds to the region in


which no information is obtained.

Vertical Distance Accuracy

Horizontal Distance Accuracy

169

5.1 Performance Testing

170

5.1 Performance Testing

3-D Egg-Phantom

Axial Resolution

171

5.1 Performance Testing

172

5.1 Performance Testing

43

Lateral Resolution

Penetration (Maximum Depth)

173

5.1 Performance Testing

174

5.1 Performance Testing

5.2. Safety Testing

Visual Inspection

Visual / Mechanical: Probe, Cable, Connector


Electrical Safety
Mechanical Index & Thermal Index
Acoustic Power

Transducers free of cracks, delaminations


Cables in good shape
Transducers cleaned after each use
Viewing monitors clean
Air filters
Wheels, locks

175

5.2. Safety Testing

176

5.2. Safety Testing

44

Excessive Acoustic Power

Cable Failure

177

5.2. Safety Testing

178

5.2. Safety Testing

Probe Failure

Probe Failure
Array housing separating from the seam normally
occurs when the probe has been in use for some time.
This glue can deteriorate as a function of time as well
as to exposure to cleaning agents. This probe should
not be used as it may present an electrical shock
hazard to the patient and the sonographer. Further, if
cleaning agents get inside the probe they will normally
destroy the array. This type of failure can be repaired if
caught early.
Seam line separating from nose cone becoming the
site of fluid infiltration. Less obvious than the example
above, this is the beginning of the cap detachment.
The compromise of the seam can be seen with the
aide of a small magnifying glass. If the problem is
found at this stage, the probe is easily repairable.
179

5.2. Safety Testing

180

5.2. Safety Testing

45

Acoustic Power Measurement

Electrical Safety Testing

Existing ultrasound power measurement


devices generally use a force balance
approach.
The radiation force is a static force
produced by the rate of change of
momentum in the ultrasound beam.

Protective Earth Verification: Ground


Continuity Test
High Resistance: Insulation Resistance
Test
Leakage Current: Earth Leakage,
Touch/Chassis (Enclosure), Patient
Leakage & Patient Auxiliary Leakage

where

is the incident angle of the ultrasound.

181

5.2. Safety Testing

182

5.2. Safety Testing

Mechanical Index

Acoustic Power Measurement


Incident ultrasound energy is reflected off a nominally perfect
reflector, thereby applying force to the reflector. With existing meters
this force is generally measured using a sensitive electronic
balance, or using a force feedback system.
When a reflecting or absorbing target is placed in an ultrasonic
beam within a fluid, it experiences a force known as the 'radiation
force'. The radiation force is a static force produced by the rate of
change of momentum in the ultrasound beam. This force, F,
depends only upon the total acoustic power (P), the velocity of the
sound in fluid (C) and the geometry of the target.
In the case of a large reflective target, the force, F, is related to the
total power P, in the ultrasonic beam by the equation

183

The mechanical index is an estimate of the maximum


amplitude of the pressure pulse in tissue.

It may be calculated using the spatial peak value of the


rarefactional pressure P3, derated by 0.3 dB/cm/MHz at
each point along the beam axis, divided by the square
root of the centre frequency fc
It gives an indication as to the relative risk of mechanical
effects (streaming and cavitation).
The FDA regulations allow a mechanical index of up to
1.9 to be used for all applications except ophthalmic
(maximum 0.23).
184
5.2. Safety Testing

46

Thermal Index

Thermal Index
For fetal scanning, the highest
temperature increase would be expected
to occur at bone and TIB would give the
worst case conditions.
The mechanical index and thermal index
must be displayed if the ultrasound system
is capable of exceeding an index of 1.
The displayed indices are based on the
manufacturers experimental and modelled
data.

The thermal index is the ratio of the power used to that


required to cause a maximum temperature increase of
1C.
A thermal index of 1 indicates a power causing a
temperature increase of 1C.
Because temperature rise is dependent on tissue type
and is particularly dependent on the presence of bone,
the thermal index is subdivided into three indices:
(1) TIS: thermal index for soft tissue;
(2) TIB: thermal index with bone at/near the focus;
(3) TIC: thermal index with bone at the surface (e.g. cranial
examination).
185

5.2. Safety Testing

186

5.2. Safety Testing

5.3. Test Record

Visual Inspection

Acceptance Testing for New Equipment by


Manufacturer / Supplier / Commissioner /
Technical Person
Acceptance Testing for After Repair Equipment
by Manufacturer / Supplier / Commissioner /
Technical Person
Routine Operation Testing by User / Operator
(daily or weekly)
Routine Maintenance Testing by Technical
Person / Engineer / Technician / Accreditation
Body (Every 1/3/6/12 months)
Calibration / Accreditation Testing by
Accreditation Body (Every 1/2/3/5 years)

Packaging
Documentation
Marking
Overall
Accessories

187

V. Performance and Safety Testing

188

5.3. Test Record

47

Functional Inspection for


Diagnostic

Safety Inspection

Image Quality:

Main Supply Connections


Earth Continuity
Insulation
Leakage Current
Probe Condition

Axial Resolution, Lateral Resolution, Dead Zone,


Sensitivity, Uniformity, Focal Zone Distance,
Photographic System, Horizontal Depth Marker
Error, Vertikal Depth Marker Error, Beam Width,
Geometry Accuracy, TGC Characteristics,
Penetration, Low contrast detection

Numerical Display: TI, MI


Acoustic Power
Surface Temperature
189

5.3. Test Record

190

5.3. Test Record

Functional Inspection for


Therapeutic

Effective radiating area


Rated output power
Acoustic Working Frequency
Beam Non Uniformity Ratio
Pulse Duration
Pulse Repetition Period
SPTA Intensity of unwanted ultrasound radiation (less than 100
mW/cm2)
Temperature of Radiating Surface (not exceed than 41oC)
Output power and Effective Intensity
Output Stability with supply variation
Output Stability with time
Timer Accuracy
191

5.3. Test Record

48

You might also like