Professional Documents
Culture Documents
sponsored by
The College of
Optometrists
Figure 2
A heavy film deposition on a soft lens easily
seen macroscopically. The effect on visual
acuity and contrast is likely to be significant.
A new lens is also shown for comparison
27
Code
C
G
F
P
Co
D
(type)
Crystalline
Granular
Film
Plaque
Coating
Debris
Code
a
b
c
d
Figure 3
Large mulberry or jelly bump
deposits on a soft lens
Figure 4
A rust spot on a soft lens. Size is likely to be 200m
to 500m. The ferrous particle may or may not
remain in the lens
Figure 5
Two hydrogel lenses showing discolorations - one a
pink colour and the other a grey
29
Figure 7
Dark field illumination. A plan view of a soft lens in
an Optimec instrument whilst the BOZR is
measured. The lens coating would show up as a
whitish-grey under direct illumination, but there is
some fluorescence here which gives a bluish tinge to
the deposition
Figure 8
High magnification of a very heavy surface film that
has started to break off the lens surface
Surfactant cleaning
This is an essential step with lenses that are
not worn once and thrown away. If a lens
has to be used again, it should be subjected
to surfactant cleaning and a disinfection
process. Some wearers may omit the
cleaning step with a surfactant. Users of
hydrogen peroxide systems are probably
the major offenders. The main purpose of
cleaning lenses between finger and thumb
whilst using a surfactant cleaner is to
physically remove the loosely attached
deposition. This will include mucin,
bacteria, polysaccharides, debris as well as
other substances such as proteins which
have not denatured. Removing large
numbers of microbes in this way makes the
actual disinfection process more efficient.
Rubbing with a surfactant cleaner and
rinsing off well afterwards is of paramount
importance in lens hygiene. Some cleaners
are alcohol-based and these will dissolve
some organic materials more easily.
In the case of RGP lenses, a polymeric
bead cleaner has a mildly abrasive action
and lipids and protein which are firmly
attached can be physically removed from
the lens surface. However, practitioners
should always sound a note of caution to
users of these products. Some wearers are
far too vigorous in the rubbing action and
the abrasive effect has been known to
change the power of the lens and/or make
the centre so thin that the lens is fragile
and prone to warpage.
An electro-mechanical cleaner has
been developed (Ifejika, 1998) which can
significantly enhance the performance of
multipurpose solutions - both from the
cleaning and disinfection aspects. The
disinfection performance is impressive
even when plain saline is used. It is
Figure 9
Cohesion and adhesion forces, (a) with mercury on glass with a contact angle of about
130 and (b) with water on PMMA with a contact angle of about 60
Figure 10
The surface tension of a
liquid drop. Inward pull of
surface tension results in a
spherical shape
SURFACE PROPERTIES
Forces acting on molecules of the same type
are known as cohesive forces. Forces acting
between different substances are known as
adhesive forces. An example is shown in
Figure 9. Mercury has a high cohesive
force which keeps a drop in a spherical
shape. There is also relatively little
adhesive force between the mercury and
glass. Conversely, when a drop of water is
placed on PMMA, water has a lower
cohesive force and the adhesive force
between the PMMA and water is higher
than that between mercury and glass - as a
result, the water spreads further than the
mercury.
Critical surface tension
If a liquid has a surface tension below this
value, it will spread on a given solid surface.
If it has the same value or higher, the liquid
will not spread on the solid but it will form
drops. In the latter case, the surface is nonwetting.
Surface tension
This is a tension force developed on a
surface as a result of asymmetrical
molecular interactions at an interface. In
Figure 11
The attractive forces between the molecules of a homogeneous liquid in the inner bulk
are contrasted to the forces at the liquids surface. The surface tension of a liquid is
created by the unopposed pull of the surface molecules which have higher potential
energy than the bulk molecules
31
sponsored by
Figure 12
The relationship of the three interfaces when a drop of liquid is
placed on a solid surface; both the solid and liquid are in air
Figure 13a
An advancing contact angle ()
where liquid from a syringe is
increasing the volume of the drop.
The drop is encroaching on to
unwetted solid
Table 1
SOLUTION
CONTACT ANGLE
HIGHEST
LOWEST
Boston IV (Prymesoak)
26 (Delta (B&L))
63
Equalens
32 (Delta)
42 (ReNu)
32
Figure 13b
A receding contact angle () where
liquid is being withdrawn from the drop
up the syringe. The circumference of
the drop is moving to an area which
has already been wet
sponsored by
Figure 14
The arrangement of apparatus to measure the contact angle with a laser. The laser beam
strikes a point on the circumference of the liquid adherent to the lens surface. Two lines
are produced on a protractor and the angle between these lines is the contact angle
Figure 15
The wetting angles of five rigid lenses of different material. Each material is shown with a
pre-wear and post-wear wetting angle. The angles were determined using the adherent
liquid-laser method
Surface hydrophilicity
The wettability of a surface can be assessed
using the contact angle (or wetting angle)
as a measurement of this property. It relates
to how well a liquid spreads across a
surface. Hydrophilicity is different in as
much as this property determines the
strength of attraction of a liquid to a lens
surface. In Figure 17, a contact lens has its
front surface immersed in a liquid (a). The
lens is gradually withdrawn from the liquid
and a point is reached when the separation
force is at its maximum (b), and then as the
lens is further withdrawn, the separation
force declines (c) and, finally, there is a
separation of the lens surface from the
Maximum force
380mg
405mg
360mg
33
sponsored by
Figure 16
The wetting angles of eleven hydrogel materials of different water content. All the postwear values are approximately the same at 2. The angles were determined using the
adherent liquid-laser method
Figure 17
In order to assess hydrophilicity, the contact lens has its front surface immersed in liquid
(a). The lens is withdrawn gradually until a maximum force is registered (b). Further
withdrawal results in the force reducing (c) and eventually the liquid separates from the lens
Figure 18
Changes in the withdrawal force as a function of withdrawal distance.
As can be seen, the maximum force needed occurs at a specific or
critical distance and this maximum force value gives the hydrophilicity
34
Figure 19
The pre-wear and post-wear values for hydrophilicity as
represented by the maximum force values for five different rigid
lenses
Acknowledgements
The author is indebted to the following
main sources of material:
Figure 20
The pre-wear and post-wear values for hydrophilicity as represented by the maximum
force values in three different soft lens materials
Michael Port is a senior lecturer in the Department of Optometry and Visual Science at
City University. He is responsible for second year clinical practice and contact lenses.
An answer return form is included in this issue. It should be completed and returned to:
CPD Initiatives (CDM8), Optometry Today, Victoria House, 178-180 Fleet Road, Fleet, Hampshire GU13 8DA, by September 8, 1999.
35
36
d is correct
Although all of the above have been
recorded as producing a toxic ocular
response, the most frequent reports and
the most severe responses have been with
thiomersal.
a is correct
Tomlinson and Cedarstaff first
demonstrated that contact lens wear
increases the rate of tear evaporation in
1982.
b is correct
There are no reports of lenses disinfected
by heat having an adverse effect on the
eye. There may be adverse effects on the
lenses.
b is correct
Atopic patients have been found to have a
greater number of corneas in the lower end
of the normal distribution of corneal
thickness.