CPI DIGITAL INTERFACE OPTION 40


FOR CMP 200 DR X-RAY GENERATOR
CONTENTS:

1.0 INTRODUCTION............................................................................................................................................. 2
2.0 INSTALLATION............................................................................................................................................... 2
2.1 CONNECTIONS .......................................................................................................................................... 2
3.0 CONFIGURATION AND PROGRAMMING .................................................................................................... 2
THEORY OF OPERATION ........................................................................................................................................ 4
3.1 GENERAL.................................................................................................................................................... 4
3.2 GENERATOR OUTPUT SIGNALS ............................................................................................................. 4
3.2.1 PREP REQUEST (EXP_PREP)........................................................................................................... 4
3.2.2 EXPOSURE REQUEST (EXP_REQ) .................................................................................................. 4
3.2.3 EXPOSURE ON (EXON) ..................................................................................................................... 4
3.3 GENERATOR INPUT SIGNALS ................................................................................................................. 5
3.3.1 EXPOSURE CONFIRMATION (EXP_OK)........................................................................................... 5
3.4 INPUTS and outputs TIME diagram............................................................................................................ 5
4.0 Schematic........................................................................................................................................................ 6

PROPRIETARY AND CONFIDENTIAL

The design and information contained in this document is proprietary to Communications & Power Industries
Canada Inc. (CPI), has been designed and developed at private expense, and is the exclusive property of CPI.
The information contained herein is loaned in confidence solely for the installation, operation, and maintenance of
CPI equipment. This document may not be duplicated in whole or in part, or be used for re-engineering, reverse
engineering, or otherwise reproducing in any form or creating or attempting to create or permitting, allowing or
assisting others to create or manufacture CPIs product or products derived therefrom, without the express written
permission of CPI.

Copyright by Communications & Power Industries Canada Inc. All rights reserved.

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Manual Supplement # 903124 Rev. B Page 1


CPI Canada Inc. Supplement

1.0 INTRODUCTION

This supplement contains information regarding CPI digital interface option for the CMP 200 DR
generator.

2.0 INSTALLATION

2.1 CONNECTIONS

The CPI interface option is a software interface that is programmed through the generator console or
service utility.

CPI Interface Connections


CPI Interface Signal Generator Signal Type Schematics Ref.
Name Connection
Outputs From the Generator
J25-1 +24V when inactive DIGITAL_OUT1\
EXP_REQ J25-4 +24V when active DIGITAL_OUT_P1
J25-30, 12 Differential Output DIGITAL_OUT1A/B
J25-20 +24V when inactive DIGITAL_OUT2\
EXP_PREP J25-23 +24V when active DIGITAL_OUT_P2
J25-31, 13 Differential Output DIGITAL_OUT2A/B
J25-21 +24V when inactive DIGITAL_OUT6\
EXON J25-25 +24V when active DIGITAL_OUT_P6
J25-34, 16 Differential Output DIGITAL_OUT6A/B
Inputs To The Generator
EXP_OK J25-35, 17 Differential Input DIGITAL_IN5A/B
+24V and the GND
+24V J25-3, 22 +24V power supply
GND J25-M1, M2, 7, 26, 29 Signal Ground

3.0 CONFIGURATION AND PROGRAMMING

To program the Generator for the CPI interface option operation, please follow the steps below:

a) Make sure the license key supplied with the Generator unlocks the digital interface options
feature. The unlocked features can be viewed in the GenWare MP About dialog box. Start
GenWare MP and make sure it establishes communication with the generator. Press the
About button on the GenWare MP toolbar to open the About dialog box. If the Digital
Interface Options feature is enabled it will be displayed in the Features list as shown below.

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Page 2 Rev. B Manual Supplement # 903124


CPI Canada Inc. Supplement

Note: If the Digital Interface Options feature is locked on an attempt to set an interface option to any
number other than 0 the generator will leave the value of the interface option unchanged.

b) Enter the Generator programming mode (See service manual chapter 3C).
c) Select the GEN CONFIGURATION menu.
d) Enter the RECEPTOR SETUP menu.
e) Select the Receptor that is to be configured with the CPI interface option.
f) Set the INTERFACE OPTION parameter to 40.

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Manual Supplement # 903124 Rev. B Page 3


CPI Canada Inc. Supplement

THEORY OF OPERATION

3.1 GENERAL

The generator control board contains an additional set of inputs and outputs used to interface to a digital
system when interface option is set to CPI interface option 40.
The interface can be used with a Bucky enabled or disabled for a selected receptor. When a Bucky is
enabled the generator will first drive the Bucky, and only when the Bucky is ready the generator will drive
the detector (Figure 2). If the Bucky for a selected receptor is disabled, the generator will only drive the
detector (Figure 1). For Bucky interconnections and the instructions on programming a Bucky interface
refer to the CMP200 DR Service Manual.

3.2 GENERATOR OUTPUT SIGNALS

3.2.1 PREP REQUEST (EXP_PREP)

The Prep Request signal (EXP_PREP) is an output signal from the Generator to a digital system. The
Generator will activate the EXP_PREP signal when the Prep input is enabled (Prep button/switch
pushed/pressed) and it will deactivate the signal when the Prep input is disabled (Prep button/switch is
released) or if the exposure ends or an error occurs.

3.2.2 EXPOSURE REQUEST (EXP_REQ)

The Exposure Request signal (EXP_REQ) is an output signal from the Generator to a digital system. The
Generator will activate the EXP_REQ signal when the X-Ray input is enabled (X-Ray button/switch
pushed/pressed) and the Generator is ready to expose and it will deactivate the signal when the X-Ray
input is disabled (X-Ray button/switch released) or if the exposure ends or an error occurs.

3.2.3 EXPOSURE ON (EXON)

The Exposure On signal is an output signal from the Generator to a digital system. The Generator will
activate the EXON signal during the exposure phase. This signal will be inactive during all other
Generator phases.

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Page 4 Rev. B Manual Supplement # 903124


CPI Canada Inc. Supplement

3.3 GENERATOR INPUT SIGNALS

3.3.1 EXPOSURE CONFIRMATION (EXP_OK)

The Exposure Confirmation signal (EXP_OK) is an output signal from a digital system to the Generator.
For the exposure to get released, the signal must become active. If the signal does not become active
within 10 seconds from the time the generator is ready to expose, the generator will report an error 27
Receptor Timeout and return to the Standby state. If the X-Ray switch is released within the 10 seconds
that the generator is waiting for the EXP_OK signal, the software will report an error 13 Manually
Terminated Exposure. When all the conditions for the exposure release are satisfied and the input
becomes active the Generator will release the exposure using current exposure settings. If during the
exposure the signal becomes inactive the Generator will immediately terminate the exposure and report
an error 13 Manually Terminated Exposure.

3.4 INPUTS AND OUTPUTS TIME DIAGRAM

Figure 1: Time Diagram With a Bucky Disabled

Figure 2: Time Diagram With a Bucky Enabled

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Manual Supplement # 903124 Rev. B Page 5


CPI Canada Inc. Supplement

4.0 SCHEMATIC

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Page 6 Rev. B Manual Supplement # 903124


CPI Canada Inc. Supplement

(This page intentionally left blank)

Use and disclosure is subject to the restrictions on the title page of this CPI document.

Manual Supplement # 903124 Rev. B Page 7

You might also like