ACTIMED User Guide

User Guide english

Version 3
Test integration- and equipment management software

Instructions for the user: This manual is supplied in the form of a loose-leaf
publication. Any necessary updates and additional pages can thus be included
without difficulty later so that you always have the latest version of the manual. We
provide you with supplements and/or changes to your manual along with technical
modifications that are made to your testing system.

The DIN/VDE symbols and drawings shown in this manual have been reproduced
with the permission of the Deutsches Institut fr Normung e.V. (DIN: German
Institute for Standardization) and the Verband Deutscher Elektrotechniker e.V.
(VDE: German Association of Electrical Engineers). In applying the relevant
standards, it is important to use the versions with the most recent date of issue.

We reserve the right to make any changes to the information in this manual without
prior notice. S.P.L. Elektronik accepts no responsibility in this regard. The software
that comes with this product is supplied within the framework of a License
Agreement. The software may be used and copied only in accordance with the
terms and conditions of the License Agreement.

All rights pertaining to reprinting and reproducing the manual, or parts thereof, are
reserved. No part of the manual may be reproduced in any form whatsoever
(photocopy, microfilm or any other procedure) or processed, reproduced or
distributed using electronic systems without prior written approval from S.P.L.
Elektronik.

Other products mentioned by name are trademarks or registered trademarks of


their respective companies.

S.P.L. Elektronik
Forstweg 1
D - 14656 Brieselang
GERMANY

Tel. +49 33232 39915


Fax +49 33232 39911

13th. Edition, 2013

1
S.P.L. ACTIMED User Guide

1 INTRODUCTION TO ACTIMED...........................................................................4

2 INSTALLATION / SYSTEM..................................................................................5

2.1 ENTERING LICENSE INFORMATION .......................................................................7


2.2 BACKUP DATA ....................................................................................................7

3 DEVICE SUPPORT .............................................................................................8

4 HINTS FOR USE .................................................................................................9

4.1 NAVIGATION BAR ................................................................................................9


4.2 RECORD FILTER IN TABLE VIEW ........................................................................10
4.3 PULL DOWN MENUS..........................................................................................10
4.4 ONLINE HELP ...................................................................................................10
4.5 PRINT FUNCTIONALITY ......................................................................................10

5 ACTIMED STRUCTURE ....................................................................................11

6 FILE MENU........................................................................................................12

6.1 CUSTOMER TABLE ............................................................................................12


6.2 MANUFACTURER TABLE ....................................................................................13
6.3 SUPPLIER TABLE ..............................................................................................14
6.4 DEVICE TABLE ..................................................................................................15
6.5 TEST PROTOCOL TABLE ....................................................................................18
6.6 DEVICE TYPE TABLE .........................................................................................19
6.7 DEVICE KIND TABLE .........................................................................................20
6.8 ACTIVITY TABLE ...............................................................................................21
6.9 TEST INSTRUCTIONS TABLE...............................................................................22
6.10 WORK STEPS AND FUNCTIONS TABLES............................................................25
6.11 COST CENTER LIST .........................................................................................26
6.12 LOCATION TABLE............................................................................................27
6.13 RESPONSIBLE PERSON TABLE ........................................................................27
6.14 THE TESTER TABLE .........................................................................................28

7 EXTRAS MENU .................................................................................................29

2
S.P.L. ACTIMED User Guide

7.1 READ IN FUNCTIONS .........................................................................................29


7.2 DETECT DEVICES ..............................................................................................29
7.3 OPTIONS ..........................................................................................................30
7.4 REMOVE RECORD LOCKS ..................................................................................30
7.5 EXPORT ...........................................................................................................31
7.6 IMPORT ............................................................................................................32
7.7 BACKUP DATA / RESTORE DATA .......................................................................33

8 TEST PROCEDURE ..........................................................................................34

8.1 GENERAL PROCEDURE .....................................................................................34


8.2 EXAMPLE PROCEDURE ......................................................................................39

9 GLOSSARY .......................................................................................................43

3
S.P.L. ACTIMED User Guide

1 Introduction to ACTIMED

ACTIMED is a Test integration- and equipment management software, which can


be brought into action in medicine and industry. ACTIMED has been developed for
technicians, who are responsible for the realization of testing. Hospitals as well as
self-employed inspectors and companies doing tests can use ACTIMED.
ACTIMED serves for preparation, carrying out and administration of test-runs with
test instruments of S.P.L. Elekronik, GERB Elektronik GmbH and other producers.
ACTIMED has been developed by S.P.L. Elektronik and is based on GMEP and
ACTIMED 1.0.
Customer- and instrument administration as well as a programme for self-made
forms are integrated components of ACTIMED.
The stored data is presented in clearly arranged tables.
This data can be sorted and filtered in alphabetical or numbered order
(descending/ascending).
For treatment of tables an input mask in file card size is accessible.
A file card called MEMO is integrated in some dialogs. This memo can be seen as
a notebook, in which the most important information can be imported.
An ODBC-database intersection allows ACTIMED on the client-side access to
professional database servers, for example the database InterBase.

4
S.P.L. ACTIMED User Guide

2 Installation / System

An IBM-compatible personal computer and MS WINDOWS version up to W95 are


the basis for the software installation. The necessary memory is 50 MB hard disc-
memory and 16 MB main memory.
To install ACTIMED you have to call the set-up routine. By choosing the standard
installation all available device drivers become installed. Choosing the custom
installation you can select which device drivers should be installed before the
installation takes place. ConActX as a required software interface is installed in both
cases.
ConActX
ConActX Test Module
ACTIMED 3

After installing the software a restart of your Windows system is required to finish
the installation successfully.
During installation of ACTIMED under up to W-NT it can be necessary to start
ConActX and ConActX Test Module once from the current user to set user specific
entries in Windows registry.

5
S.P.L. ACTIMED User Guide

The Database used by ACTIMED is a MS-Access Database. To access the data


ODBC settings were done by the installation routine. You can change the database
by changing the corresponding ODBC settings.

Doing that you can point to equal or different physical databases with the ODBC
entries ACTIMED_3DB and CxTmDB. That gives you the opportunity to put the
database to the network and let multiple ACTIMED installations access one
database.
If you cant access the ODBC entries using the ODBC-Data source-Administrator
because they are invisible, you can configure them editing the registry directly.
On request it is possible to connect ACTIMED to different databases like Oracle or
Interbase.

6
S.P.L. ACTIMED User Guide

2.1 Entering License Information


To use ACTIMED after the initial trial period of 30 days you have to have a valid
license. The license has to be entered in the license dialog and is effective after a
restart of ACTIMED. Entering a valid license in ACTIMED you automatically get a
valid license for ConActX-Test Module and for ConActX.

2.2 Backup Data


To backup the data entered and measured with ACTIMED you have to save the file
C:\Program Files\ConActX\Data\ACTIMED3DB.mdb regularly. This file contains all
the data displayed in the data tables and ACTIMEDs dialogs.
You can use typical backup tools like the backup tool of Windows to save the
database file.

7
S.P.L. ACTIMED User Guide

3 Device Support

ConActX supports the following Safety Tester and Functionality Tester.

Device description

GM-50 Measurement- and test device for electrical safety


GM-100 Measurement- and test device for electrical safety in accordance to FW Version 3
IEC 60601, IEC 62353, IEC 61010, VDE 0701-0702, BGV A3
GM-150 Measurement- and test device for electrical safety in accordance to FW Version 3
IEC 60601, IEC 62353, IEC 61010, VDE 0701-0702, BGV A3
GM-200 Measurement- and test device for electrical safety in accordance to FW Version 3
IEC 60601, IEC 62353, IEC 61010, VDE 0701-0702, BGV A3
GM-300 Measurement- and test device for electrical safety in accordance to FW Version 3
IEC 60601, IEC 62353, IEC 61010, VDE 0701-0702, BGV A3
GM-410 Measurement- and test device for electrical safety in accordance to FW Version 3
IEC 60601, IEC 62353, IEC 61010, VDE 0701-0702, BGV A3
GM-610 Measurement- and test device for electrical safety in accordance to
IEC 60601, IEC 62353, IEC 61010, VDE 0701-0702, BGV A3

SAFE Measurement- and test device for electrical safety in accordance to FW Version 3
IEC 62353, IEC 61010, VDE 0701-0702, BGV A3

GS-X Basic unit FW Version 3

DEFI GS Test Module for functional testing of external defibrillators

STIMU GS Test Module for functional testing of electro stimulation equipment


INFU GS Test Module for functional testing of infusion pumps

PACE GS Test Module for functional testing of external pacemakers

RF GS Test Module for functional testing of high-frequency surgical devices


HF GS Test Module for functional testing of high-frequency surgical devices
ES GS Test Module for functional testing of ECG equipment

HF-150 Test device for functional testing of high-frequency surgical devices

HF-300 Test device for functional testing of high-frequency surgical devices

HF-400 Test device for functional testing of high-frequency surgical devices

DC-50 S Test device for functional testing of external defibrillators

DP-300 Test device for functional testing of external defibrillators

ES-300 Test device for functional testing of ECG equipment

IN-300 Test device for functional testing of infusion pumps

IN-400 Test device for functional testing of infusion pumps

other

8
S.P.L. ACTIMED User Guide

4 Hints for use

You can use ACTIMED with or without a mouse without restrictions. Using the
keyboard is very useful when you have to test in very cramped spaces and there is
not enough space to use a mouse. Editing the entries in a table can be
accomplished by pressing F2.

4.1 Navigation bar

You can find the navigation bar on top of each table. The navigation bar is used to
browse through the records. Therefore you can use the following functionalities:
First Record:
The record selector is set to point to the first record of the table.
Next Record:
Increments the record selector to point to the next record of the table.
Previous Record:
Decrement the record selector to point to the previous record of the table.
Last Record:
The record selector is set to point to the last record of the table.
Insert Record:
Adds a new record to the table. The record selector points to the new record after it
is added.
Delete Record:
Removes the current record from the table.
Edit Record:
A dialog is being opened and displays the current record. You can now change the
entries of the record in this dialog.
Duplicate Record:
The current record will be duplicated and opened in edit mode. You can now
change the new record.

9
S.P.L. ACTIMED User Guide

4.2 Record Filter in Table View

A powerful tool when viewing data in table view is the record filter. You can use it to
display only records of interest. You can rapidly reduce the number of visible
records by defining conditions used to filter records. To select a record you will have
to place the record selector to any field in the record line.
The comparison operator is used to specify the conditions for filtering records. The
filter options are applied to the fields you specify. Depending on whether a condition
is true or not a record will be displayed or not.
The following list shows all valid comparison operators:

= equal
<> not equal
> greater than
>= greater than or equal
< lower than
<= lower than or equal

4.3 Pull Down Menus

You can select pull-down menu entries to display tables or dialogs. Bold menu
entries are the default entries and you can select them by double-clicking the parent
menu.

4.4 Online Help

ACTIMED contains a comprehensive online help system based on HTML. You can
access it by clicking the function key F1.

4.5 Print Functionality

ACTIMED supports comprehensive print out functionalities. You can access them
from the table views. You can edit the sample reports that come with ACTIMED or
you can define your own reports. For further descriptions of print out functionalities
please refer to the online help or to the document designer.pdf.

10
S.P.L. ACTIMED User Guide

5 ACTIMED Structure

This figure shows the schema of ACTIMED data. All test and administrator
functionalities are based on this schema.

11
S.P.L. ACTIMED User Guide

Main menu of ACTIMED 3.

6 File menu

6.1 Customer Table

The customer table displays all data of customers stored in ACTIMED. You can get
an overview of all customers you attend to in a comfortable table view. The
customer name and the shorthand are mandatory fields.

12
S.P.L. ACTIMED User Guide

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be edited or printed. All print out
functionalities provided by the integrated form generator can be used for customer
data.

6.2 Manufacturer Table

The manufacturer table show all manufacturers stored in ACTIMED. You can get an
overview of all manufacturers of the stored devices. Mandatory fields are Name1
and the shorthand.

13
S.P.L. ACTIMED User Guide

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be edited or printed. All print out
functionalities provided by the integrated form generator can be used for
manufacturer data.

6.3 Supplier Table

The supplier table show all suppliers stored in ACTIMED. You can get an overview
of all suppliers of the stored devices (equipment). Mandatory fields are Name1 and
the shorthand.

14
S.P.L. ACTIMED User Guide

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be edited or printed. All print out
functionalities provided by the integrated form generator can be used for
manufacturer data.

6.4 Device Table

The device table show all devices of all customers stored in ACTIMED. You can get
an overview of all suppliers of the stored devices (equipment). The inventory
number is a mandatory field. You can enter all field values directly or by clicking the
corresponding button to select a value out of a value list.

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be edited or printed. All print out
functionalities provided by the integrated form generator can be used for device
data.

15
S.P.L. ACTIMED User Guide

16
S.P.L. ACTIMED User Guide

17
S.P.L. ACTIMED User Guide

6.5 Test Protocol Table

The Test Protocol List show all test protocols of all devices of all customers stored
in ACTIMED. You can get an overview of all test protocol available.

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be viewed or printed. All print out
functionalities provided by the integrated form generator can be used for test
protocol data.

18
S.P.L. ACTIMED User Guide

6.6 Device Type Table

The Device Type table show all device types stored in ACTIMED. You can get an
overview of all device types available. The Device Type, Device Kind, Model and the
Manufacturer are mandatory fields. You can enter all field values directly or by
clicking the corresponding button to select a value out of a value list.

The Device Type describes a group of devices manufactured by a special producer


and containing a unique Type Identifier. The Device Type is a generic term that
describes all devices that are similarly built and only differ in their Serial Number.

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be viewed or printed. All print out
functionalities provided by the integrated form generator can be used for Device
Type data.

19
S.P.L. ACTIMED User Guide

6.7 Device Kind Table

The Device Kind table show all device kinds stored in ACTIMED. You can get an
overview of all device kinds available. The Device Kind, Model and the Manufacturer
are mandatory fields. You can enter all field values directly or by clicking the
corresponding button to select a value out of a value list.

The Device Kind is a global grouping of medical devices. All medical devices can be
roughly grouped in generic classes with similar properties. The device kind
defibrillator describes all defibrillators independent from their construction and
manufacturer.

20
S.P.L. ACTIMED User Guide

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be viewed or printed. All print out
functionalities provided by the integrated form generator can be used for device
kind data.

6.8 Activity Table

The Activity table show all activities stored in ACTIMED. You can get an overview of
all activities available. The Activity and Test Instruction are mandatory fields. You
can enter all field values directly or by clicking the corresponding button to select a
value out of a value list.

The activity consists of a name, a selected test instruction and an interval. The
interval specifies the time between two tests. After that interval the test for the
device should be repeated. The intervals unit is month.

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be viewed or printed. All print out
functionalities provided by the integrated form generator can be used for activity
data.

21
S.P.L. ACTIMED User Guide

6.9 Test Instructions Table

The Test Instructions Table show all test instructions. You can get an overview of all
test instructions available. The Test Instruction is the mandatory field.
The Activity consists of a list of Test Steps. These test steps are Work steps with a
corresponding Assessment and the necessary Limits. Work steps are sets of
functions and parameters.
A Test Instruction describes how a test is performed. It consists of a list of Test
Steps that are worked through step by step. This work through is usually performed
automatically. Additionally the user has the opportunity to set a stop mark S to
stop the automatic processing at the marked (stop mark) Test Step. Besides the
stop mark S there exists another mark labelled M that indicates that this Test
Step should be entered manually. The automatic execution stops, a dialog is
displayed and the user gets the chance to enter the measured value manually
independent from the connected Safety or Functional Tester.
Test Instructions can separately be imported and exported. This functionality is
available in the corresponding menu.
The ACTIMED 3 Test Instructions are identically to the Test Instructions found in
ConActX Test Module.

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be viewed or printed. All print out
functionalities provided by the integrated form generator can be used for test
instructions data.

22
S.P.L. ACTIMED User Guide

Structure in principle of a test instruction

23
S.P.L. ACTIMED User Guide

Example test instruction

24
S.P.L. ACTIMED User Guide

6.10 Work Steps and Functions Tables

The Work Steps tables show all Work Steps. You can get an overview of all work
steps available. The Functions tables show all Functions. You can get an overview
of all functions available.

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be viewed or printed. All print out
functionalities provided by the integrated form generator can be used for work steps
and functions data.

Functions Electrical Safety

This table displays all functions for testing the electrical safety provided by test
devices. Only using new device drivers can extend this list.

Functions Functionality Tester

This table displays all functions for testing the functionality provided by test devices.
Only using new device drivers can extend this list.

Functions User defined Steps

This table displays all functions for entering values manually. You can extend the list
and add as many new functions as you like. The result usually specifies a physical
value.

Functions Visual Inspection

This table displays all functions for visual inspection. You can extend the list and
add as many new functions as you like. The result is intended to be a boolean
(yes/no) value.

Work Steps Electrical Safety

This table show all work steps available for testing the electrical safety. You can
extend this list, as you like.

Work Steps Functionality Steps

This table show all work steps available for testing the functionality. You can extend
this list, as you like.

25
S.P.L. ACTIMED User Guide

Work Steps User defined Steps

This table show all work steps available for entering values manually. You can
extend this list, as you like.

Work Steps Visual Inspection

This table show all work steps available for visual inspections. You can extend this
list, as you like.

6.11 Cost Center List

The Cost Center table show all cost center data stored in ACTIMED. You can get
an overview of all cost centers available. The Cost Center is a mandatory field.

26
S.P.L. ACTIMED User Guide

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be viewed or printed. All print out
functionalities provided by the integrated form generator can be used for cost center
data.

6.12 Location Table

The Location table show all location data stored in ACTIMED. You can get an
overview of all locations available. The Location is a mandatory field.

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be viewed or printed. All print out
functionalities provided by the integrated form generator can be used for cost center
data.

6.13 Responsible Person Table

The Responsible Person table show all responsible persons data stored in
ACTIMED. You can get an overview of all responsible persons available. The
Resp.Pers.Name is a mandatory field.

27
S.P.L. ACTIMED User Guide

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be viewed or printed. All print out
functionalities provided by the integrated form generator can be used for Location
data.

6.14 The Tester table

The Tester table show all tester data stored in ACTIMED. You can get an overview
of all testers available. The Tester Name is a mandatory field.

Remark: Changing the tester name does not affect the test protocols tested by the
Tester.

Using the record filter and sort functionalities you can select single records or
groups of records. The displayed records can be viewed or printed. All print out
functionalities provided by the integrated form generator can be used for location
data.

28
S.P.L. ACTIMED User Guide

7 Extras Menu

7.1 Read In Functions

After installing a new previously unknown test device and its corresponding device
driver the new functions are learned from ACTIMED 3 by reading them in.

7.2 Detect Devices

To detect the devices connected with your computer. You have to call the device
detection. Only detected devices can be used to perform Test Instructions.

29
S.P.L. ACTIMED User Guide

7.3 Options
This dialog is used to enter license information.
Additionally you can specify the behaviour of ACTIMED 3. This setting is important
for communication between several ACTIMED 3 installations and specifies whether
the installation should behave like a client or a server installation. For further
information please refer to the import / export section of this documentation.
You can specify if Device Detection should be performed automatically starting
ACTIMED 3.

7.4 Remove Record Locks

It is possible to use ACTIMED 3 in a local area network. This means that used
records must me locked to protect them against multiple use. If a record is not
closed correctly the record is still locked. This means it cannot be used before it is
unlocked. The Remove Record Locks functionality is a possible way to unlock
accidentally locked records that were not unlocked automatically.

30
S.P.L. ACTIMED User Guide

7.5 Export

The Export is used for output of all data base information of ACTIMED 3. Selecting
an output specification file with the dee file extension you can specify the export
format.
Export data using the Extras / Export is a full data export. That means that all
devices available are exported in the format described by the dee file. It is not
possible to select sub data sets using selection criterions.

To export only those devices that fulfil a special criterion you can start the export
from the Device List by choosing Data Exchange Export. There the export format is
automatically detected using the Client/Server setting from the Extras / Options
menu.

31
S.P.L. ACTIMED User Guide

The server creates a file that can only be read from the client. The client creates a
file that can only be read by the server performing data import.
This Client / Server functionality enables testing organizations or workgroups to use
a central server installation and administer and evaluate device data.
Device data for devices to be tested can be exported from such a central
installation. The export file can then be imported into a client installation, e.g. on a
notebook. The clients now are used to test the devices following the Test
Instructions. After testing the Test Protocols can be transferred to the server
installation using import and export functionalities. Changes regarding the device
data should be done using the server installation.
The export and import dont contain functionalities for importing or exporting Test
Instructions. This has to be performed separately using the menu in the Test
Instruction list.

7.6 Import

The import is used to read in data base information into ACTIMED 3 stored in files.
The import format is stored in an import specification file with the dei file
extension.

32
S.P.L. ACTIMED User Guide

Device data that was exported using the client server functionality can be imported
from the Device List Data Exchange Import menu option.

The import format is selected automatically depending on the Client / Server setting
done in the Extras Option menu.
After importing device data you have to re-open the device table to see the imported
devices.
Optionally it is possible to remove all device data stored in the database before the
import takes place. That is useful if a client should only contain data of devices that
actually have to be tested and therefore have to be imported before. You delete the
devices by checking the Delete all device data before import option.

7.7 Backup Data / Restore Data

This option is useful to easily backup and restore all ACTIMED 3 data.
Test Instructions and dialog settings are not saved using this functionality. Please
do a complete security copy of the ACTIMED program and database to save all
data regularly.

33
S.P.L. ACTIMED User Guide

8 Test Procedure

The following example describes the test procedure to perform a test correctly.

8.1 General Procedure

To prepare a test please be sure that the device you want to test exists in your
ACTIMED 3 database.

Otherwise you have to add a new device using the:

File Device
menu. There you can see the device table that displays all devices independent
from the customer they belong to. Using the device dialog you can add the device
data.

34
S.P.L. ACTIMED User Guide

35
S.P.L. ACTIMED User Guide

Activities
Using the Activities tab in the Add/Edit Device dialog you have to add at least one
activity. This Activity specifies a Test Instruction and an Interval. You can add as
many activities as you like for each device. You could add an activity to test the
electrical safety selecting a corresponding test instruction and an interval of 12
months. Additionally you could add a functionality test that tests the internal
accumulators all 24 months. A visual test that is repeated after 2 months could be
also added.

36
S.P.L. ACTIMED User Guide

By pressing the:

Test
button you start the Test Instruction.

You can start the automatic test pressing the Start F9 button. Double clicking a
test step opens the circuit analyzer dialog to test the test step separately.
To get assessment information select the Assessment button. To detect devices
you can click the Detect Devices button.
It is not possible to change the order or the assessment in this dialog. To do that
you have to open the Test Instruction in the Test Instruction dialog out of the file
menu.

37
S.P.L. ACTIMED User Guide

After the test was performed you can change the values in the Report tab of the
Test Protocol.

The Test Protocol Number is generated in ACTIMED 3 and should not be altered.
The Test Date is set to the actual system date.
The Tester Name is set by ACTIMED 3 and should not be altered.
The Test Result has to be set manually.
The General Judgement is a piece of additional information that is not evaluated by
ACTIMED 3.
Pressing the Print button prints a basic report on Windows standard printer.
Pressing the Export button an export different to the standard export of the Test
Protocol is performed. Please note that this export is not repeatable.
Pressing OK F10 the Test Protocol is saved in ACTIMED. You can now view and
print the test protocol as often as you like. The reports can be found at the Test
Protocols tab in the Device dialog and in the Test Protocols List.

38
S.P.L. ACTIMED User Guide

The following example describes the test procedure to perform a test correctly.

8.2 Example Procedure

These hints shall give an example for an course of a test under application and
processing of a Test Instruction with ACTIMED 3 / ConActX Test Module.
This example does not lay a claim on completeness and correctness

Used inspection devices:

GS-X SAFE, GS-X DEFI, circuit analyser with battery test function

Visual inspection

At first a visual check is accomplished at the test object. This visual check has e.g.
following points:

- are there existing damages on the housing?


- is the equipment in a good general state?
- are there existing damages to the lines?
- are serial number, inventory number, etc. correct?

For all steps specified here only one function is produced, which expects as positive
result "Yes". Then the appropriate work steps are generated from this function.
These work steps then are assigned to the appropriate test instruction plus the
evaluation.

e.g.
function: View 1 - visual inspection
work step: visual inspection of housing
evaluation: yes, absolute tolerance, target value 1, deviation 0

Electrical safety

The examination of electrical safety takes place in accordance to the demanded


points of IEC 60601, IEC 62353, VDE 0701 or VDE 0702. Additionally, depending
on used testing unit, the following further measurements can be accomplished.

- insulation resistance test


- line voltage measurement
- measurement of the current output
- measurement of the power output

39
S.P.L. ACTIMED User Guide

e.g.
function: SPL20
work step: measurement of protective ground resistance
parameter: cyclically
evaluation: 0 - 0,5 ohms

For all steps specified here the functions SPL01 - SPL44 are deposited, which
expects a measured value in the appropriate unit as result. Then from these
functions corresponding work steps under allocation of the appropriate parameters
(set up time, measuring time) were generated. These work steps then become in
the appropriate test instruction with the allocation of an evaluation a test step.

Function test

The function test takes place according to the manufacturer data.

- defibrillation asynchronously
- defibrillation paddle synchronously
- defibrillation monitor synchronously

For all steps specified here the function xxx DEFI2 is deposited, which expects as
result one or more single measured values in the appropriate unit. Then from these
functions corresponding work steps under allocation of the appropriate parameters
(measuring range, pulse form, pulse amplitude, pulse time) are generated. These
work steps then become in the appropriate test instruction with the allocation of an
evaluation a test step.

e.g.
function: SPL215 DEFI 2
work step: defibrillation asynchronously
parameters: measuring range 2
evaluation: energy; 350 - 370 Joule
maximum voltage: 2900 - 3100 V
pulse time: 2 - 4 ms

User-defined examinations

The user-defined examination can be e.g. the test of an accumulator with an


appropriate battery test equipment.

- battery test

For this step at first a function with appropriate unit is put on. From this function a
work step is generated. This work step is then assigned to the appropriate test
instruction plus the evaluation.

40
S.P.L. ACTIMED User Guide

Test procedure
The test instruction is selected. Then this test instruction is started in automatic
operation.

At first the visually inspecting steps are done. If the result is positive the next test
step is called automatically.

Then the test steps of electrical safety are processed. These obtain a measured
value as result. If the result is positive the next test step is called automatically.

Following the function test steps are called. These have usually one or more
measured values as result e.g..

- energy in Joule (e.g. 360)


- voltage in Volt (e.g. 3000)
- pulse length in ms (e.g. 3)
- pulse duration in ms (e.g. 10)

41
S.P.L. ACTIMED User Guide

These results are indicated in a result dialogue. Furthermore this dialogue offers the
possibility (depending on possibly function) to show a taken up measured value row
(e.g. 1024 measured values) in form of a curve. This curve is then shown with the
help of a special diagram. There this curve can be measured or printed out.
The taken up measured value row and / or the curve resulting from it, cannot be
stored at the moment. This applies in principle also to all other functions with curve
displaying. With none use of the curve displaying the results of the measurements
are transferred automatically after an adjustable time to the test instruction. The
single results of measurement are evaluated in the test provision singly.
At the end of the measurement the user-defined test is called and the value
determined with the circuit analyser is registered. The result of the measurement is
evaluated in the test instruction.

Now the test report in the window test report 2 can be classified depending on
examination.

Now the test report with "Print" can be printed out, be exported with "Export" into
any file or be spent with OK to ACTIMED 3.

42
S.P.L. ACTIMED User Guide

9 Glossary

Work Step
A Work Step is a Function from the Function Table with additional Assessment
information. Work Steps are stored in the Work Step table in the database.
Work steps are no direct ancestors of test steps
Example
Work step earth connector resistance
measuring time
Device
A device in ACTIMED describes a device to be tested. It has an unique Inventory
Number. Each device belongs to a Device Type. Additional information e.g. the date
of the next STK can be provided. A device is tested using an assigned Test
Instruction. Results then are stored in Test Protocols that are assigned to the
device.

Device Type
Devices are assigned to a Device Type. A Device Type describes all properties that
are relevant for a Test. Those are e.g. Manufacturer, Protective Class, Med. GV
Group, Rated Voltage, etc. The Device Types are stored in the Device Type table in
the database.

Function
The Function Table shows all Functions that can be used for testing in ACTIMED.
Functions are grouped into the following sections: Visual Test functions, Electrical
Safety functions, Functionality Test functions and User Defined functions.

Manufacturer
The Manufacturer Table stores all relevant data of manufacturers. Those are the
manufacturers name, address, contact person, etc. The Device Type points to a
manufacturer.

43
S.P.L. ACTIMED User Guide

Memo
You can make notes using the memo that is directly assigned to a data record. You
can edit memos. Memos that are assigned to Functionality Test functions or to
Safety Test functions are not editable. Memos are very useful data items because
they are provided for nearly each record type.

Test Instruction
A Test Instruction describes an ordered test procedure to test a device. A Test
Instruction consists of several Work Steps that specify the action to take and the
necessary assessment. Test Instructions in ACTIMED can be executed
automatically or manually.

44