Professional Documents
Culture Documents
Trial and error method is a tuning method that requires the system to be online.
This method begins by setting and values to zero. Only proportional gain, is
set and changed by increasing or decreasing the value until an acceptable oscillatory
output is produced. A small overshoot and then obtains a well damped response is
considered as a good oscillation output. As the flow process control unit used in this
experiment has a similar and values, the and values must be set
simultaneously. Besides that, increasing or decreasing the values of and can be
set and done by using trial and error method to obtain a faster stability .
There are several precaution steps and recommendations to improve the results
for this experiment. First of all, the pressure value should not be adjusted more than
0.4MPa to the electro-pneumatic proportional valve and the system tanks. Next,
manual valves should be in correct operation whether open or close. This is to ensure
that more accurate result can be obtained. The reading of the manual valve and
pressure valve must in range so that there will no error occurred during the
experiment. When changing a new value or value, the time interval should be
as long as possible in order to get an accurate stable system. Lastly, the setting of the
sensor and the control valve should not be changed because it may disturb both sensor
and valve because it may affect the performance and operation range.