You are on page 1of 9

Int. J. Ion Mobil. Spec.

(2008) 11:61–69
DOI 10.1007/s12127-008-0006-5

ORIGINAL RESEARCH

A rapid analytical method for hair analysis using ambient


pressure ion mobility mass spectrometry with electrospray
ionization (ESI-IMMS)
Prabha Dwivedi & Herbert H. Hill Jr

Received: 13 September 2007 / Revised: 8 May 2008 / Accepted: 9 May 2008 / Published online: 10 June 2008
# Springer-Verlag 2008

Abstract Analysis of hair is often applied to assess drug Keywords Hair analysis . Ion mobility spectrometry-mass
abuse history, exposure to environmental and industrial spectrometry . Metabolomics . Drugs . Cosmetics .
pollutants, heavy metals, gestational drug exposure and Pharmaceuticals
various other screening purposes. This manuscript reports
the application of ambient pressure ion mobility spectrom-
etry mass spectrometry (IMMS) with electrospray ionization Introduction
(ESI) source as a rapid analytical tool for hair analysis. The
study demonstrated that ion mobility spectrometry (IMS) as a Analysis of hair is used in various toxicological, clinical, or
pre-separation technique prior to analysis by mass spectrom- forensic investigations. The advantages of performing hair
etry (MS) provides detection and determination of com- analysis over other biological samples are that hair analysis
pounds of interest present in hair at nano-molar concentration (1) is a less intrusive method, (2) provides a history of drug
level. After extraction of analytes from hair, the ESI-IMMS use, (3) is more resistant to adulteration, (4) can be easily
method of analysis does not require the derivatization or collected and stored, (5) has long shelf life and (6) can be
sample treatment that is often required for other separation easily decontaminated [1]. Determination of cause of death,
methods such as gas chromatography. One advantage of IMS or assessment of drug abuse and exposure levels to
over chromatography separation is that resolving powers are environmental and industrial pollutants [2–4] or monitoring
similar to those in GC and much greater than those possible therapeutic compliance and treatment [5, 6] are achieved by
by liquid chromatography. In addition, separation speed is performing analysis of hair samples. Identification, detec-
faster than both gas and liquid chromatographic methods. tion and/or determination of metabolites present in hair can
Four of the nine hair samples anonymously donated by also be used to predict, detect, and treat various diseases
customers at a local hair salon tested positive for caffeine and [7–14]. In addition, segmental analysis of hair yields
two of the four samples that tested positive for caffeine also information that can be related or associated with chronol-
tested positive for nicotine. A positive response based on ogy of diseases, poisonings, or drug use [15, 16].
mass analysis for methamphetamine was obtained for one of The use of hair for forensic analysis is not a new concept.
the hair samples. Further investigation using the mobility In 1858, hair obtained from a dead body, exhumed after
data demonstrated that the response was a false positive and 11 years was analyzed for arsenic and cause of death by
that it may have occurred from the use of a hair gel. This arsenic poisoning was reported [17]. Hair analysis as a
article reports the potential of IMMS as an analytical means to determine histories of drug abuse gained popular-
technique for rapid and routine screening of hair samples, ity only in 1970s and 1980s when methods to detect opiates
cosmetics, and pharmaceuticals. in digested hair samples were reported [18, 19].
Currently, analysis of hair to determine drugs of abuse is
performed by separation techniques such as gas chromatog-
P. Dwivedi : H. H. Hill Jr (*)
raphy [20, 21], liquid chromatography [22–24] and capillary
Department of Chemistry, Washington State University,
Pullman, WA 99164, USA electrophoresis [25–27] with or without mass spectrometry.
e-mail: hhhill@wsu.edu However, the steps required for these analyses such as
62 Int. J. Ion Mobil. Spec. (2008) 11:61–69

extraction, concentration, derivatization, separation and coupled to ESI as ion source and mass spectrometry (ESI-
detection, of targeted analytes reduce sample throughput. IMMS) can be applied for rapid analysis of hair samples.
Derivatization procedures are generally applicable to As a “proof of concept study” the study focused on
certain chemical classes and often render chromatographic determination of caffeine and nicotine present in hair
methods suitable only for a specific class of analytes. For samples that were obtained as donations from customers
example, preparation of hair sample for drug analysis at a local hair salon. These two drugs were selected as the
by gas chromatography-mass spectrometry (GC-MS) target compounds for this study because caffeine and
involved acid hydrolysis of hair, followed by a three-step nicotine are the most common drugs that can be found in
liquid–liquid extraction, and derivatization with N,O-bis hair samples from the general population.
(trimethylsilyl) trifluoroacetamide plus 1% trimethylchlor-
osilane [28]. Similarly, for sensitive determination of
histamine HA in hair by high pressure liquid chromatography– Materials and methods
electrospray ionization mass spectrometry (HPLC–ESI-MS),
HA was first extracted from hair by acid hydrolysis, followed Samples and solvents
by fluorescence labeling for HPLC separation and then
detected by ESI-MS [29]. Thus, there is a need for develop- High performance liquid chromatography grade solvents
ment of innovative approaches for hair analysis in order to (methanol, water and acetic acid) were purchased from J. T.
simplify the measurement, increase sample throughput and Baker (Phillipsburg, NJ, USA). Caffeine and nicotine were
minimize diversity in sample preparation. purchased from Sigma-Aldrich (Sigma Aldrich Chemical
Ion mobility spectrometry (IMS) with its millisecond Co., St. Louis, MO, USA). Hair samples were anonymous-
separation time provides 105 times greater number of plates ly donated by customers at a local hair salon.
per second when compared to chromatographic separation
methods such as GC, LC, and CE [30]. Currently, IMS is Sample preparation
widely used as a rapid detection method for vapor phase
analytes such as chemical warfare agents, explosives, drugs Hair samples collected at a local hair salon from nine
and air contaminants [31]. Pyrolysis-IMS and thermal customers were cut into about 5 mm lengths, washed three
desorption-IMS have also been applied for analysis of hair times with 50:50 methanol water solution, and left to dry
for drugs [32–35]. However, determination of non-volatiles overnight at room temperature. After the sample was
present in hair is not feasible using pyrolysis-IMS and evaporated to dryness, 50 mg of each hair sample was
thermal desorption-IMS. With the introduction and devel- extracted with 4 ml of methanol in a screw cap culture tube
opment of electrospray ionization (ESI) as the ion source by heating in a water bath for 8 h at ∼80 °C and then cooled
for IMS [36, 37], application of IMS have been extended to to room temperature. The extract were then filtered using a
the analysis of biological samples and structural studies of 0.2 μm size syringe filter, and diluted with HPLC grade
non-volatile ionic species [38–41]. Thus, ESI-IMMS seems water and acetic acid to make solutions of composition
to be viable analytical option that can be used for rapid 47.5:47.5:5 (methanol–hair extract/water/acetic acid). The
determination of volatile and non-volatile species present in extract was diluted to approximately half its original
hair. Two dimensional information (mobility and m/z) concentration (110% dilution). Samples were stored at
acquired through an IMMS experiment allows the analysis −4 °C until analysis.
of complex biological samples. In a typical IMMS
experiment, ions produced at atmospheric pressure by an ESI-IMMS analyzer
ion source are injected in the drift region of an IMS where
ions are separated based on their size-to-charge ratio. These The ESI-IMMS analyzer comprised of an electrospray
ions are then introduced into the MS through a pin-hole ionization source (ESI), an ion mobility analyzer (IMS)
leak, guided by einzel lenses to the mass analyzer at and a quadrupole mass analyzer (MS), the schematic of
pressure of ∼10−5 torr, and are separated based on their which is shown in Fig. 1.
mass-to-charge ratio. Due to high velocity of ions in
vacuum compared to that at ambient pressures, time spent
by an ion in the mass analyzer is considered negligible with ESI Source For the generation of electrospray ions, sample
respect to the time that the ion spends in the drift region of was infused into a 30 cm long, 50 μm inner diameter silica
the ambient pressure IMS. capillary by a KD Scientific 210 syringe pump (New Hope, PA,
Hair, as any other biological sample, is expected to be USA) at a flow rate of 5 μl/min. The capillary was then inserted
complex in nature with multiple constituents. The objective into a water-cooled 22-gauge stainless steel needle, the tip of
of this study was to investigate if ion mobility spectrometry which was centered ∼1.0 cm from the target screen of the IMS
Int. J. Ion Mobil. Spec. (2008) 11:61–69 63

Fig. 1 Schematics of the ion


mobility mass spectrometer with
electrospray ionization source.
The mass spectrometer was a
quadrupole mass analyzer with
electron multiplier as detector.
The ESI and the IMS were
operated at ambient pressure and
the MS was at a pressure of
10−5 torr. The IMS and MS were
coupled through a 40 μm pin-
hole interface

analyzer and the capillary protruded ∼2 mm out of the stainless (Keithley Instruments, Cleveland, OH, USA). The ampli-
steel needle tip. The ESI needle was maintained at a potential of fied signal was then sent to either the MS data acquisition
+3.0 kV with respect to that on the target screen of the IMS. system or IMS data acquisition system.

IMS analyzer The IMS analyzer was constructed at Instrument control, data acquisition and processing Merlin
Washington State University, and used a stacked-ring software (ABB Extrel, Pittsburgh, PA, USA) was utilized for
design that has been described in previous publications all MS data acquisition, analysis and control. For the IMS
[30]. In summary, the IMS tube was divided into a gating and data acquisition, the electronic controls were built
desolvation region (7.2 cm in length) and a drift region at Washington State University. Labview (National Instru-
(21.8 cm in length) separated by a Bradbury–Nielsen-style ments, Austin, TX, USA) based software developed at
ion gate. Both regions consisted of alternating alumina Washington State University, Pullman WA, USA; was used
spacers and stainless steel rings with high temperature for IMS data acquisition and IMS gate-control. Two types of
resistors connecting the stainless steel rings (Caddock ion mobility spectrum were generated through ESI-IMMS
Electronics, Riverside, CA, USA; 500 kΩ resisters for the experiments: non selected ion mobility (NSIM) spectrum
desolvation region, 1 MΩ resisters for the drift region, and selected ion mobility (SIM) spectrum. In the first case,
250 °C, 0.1% tolerance). The temperature of the drift the MS was operated in the “scan mode” with the DC voltage
region, desolvation region, and the drift gas was maintained turned off and served as an ion transfer line between the IMS
at 250 °C, the ion gate held at a potential of 7.51 kV, and and the detector and provided a non-selected ion mobility
the instrument was operated at ambient pressure (696 to (NSIM) spectrum for the analyte(s). In the second case, the
703 torr in Pullman, WA, USA). With a potential of 9.5 kV MS was operated in the “single ion monitoring mode” and
at the IMS target screen, electric field in the desolvation served as a filter that allowed transfer of ions with a
region was ∼276 V/cm and ∼344 V/cm in the drift region of preselected m/z value reach the detector. Under these
the IMS (E/N between 2–3 Td). Preheated counter flowing conditions, a selected ion mobility spectrum (SIM) was
drift gas (nitrogen) at a flow rate of ∼1,300 ml/min was obtained. Operation of IMMS in SIM mode allows one to
introduced at the end of the drift region (preheated to the monitor the drift time(s) of ions associated with preselected
temperature of the IMS). The ions were injected into the m/z values.
IMS drift region in pulses of 0.2 ms using an ion gate
operating at a frequency of 25 Hz.
Calculations
MS analyzer The IMS analyzer was interfaced to a model
150-QC ABB Extrel (Pittsburgh, PA, USA) quadrupole Reduced mobility value for ions (Ko in the units of square
analyzer (m/z range of 0–4,000 amu) via a 40-μm pinhole centimeter per volt second) was calculated using the
interface. Earlier publications from our research group following equation:
document the detailed description and schematics of the  2   
IMMS analyzer [30, 42]. The output signal from the L 273:15 P
K0 ¼
multiplier was amplified by a Keithley model 427 amplifier Vtd T 760
64 Int. J. Ion Mobil. Spec. (2008) 11:61–69

where, L is the length of the drift region in centimeter, V is of ions detected between hair samples A and E. Absence of
the voltage applied to the ion-gate in volts, td is the ion drift IMS peaks in the drift time range of 15–30 ms in the NSIM
time in seconds, T is the drift region temperature in Kelvin, spectrum of the control and their presence in the NSIM
and P is the ambient pressure in torr. spectra of hair extracts shows that various analytes were
extracted from the hair samples and were detected by ESI-
IMS. Incomplete resolution of IMS response peaks
Results and discussion corresponding to 30–55 m/z ions that were detected by MS
resulted in a broad “hump” like peak shape in the NSIM
Analysis of hair samples spectra of each hair sample. Selected ion mobility (SIM)
spectrum provided specific details (m/z, mobility, and
Non selective ion mobility (NSIM) spectra of hair samples intensity) for each of the detected analytes. SIM spectra of
identified as A, B, C, D, E, F, and G were acquired by few target analytes detected in the extracts of hair samples
operating the MS in the “scan mode”. Figure 2 shows the are discussed in the following sections. Ability to rapidly
superimposed NSIM spectra of (1) negative control sample analyze hair samples without any further sample treatment
(m/z 10–400 Da), (2) hair sample A (m/z 100–400 Da, and after extraction, allows one to use ESI-IMS as a screening
(3) hair sample E (m/z 100–400 Da). Since most of the ions tool for rapid analysis of hair samples. ESI-IMMS can be
detected in the negative control sample were of m/z values used to rapidly characterize complex biological samples such
less than 100 Da, the cutoff m/z value for NSIM spectra of as hair and determine the identity of each constituent using
hair samples was set at 100 Da to exclude background ion two dimensional IMMS data.
peaks from the NSIM spectra. Depending on the identity of
the hair sample, number of ions detected in the m/z range of Targeted component analysis of hair samples
100–400 Da varied between 30 and 55. Peak patterns
(NSIM spectra) and number of peaks detected were Since the hair samples collected were anonymous and thus
reproducible for each hair sample in two replicate measure- provided no prior information about the subjects, the
ments. Qualitative differences of the hair samples were experiments were directed towards detection of most
observed by comparing their respective NSIM spectra. For common drugs that can be potentially present in hair samples
example, Fig. 2 shows the difference in peak pattern/number

1.40
Negative control (e)
Hair Sample A Hair sample A
Hair Sample E
1.15

(d)
IMS Response (nA)

IMS Response

0.90 Hair sample E

0.65 (c)
Hair Sample B
Negative
0.40 A E control

(b)
Hair sample G
0.15

(a)
-0.10 Negative control
0 10 20 30 40 50
Drift time (ms)
Fig. 2 Non selective ion mobility (NSIM) spectra of negative control, 0 10 20 30 40 50
hair sample A and hair sample E. Each of the hair samples produced Drift Time (ms)
characteristic and reproducible IMS spectrum as illustrated for hair Fig. 3 Selected ion mobility (SIM) spectra obtained at m/z 195 of (a)
samples A and E. The IMS spectra were obtained in the scanning negative control, (b) hair sample G, (c) hair sample B, (d) hair sample
mode of the MS. The negative control NSIM spectrum was obtained E, and (e) hair sample A. The drift time of ion at m/z 195, the (MH)+
in the mass range of 10–400 Da and the NSIM spectra of hair samples ion of caffeine, was monitored in each sample. Variation in intensity
were obtained in the mass range of 100–400 Da of the IMS peak at m/z 195 in each hair sample was observed
Int. J. Ion Mobil. Spec. (2008) 11:61–69 65

collected from general population. The focus of the study Figure 3 shows selected ion mobility (SIM) spectra of
was to investigate if drugs such as nicotine (m/z=162) and or hair samples A, B, E, and G at m/z value of 195 acquired by
caffeine (m/z=194) in the collected hair samples could be operating the MS in the single ion monitoring mode. The
extracted and detected by ESI-IMMS. Out of nine hair drift times of the IMS peak at m/z 195 was measured to be
samples analyzed,
 based on m/z values
of protonated
 ions of 19.84, 19.81, 19.88, and 19.86 ms in samples A, B, E, and
nicotine ðMHÞþ  ðm=z ¼ 163Þ and caffeine ðMHÞþ  G, respectively, with Ko values of 1.53±0.02 cm2 V−1 s−1.
ðm=z ¼ 195Þg, peaks corresponding to caffeine were With three consecutive measurements, the standard devia-
detected in hair samples identified as A, B, E, and G and tion in drift time for all samples was measured to be less
peaks corresponding to nicotine were detected in hair than ±0.06 ms. The Ko for standard solution of caffeine at
samples identified as A and E. The peaks detected at m/z m/z 195 was measured to be 1.53±0.01 cm2 V−1 s−1 which
values of 163 and 195 were identified as that of nicotine and compared well with the reduced mobility measured for the
caffeine extracted from hair because (1) the m/z values and detected ion of m/z 195 in each of the hair samples. SIM
reduced mobility values (Ko) of these peaks matched those spectra for the negative control at m/z 195 is also shown.
measured for standard solutions of nicotine and caffeine Hair samples containing both caffeine and nicotine also
analyzed by ESI-IMMS and (2) these peaks were not produced a peak at m/z value of 231 Da and drift time of
detected in the negative control. The Ko values of the ions 21.26±0.04 ms. The ion was tentatively identified as water
at m/z 163 and m/z 195 in hair sample A also matched those adduct peak of caffeine fM 2ðH2 OÞHgþ  ðm=z ¼ 231Þ. By
measured in hair sample E. comparing mobility values measured for the standard

Fig. 4 Ion mobility spectra il- 150 163 195 231


lustrating the detection of nico-
0.8
tine in hair samples. On the left
panel are shown the NSIM
spectra of hair samples A, and E 0.05 Hair sample A
with highlighted nicotine peak at m/z 163 amu
m/z value of 163. Also shown 0.6
drift time 19.24 ms
IMS Response (nA)

are the peaks at 150, 195, and


231 Da for reference. The right
hand panel shows the SIM 0.03
spectra obtained at m/z 163 with 0.4
drift times of 19.24 and
19.26 ms in hair samples A and
E, respectively
0.01
0.2
Hair sample A

0.0 -0.01
17 18 19 20 21 22 23 0 10 20 30 40 50

0.8

0.05 Hair sample E


m/z 163 amu
0.6
drift time 19.26 ms
IMS Response (nA)

0.03
0.4

0.01
0.2

Hair sample E

0.0 -0.01
17 18 19 20 21 22 23 0 10 20 30 40 50
Drift Time (ms) Drift time (ms)
66 Int. J. Ion Mobil. Spec. (2008) 11:61–69

caffeine solution and the peak observed in the hair samples


A A, B, E, and G at m/z 195 it was concluded that the samples
Hair Sample A; m/z 150
Drift time 18.14 ms contained caffeine and can be determined by ESI-IMMS
method.
NSIM spectra of hair sample extracts A and E in the drift
time range of 17–22 ms are shown in Figure 4, left hand
side panel. The IMS peaks at m/z values of 150, 163, 195
and 231 were identified in the two spectra for reference. As
shown, a peak at m/z value of 163 was detected in both
samples. From the acquired SIM spectra the drift times of
IMS Response

B the peak at m/z 163 was measured to be 19.24 and 19.26 ms


Methamphetamine standard
m/z 150; Drift time 18.66 ms in the samples A and E respectively as shown in the right
hand panel of Fig. 4. The reduced mobility for nicotine in
the standard solution was determined as 1.60±0.02 cm2
V−1 s−1. This peak was tentatively identified as that of
nicotine by comparing the mobility of the m/z 163 peak
C
detected in hair samples to the m/z 163 peak detected for a
Methamphetamine standard solution of nicotine. A Ko value of 1.61±0.01 cm2
in Hair Sample A; m/z 150 V−1 s−1 was measured for the standard nicotine solution.
Drift time 18.11 and 18.65 ms
Figure 4, left hand side panel, also shows an intense
peak at m/z value of 150 that was detected in hair sample A.
Since methamphetamine (m/z=149), a commonly used drug
of abuse, produces a peak at m/z 150 Da ðMHÞþ 
10 15 20 25 30 35 40
ðm=z ¼ 150Þg, IMMS measurements were performed to
Drift Time (ms)
tentatively identify the peak. Figure 5 shows the SIM
Fig. 5 SIM spectra at m/z 150 showing the separation of isobaric
ions. SIM spectra of hair sample A (a), standard solution of spectra at m/z 150 obtained for (a) hair sample A, (b)
methamphetamine (b) and hair sample A spiked with methamphet- standard methamphetamine solution, and (c) hair sample A
amine (c) are shown spiked with methamphetamine. The reduced mobility value

Fig. 6 NSIM spectra of vitamin


C (a), vitamin B (b), aspirin (c), 3.4
hair gel (d), and hair sample A (a)
(e). IMS peak with a drift time Vitamin B
of 18 ms at m/z 150 observed in
hair sample A is aligned with
that observed in the NSIM (d)
1.6
spectrum of the hair gel Hair gel
IMS Response (nA)

-0.2
(b)
Vitamin C 3.4

(e)
1.6 Hair sample A
(c)
Aspirin

-0.2

10 15 20 25 30 10 15 20 25 30
Drift time (ms) Drift time (ms)
Int. J. Ion Mobil. Spec. (2008) 11:61–69 67

Calibration curve for caffeine


of the m/z 150 peak in the hair sample was measured to be 4.0
1.63±0.02 cm2 V−1 s−1. However, the standard solution of 3.5
methamphetamine produced a peak at m/z 150 with a y = 0.721x -0.000

IMS Response (nA)


3.0
reduced mobility of 1.69±0.01 cm2 V−1 s−1 was produced. 2.5
R2 = 0.999

Mass-mobility comparison between standard and hair 2.0


sample A suggested that the m/z 150 peak detected in the 1.5
hair sample A was not methamphetamine. The separation of 1.0
isobaric ions based on mobility measurements also demon- 0.5
strates one of the advantages of using IMS as a rapid 0.0
separation method prior to MS analysis. 0 1 2 3 4 5 6
In an effort to identify the m/z 150 peak detected in hair Mixing Ratio of Caffiene in ppm
extract, various commonly used hair cosmetics such as gels,
Calibration curve for nicotine
shampoos and creams, and over the counter drugs and 1.6
health supplements that were suspected to be the source of 1.4
y = 0.138x + 0.0231
the peak at m/z 150, were analyzed by ESI-IMMS.

IMS Response (nA)


1.2
R2 = 0.996
Although all samples produced distinguishable ion mobility 1.0
patterns, the drift time of the m/z 150 peak observed in the 0.8
hair sample A matched only with a peak observed in one of 0.6
the hair gels analyzed. Non selective ion mobility spectra of 0.4
vitamin B, vitamin C, aspirin and hair gel are shown in 0.2
Fig. 6. On the right hand side panel of the figure is shown 0.0
the NSIM spectrum of the hair sample A aligned with the 0 2 4 6 8 10 12
NSIM spectrum of hair gel. The presence of a m/z 150 peak Mixing Ratio of Nicotine in ppm

in the hair gel at a drift time that matched with that Fig. 7 Calibration curve for caffeine and nicotine. Standard solutions
of caffeine and nicotine in the mixing ratio range of 0.01–5.0 ppm for
measured in the hair sample suggested that the peak caffeine and 0.05–10.0 ppm for nicotine were used. Measured
observed in the hair sample at m/z 150 could have concentration of caffeine in the hair samples A, B, E, and G varied
originated from the hair gel either as adsorbed or absorbed between ∼8.2 to ∼0.5 ng/mg. Nicotine concentration was measured to
hair gel component. In the absence of MS-MS capability be ∼12 and ∼ 6 ng/mg in hair samples E and A respectively
the identity of the m/z 150 peak was not confirmed.
However, the basic inference drawn from the experiment measured to be 0.5 ng/mg of hair sample. Concentration
was that IMS can rapidly provide a confirmatory piece of of nicotine in hair sample E was measured to be ∼12 ng of
data to distinguish ions with same m/z values present in nicotine in 1 mg of hair whereas concentration of nicotine
complex samples. in hair sample A was measured to be ∼6 ng/mg of hair. For
more accurate quantification of analytes, calibration curves
Determination of concentration of extractible target should be generated using control hair samples with known
compounds in hair samples amount of target compounds, and recovery data. The study
illustrates the basic concept that ESI-IMMS method of
To determine the extractible concentration of caffeine and analysis has the ability to rapidly analyze hair samples for
nicotine incorporated in hair, calibration curves for both either target compound measurement or comprehensive
were constructed by spiking hair samples in which peaks determination.
for caffeine and/or nicotine were not detected with known
amount of caffeine and nicotine. Standard solutions of
caffeine and nicotine in mixing ratio range of 0.01–5.0 and Conclusions
0.01–10.0 ppm respectively were analyzed by ESI-IMMS.
Intensity of IMS spectra obtained from an average of 2,500 Ion mobility spectrometry in tandem with mass spectrom-
IMS spectra was measured. Average signal intensity of etry and electrospray ionization (ESI-IMMS) can be used as
three consecutive measurements was plotted against the a rapid analytical tool for hair analysis. The ability of ion
concentration of the standard caffeine and nicotine solu- mobility spectrometry to rapidly provide confirmatory
tions (Fig. 7). In the concentration range studied, linear information in a second dimension allows one to exclude
relationship between the concentration and IMS signal false positives and increase confidence in the identification
intensity was observed. Concentration of caffeine in hair of ions of interest. Also, the method does not require
sample A was measured to be ∼8.2 ng of caffeine in 1 mg derivatization or extensive sample preparation. This method
of hair. Caffeine concentration in hair sample G was can be used either for measurement of target compound or
68 Int. J. Ion Mobil. Spec. (2008) 11:61–69

for the determination of all constituents extracted from hair 13. Brochart M (1978) Investigation of nutritional deficiencies and
imbalances by the chemical analysis of hair. Vet Sci Commun
samples. ESI-IMMS method can thus be extended to target
2:183–192
metabolite analysis or studies investigating the whole hair 14. Afridi H, Kazi T, Jamali M, Kazi G, Arain M, Jalbani N, Shar G
metabolome. With ability to analyze hair as a biological (2006) Analysis of heavy metals in scalp hair samples of
matrix and detect caffeine and nicotine at low nano-molar hypertensive patients by conventional and microwave digestion
methods. Spectrosc Lett 39:203–214
concentrations demonstrated, further modifications in sam-
15. Takeuchi T (1972) Relation between mercury concentration in
ple preparation methods and/or improvement in instrument hair and the onset of Minamata disease. Environ Mercury Contam,
sensitivity can facilitate analysis of hair for determination Int Conf 302–304
of constituents (metabolites, drugs) present in sub nano- 16. Jurado C, Rodriguez-Vicente C, Menendez M, Repetto M (1997)
Time course of cocaine in rabbit hair. Forensic Sci Int 84:61–66
molar concentrations. Additionally, ESI-IMMS or ESI-IMS 17. Sachs H (1997) History of hair analysis. Forensic Sci Int 84:7–16
as a stand-alone method can be exploited for rapid screen- 18. Baumgartner AM, Jones PF, Baumgartner WA, Black CT (1979)
ing of hair samples, cosmetics and pharmaceuticals. Radioimmunoassay of hair for determining opiate-abuse histories.
J Nucl Med 20:748–752
Acknowledgements The authors would like to thank Fantastic Sams 19. Baumgartner WA, Hill VA, Blahd WH (1989) Hair analysis for
of Pullman, Washington for providing hair samples. Also, the authors drugs of abuse. J Forensic Sci 34:1433–1453
would like to thank the Alcohol and Drug Abuse Research Program 20. Uhl M (1997) Determination of drugs in hair using GC/MS/MS.
(ADARP) of Washington State University (WSU) and National Forensic Sci Int 84:281–294
Institutes of Health (NIH) for providing financial support. This work 21. Lin DL, Yin RM, Liu RH (2005) Gas chromatography-mass
was supported by an ADARP grant and a Road Map Grant from NIH spectrometry (GC-MS) analysis of amphetamine methamphet-
the (R21 DK 070274). amine, 3,4-methylenedioxyamphetamine and 3,4-methylenediox-
ymethamphetamine in human hair and hair sections. Journal of
Food and Drug Analysis 13:193–200
22. Kronstrand R, Nystrom I, Strandberg J, Druid H (2004) Screening
for drugs of abuse in hair with ion spray LC-MS-MS. Forensic Sci
References Int 145:183–190
23. Welch MJ, Sniegoski LT, Allgood CC, Habram M (1993) Hair
1. Gibson GS, Manley S (1991) Alternative approaches to urinalysis analysis for drugs of abuse—evaluation of analytical methods,
in the detection of drugs. Soc Behav Pers 19:195–204 environmental-issues, and development of reference materials. J
2. Katz SA, Wood JD (1980) Hair. A monitor of chemical exposure Anal Toxicol 17:389–398
and disease. Chem Int 6:12–15 24. Miyazawa N, Uematsu T, Mizuno A, Nagashima S, Nakashima M
3. Sabbioni E, Minoia C, Pietra R, Mosconi G, Forni A, Scansetti G (1991) Ofloxacin in human hair determined by high-performance
(1994) Metal determinations in biological specimens of diseased liquid-chromatography. Forensic Sci Int 51:65–77
and non-diseased hard metal workers. Sci Total Environ 150:41–54 25. Tagliaro F, Smyth WF, Turrina S, Deyl Z, Marigo M (1995)
4. Metcalf SW, Orloff KG (2004) Biomarkers of exposure in Capillary electrophoresis—a new tool in forensic toxicology—
community settings. J Toxicol Environ Health—Part a—Current applications and prospects in hair analysis for illicit drugs.
Issues 67:715–726 Forensic Sci Int 70:93–104
5. Akanle O, Ho J, Muhiddin K, Admans L, Croft N, Spyrou N 26. Tagliaro F, Poiesi C, Aiello R, Dorizzi R, Ghielmi S, Marigo M
(2004) Elemental composition of hair and bone density measure- (1993) Capillary electrophoresis for the investigation of illicit
ments at diagnosis in paediatric inflammatory bowel disease. J drugs in hair: determination of cocaine and morphine. J
Radioanal Nucl Chem 259:355–360 Chromatogr 638:303–309
6. Williams J, Patsalos PN, Wilson JF (1997) Hair analysis as a 27. Zaugg S, Thormann W (2000) Enantio selective determination of
potential index of therapeutic compliance in the treatment of drugs in body fluids by capillary electrophoresis. J Chromatogr A
epilepsy. Forensic Sci Int 84:113–122 875:27–41
7. Mogos T, Popescu C, Tanase I, Panus C, Mincu I (1993) 28. Kintz P, Cirimele V, Sengler C, Mangin P (1995) Testing human
Predictive value of glucose and proteic sulphur levels in the hair hair and urine for anhydroecgonine methyl ester, a pyrolysis
for an early detection of the preclinical stage of IDDM. Rom J product of cocaine. J Anal Toxicol 19:479–482
Internal Med 31:287–290 29. Toyo'oka T, Suzuki A, Fukushima T, Kato M (2004) Hair analysis
8. Kimura A, Kanekura T, Saito Y, Sagawa K, Nosaka M, Kanzaki of histamine after fluorescence labeling by column-switching
T, Tsuji T (2005) Blood group A glycosphingolipid accumulation reversed-phase liquid chromatography with electrospray ioniza-
in the hair of patients with [alpha]-N-acetylgalactosaminidase tion mass spectrometry and application to human hair. Anal
deficiency. Life Sciences 76:1817–1824 Biochem 333:236–245
9. Silengo M, Valenzise M, Sorasio L, Ferrero GB (2002) Hair as a 30. Asbury GR, Hill HH (2000) Evaluation of ultrahigh resolution ion
diagnostic tool in dysmorphology. Clin Genet 62:270–272, Section mobility spectrometry as an analytical separation device in
Editor: Albert E Chudley, e-mail: chudley@cc.umanitoba.ca chromatographic terms. Journal of Microcolumn Separations
10. Anon (1999) Hair analysis may one day screen for breast cancer. 12:172–178
Lab Med 30:440–441 31. Hill HH, Siems WF, Stlouis RH, Mcminn DG (1990) Ion mobility
11. Piccinini L, Borella P, Bargellini A, Medici CI, Zoboli A (1996) A spectrometry. Anal Chem 62:A1201–A1209
case-control study on selenium, zinc, and copper in plasma and 32. Keller T, Miki A, Regenscheit P, Dirnhofer R, Schneider A,
hair of subjects affected by breast and lung cancer. Biol Trace Tsuchihashi H (1998) Detection of designer drugs in human hair
Elem Res 51:23–30 by ion mobility spectrometry (IMS). Forensic Sci Int 94:55–63
12. Qin J, Li Z (2003) Diagnosis of atherosclerosis and hair analysis. 33. Miki A, Katagi M, Tsuchihashi H (2003) Recent improvements in
Guangdong Weiliang Yuansu Kexue 10:1–10 forensic hair analysis for illicit drugs. J Health Sci 49:325–332
Int. J. Ion Mobil. Spec. (2008) 11:61–69 69

34. Miki A, Keller T, Regenscheit P, Bernhard W, Tatsuno M, Katagi 38. Taraszka JA, Counterman AE, Clemmer DE (2001) Gas-phase
M, Nishikawa M, Kim L, Hatano S, Tsuchihashi H (1997) separations of complex tryptic peptide mixtures. Fresenius J Anal
Detection of internal and external methamphetamine in human Chem 369:234–245
hair by ion mobility spectrometry. Japanese Journal of Toxicology 39. Lee DS, Wu C, Hill HH (1998) Detection of carbohydrates by
and Environmental Health 43:15–24 electrospray ionization ion mobility spectrometry following micro-
35. Miki A, Keller T, Regenscheit P, Dirnhofer R, Tatsuno M, Katagi bore high-performance liquid chromatography. J Chromatogr A
M, Nishikawa M, Tsuchihashi H (1997) Application of ion 822:1–9
mobility spectrometry to the rapid screening of methamphetamine 40. Wu C, Siems WF, Klasmeier J, Hill HH Jr. (2000) Separation of
incorporated in hair. J Chromatogr. B, Biomedical Sciences and isomeric peptides using electrospray ionization/high-resolution
Applications 692:319–328 ion mobility spectrometry. Anal Chem 72:391–395
36. Shumate CB, Hill HH, Jr. (1987) Electrospray ion mobility 41. Dwivedi P, Matz LM, Atkinson DA, Hill HH (2004) Electro-
spectrometry after liquid chromatography. In: 42nd Northwest spray ionization-ion mobility spectrometry: a rapid analytical
Regional Meeting of the American Chemical Society, Bellingham, method for aqueous nitrate and nitrite analysis. Analyst 129:
WA. 139–144
37. Shumate CB, Hill HH Jr. (1989) Coronaspray nebulization and 42. Wu C, Siems WF, Asbury GR, Hill HH (1998) Electrospray
ionization of liquid samples for ion mobility spectrometry. Anal ionization high-resolution ion mobility spectrometry-mass spec-
Chem 61:601–606 trometry. Anal Chem 70:4929–4938

You might also like