Professional Documents
Culture Documents
Introduction Page 1
User Manual
NEUROWERK EMG
EMG/NCV/EP System
MANUFACTURER
Table of Contents
A Introduction ....................................................................................................................................7
A.1 Field of application, indications/contraindications................................................................................7
A.2 Warnings and notes ...........................................................................................................................................8
A.3 Electrical safety ....................................................................................................................................................8
A.4 Safety notes on the EMC regulations...........................................................................................................9
A.5 Safety notes on the electrical stimulator....................................................................................................9
A.6 Further safety notes ........................................................................................................................................ 10
A.7 CE marking ......................................................................................................................................................... 10
A.8 Equipotential bonding................................................................................................................................... 10
A.8.1 Place of installation........................................................................................................................ 10
A.8.2 Combination with other equipment....................................................................................... 11
A.8.3 Using a defibrillator ....................................................................................................................... 11
A.8.4 Using high-frequency surgical devices .................................................................................. 11
A.9 Conditions for proper use............................................................................................................................. 11
A.9.1 Before getting started................................................................................................................... 11
A.9.2 When working with the equipment ........................................................................................ 12
A.9.3 After finishing work ....................................................................................................................... 12
A.9.4 Inspection and maintenance ..................................................................................................... 12
A.9.5 Hygiene and care of the equipment ....................................................................................... 12
A.9.6 Cleaning and sterilisation of electrodes................................................................................. 12
A.9.7 General notes................................................................................................................................... 14
A.9.8 Environmental protection........................................................................................................... 14
A.9.9 Symbols used and their meaning............................................................................................. 15
B Design and Installation ...............................................................................................................17
B.1 Hardware............................................................................................................................................................. 17
B.1.1 Technical description.................................................................................................................... 17
B.1.2 System components:..................................................................................................................... 20
B.1.3 Connections on the electrode input box............................................................................... 22
B.1.4 Connector panel on the central unit....................................................................................... 23
B.1.5 Connections on the PC................................................................................................................. 27
B.1.6 Remote control ............................................................................................................................... 28
B.2 Installation.......................................................................................................................................................... 29
B.2.1 Place of installation........................................................................................................................ 29
B.2.2 Mains power connection and equipotential bonding...................................................... 29
B.2.3 Connecting electrodes and peripherals................................................................................. 30
B.2.4 Switching on the equipment ..................................................................................................... 30
0 NEUROWERK EMG
Page 4 CONTENTS
A Introduction
When compiling this User Manual, particular importance was attached to a clear and
readily understandable description of the individual applications.
The User Manual is intended to enable effective use of the NEUROWERK EMG system
and of the application possibilities it offers.
The User Manual is an integral part of the measuring system. It should always be
kept in the immediate vicinity of the system. Careful adherence to the User Man-
ual is imperative, as a condition for proper use and correct handling of the meas-
uring system.
The manufacturer reserves the right to implement modifications for technical im-
provement of the system without prior notice.
Please read this User Manual carefully before you install and start up the system. The
system must only be used by persons whose training, knowledge and practical experi-
ence constitute a guarantee for proper handling; these persons must have been in-
structed accordingly on proper handling of the system in accordance with the con-
tents of this User Manual. Such instruction must only be given by persons who, by way
of their knowledge and practical experience, are suitably qualified to provide instruc-
tion on the handling of the system.
Exact examination results and safe, trouble-free operation are only guaranteed if the
system is operated in accordance with the Instruction Manual.
4. The MEP examination (which requires an optional magnetic stimulator not included
in the NEUROWERK EMG) uses the transcranial or spinal magnetic stimulation to assess
proximal sections of the neuromuscular system.
Contraindication is to be noted for persons with cardiac pacemakers or intracranial fer-
romagnetic materials, and for patients suffering from epilepsy.
Due to the increased safety requirements applicable, the NEUROWERK EMG system
must not be connected via an extension cable or multi-socket distribution unit.
The device earthing is provided by the PE conductor of the mains cable.
A.7 CE marking
SIGMA Medizin-Technik maintains a certified quality management system compliant
with the European standards EN ISO 9001 and EN ISO 13485 to monitor the whole life
cycle of the equipment it develops and manufactures. With our quality control proce-
dures, we are able to guarantee application of the fundamental standards defined in
the European directives, a maximum level of safety for the patients and users of the
equipment, and the optimum serviceability of our equipment.
The equipment meets the specifications of Hazard Class IIa (directive 93/42/EEC, Annex
IX).
Damaged electrode cables cannot be repaired. Use a multimeter to check any cable
thought to be defective. If the measurement shows that the cable is indeed damaged,
provide for proper disposal of the electrode and replace it with a new one.
L
CAUTION:
EMG needle electrodes must always be sterilised before use (also before
their first use), in order to eliminate the risk of infections with hepatitis or
HIV. Cold sterilisation with chemical means is not sufficient.
CAUTION:
Earth electrode
Clean the earth electrode with water after use and allow it to dry to avoid oxidation of
the metal parts.
Stimulation electrodes
Remove the felts from the stimulation electrodes and leave them to dry. Clean the
mounting holes for the stimulation electrodes with a cotton bud.
Finger electrode
Rinse the loops under running water. Subsequently dry the electrode carefully with a
cloth.
Symbol Description
B.1 Hardware
The NEUROWERK EMG system comprises essentially the following components:
o Personal computer or laptop (option)
- CPU: Intel processor
- ≥512 MB RAM
- Hard disk > 80 GB
- DVD-RW drive
- 2 PCI slots
- Graphics adapter: max. 1,280 x 1,024 pixels
- Software: Windows XP Professional SP2
- TFT monitor, mouse, keyboard
- Safety isolating transformer
- Power supply: 230 V / 50 - 60 Hz, connected load max. 400 W
o Central unit with stimulators
o 2-channel, optionally 4-channel, AC amplifier
o Instrument trolley with support arm for amplifier (optional)
o EMG monitor speakers
o AEP headphones (EMG/NCV/EP system only)
o VEP monitor (EMG/NCV/EP system only)
o Remote control unit (optional)
o Inkjet printer (optional)
o NEUROWERK EMG software
o Electrode starter kit: EMG, NCV and EP electrodes as per SIGMA Medizin-
Technik accessories catalogue
o GDT interface, linking to surgery EDP (optional)
The conductive link-up to the NEUROWERK EMG system is to be entrusted exclusively
to the SIGMA Medizin-Technik service department or a person authorised by the ser-
vice department, observing the requirements of IEC 601-1-1.
erator in the central unit. The traced EP signals are amplified accordingly in the elec-
trode input box, filtered, digitised, and issued to the PC software via the central unit.
The system is operated via the PC software.
The central unit comprises the pattern, goggle and headphone stimulators, and the
electrical stimulator.
The central unit communicates with the PC via a data line (network). This line transfers
bidirectional data for the settings of the stimulator (central unit) and the amplifier
(electrode input box), checkback messages to the PC, and the actual measured data
(evoked potentials measured using the electrode input box).
The central unit transmits control signals to the electrode input box via a serial inter-
face.
LEDs in the central unit indicate the following states:
Electrical stimulator
This module generates the stimulation current for the electrical stimulator. The mod-
ule can be activated or deactivated. Furthermore, it is possible to activate or deactivate
the 400 V generation separately. The polarity can be inverted. The current can be set
between 0mA and 100mA in 25µA steps. If this current is not reached at the stimula-
tion electrode, the module issues a checkback message. This is the case if 100mA is set
for electrode impedances of >4kΩ. If the current is only halved, the critical electrode
impedance is doubled.
Goggle module
This module generates the stimuli for the visual stimulator (flash goggles). The module
can be activated or deactivated.
Visual stimulator
NEUROWERK EMG B
General handling Page 19
The monitor displaying the chequered pattern is connected to this module. It is possi-
ble to display a chequered pattern, horizontal and vertical lines. Furthermore, user-
defined fixation points (images) can be displayed, and it is possible to move the posi-
tion of the fixation point. The chequered pattern may consist of 4, 8, 16, 32, 64 or 128
fields. In addition to 0% (black screen), three contrast values (33%, 67% and 100%) can
be set.
Acoustic stimulator
The acoustic stimulator can output stimulus volumes of 0..125dBSPL or up to 100dBHL.
The masking of the contralateral ear is achieved by way of a white noise at a volume
below the stimulus volume. The volume is set in the trigger matrix in 0.375dB. The
stimulus types are "click" (positive stimulus, condensation), "clack" (negative stimulus,
rarefaction), alternating, random, tones (which are defined by the user) or white noise
(for masking of the contralateral ear).
B NEUROWERK EMG
Page 20 General handling
10
3
9
11
2
1
5
12
NEUROWERK EMG B
General handling Page 21
Design
1 2
3 4 5 6
7
NEUROWERK EMG B
General handling Page 23
Connectors
3 Electrode input for the ac- "-" connection for electrodes with DIN con-
tive electrode nectors
If this input is used, then the electrode
cable must be removed from socket 2.
4 Electrode input for the ref- "+" connection for electrodes with DIN
erence electrode connectors
If this input is used, then the electrode
cable must be removed from socket 2.
1 2 3
Connectors
ON
Connectors
1 2 3 4
Connectors
1 2 3 4 5 6 8 10
7 9
Connectors
1 2 3 4 5 6 7 9
Connectors
"Analysis"
mode
Stop
Trigger
B.2 Installation
CAUTION:
L
CAUTION:
Never use the gas or water piping for equipotential bonding. Always con-
nect the cable to the proper equipotential bonding connection.
C Software Installation
This chapter describes the initial installation of the NEUROWERK EMG software on a PC.
This installation is only necessary if the system is supplied without a PC (system kit).
If the system is supplied together with a PC, the NEUROWERK EMG software is already
installed.
L CAUTION:
Never perform impedance measurements with the needle electrodes in use!
After use, remove the electrode from the connection cable and clean it carefully with
alcohol or a medical cleansing solution, e.g. to prevent blood drying on the electrode.
Then slide the protective sheathing over the electrode again.
D.5 Headphones
Headphones suitable for use with the NEUROWERK EMG system:
Type DT 48 200 Ohm
L Note: Only these headphones are to be used with the NEUROWERK EMG sys-
tem!
D NEUROWERK EMG
Page 34 Electrodes and Accessories
1) Put the headphones on the patient such that the ear-pieces enclose the ears
completely.
2) Connect the headphones cable to the headphones socket on the right-hand
side of the central unit. Make sure that the left and right-hand ear-pieces of the
headphones sit on the correct ear. The headphones are marked with the let-
ters L for left and R for right.
Cleaning
If required, clean the headphones with a damp cloth. Do not use solvents.
Control
System designation Menu bar functions
Program version
On-line
Raw data parameters
window
Sequence
information
Trace window
Numerical
Cursor informa- results window
tion
Quick method
selection
Graphic results
display
Report window
E NEUROWERK EMG
Page 36 General handling
Results table
Trace window
MU MUAP
Analysis
window
Selected MUAP
Selected MUAP
NEUROWERK EMG E
General handling Page 37
The buttons can be selected either by clicking with the mouse or by way of the func-
tion keys on the PC keyboard.
Patient
selection Method
selection
Help menu
Patient Method
display display
Create a trace
report
Quit the
NEUROWERK EMG
program
The buttons can be selected either by clicking with the mouse or by way of the func-
tion keys on the PC keyboard. For certain functions, it is also possible to use the op-
tional remote control unit or the foot switch. The use of the individual control ele-
ments is explained under the corresponding method descriptions.
Stimulation-
synchronous Print out the
display measure-
Save the ex- ments
amination
Here, you can select the stimulation mode. You can work with single stimuli or cyclic
stimulation, and can also set the pulse width for the stimulation.
The buttons can be selected either by clicking with the mouse, by using the cursor
keys of the PC keyboard, or by way of the optional remote control unit.
Using the mouse: Click on the parameter, move the mouse over the parameter box
and change the parameter value with the aid of the up and down hand symbols.
The sensitivity and time basis can be modified simultaneously for all tracks by pressing
the [SHIFT] key together with the left mouse button.
Using the cursor keys: Select the parameter with the Right/Left Arrow keys, and then
change the parameter value with the Up/Down Arrow keys.
Using the remote control unit: Select the parameter with the outer wheel, and then
change the parameter value with the left/right handwheel.
Time resolution
(per division) Single stimulus /
cyclic stimulation
Stimulation pulse
Amplitude resolu- width (electrical
tion (per division) stimulation only)
Number of avera-
ged/discarded
sequences
Distance specifica- Stimulation
tion (only for NCV) intensity
Sound pressure,
Sound pressure,
right ear
left ear
E NEUROWERK EMG
Page 40 General handling
Single stimulus /
cyclic stimulation
Sequence information
After clicking on the trace point (A), stimulation point (R) or additional information (Z),
it is possible to change the corresponding entries.
Moving curves
Position the mouse pointer on the currently active curve such that the mouse arrow is
displayed as a vertical double arrow. The curve can be moved up and down by press-
ing and holding down the left mouse button.
Keyboard assignments
F7 Analysis S Save
NEUROWERK EMG E
General handling Page 41
Patient database
The patient database window is used to select a patient for an examination
or to open an existing examination for a particular patient.
The patient database is displayed automatically when the program is started.
If the display presently shows the main screen in which examinations are performed,
the patient database can be called with:
Patient database
Mouse Keyboard
"F2"
E NEUROWERK EMG
Page 42 General handling
New patient
New patient
Mouse Keyboard
Mouse Keyboard
New examination
Mouse Keyboard
F3
The method selection window is displayed automatically (see the section "Method se-
lection")
Mouse Keyboard
F6
The saved examinations of this patient are listed on the right-hand side of the screen.
The method selection window is displayed if you select a patient and then select the
"Method F3" button in the patient database window.
First select the required method.
Method selection
Mouse Keyboard
"F3"
The method selection window is displayed automatically if you select a patient and
then select the "TRACE" button in the patient database window.
First select the required method.
NEUROWERK EMG E
General handling Page 45
Selecting a method
Mouse Keyboard
Click on the "plus" sign in Arrow Right
front of the method
then
Arrow Up/Down
or
first letter
then
Arrow Right and Arrow
Up/Down
Mouse Keyboard
Double-click on the examina- Arrow Right
tion definition
then
Arrow Up/Down
or
first letter
then
Arrow Right and
Arrow Up/Down
Confirm the selection with
Enter
Once the examination program has been selected, the actual starting of the program
will vary according to the current configuration (see the section "Examination pro-
grams").
E.3 Preparation
Optionally, you can choose to precede an examination sequence with a number of
preparation steps (see the section "Setup").
Possible settings which can be made under "Preparation", depending on the selected
examination, are:
1. Trace point and additional text: Choose a name from the list or else enter a new
text of your own. You can add remarks as information on the particular sequence.
2. Stimulation point: Choose a name from the list or else enter a new text of your
own.
3. Impedance measurement: An impedance measurement is performed automati-
cally for the active channels of the examination sequence. The results are dis-
E NEUROWERK EMG
Page 46 General handling
You can move from step to step through the preparation with:
Preparation
E.4 Monitor
In monitor mode, the quality of the incoming signal can be assessed before starting
the measurement.
E.5 Trigger
The trigger mode serves to display stimulation-triggered signals. Measurements, e.g.
motor neurography, can be performed with either single stimuli or cyclic stimulation.
This mode is also used to determine the current strength and the position of the
stimulation electrode for measurements involving an averager or cascaded display.
E.6 Analysis
This mode is used to display the signals as appropriate for the selected examination
(e.g. averaging, cascaded display).
Inserting a sequence
Mouse Keyboard
-- "Ins"
Removing a sequence
Mouse Keyboard
-- "Del"
NEUROWERK EMG E
General handling Page 47
A selection window is displayed. You can here define the information to be included in
the print-out.
Starts printing
Opens a preview
window
Summarise measurements
for printing
Printing
Display Explanation
Traces as on screen Print-out with or without traces
Traces as displayed on the screen
A click with the left mouse button on Edit report opens the report you have created
for further editing.
The default setting is to print the traces, trace information and the screen analysis.
You can vary the contents of the print-out by clicking on the corresponding check
boxes with the left mouse button to activate/deactivate the options.
Clinic/practice header
Patient ID
Traces as on the
screen
Trace information
Analysis screen
Operator controls
Editing of a summary
report
If the "Report" field is selected in the "Summary Report" selection window, you can en-
ter a summary report with a double-click at the end of the summary report.
NEUROWERK EMG E
General handling Page 51
->
Existing summary reports can be called directly from the NEUROWERK CENTER data-
base.
A new summary report is created each time the "Print"/"Summary" function is used. In
this way, it is also possible to create several different versions of the summary report
(e.g. with and without curves, different examinations) from the examinations of a pa-
tient.
L
Note: If the contact resistance at the stimulation electrode is too high, the
yellow lamp STIM1 or STIM2 on the central unit will light permanently in the
case of electrical stimulation and, in addition, the
symbol is displayed on the screen in the title bar. This may have the follow-
ing reasons:
- the stimulation electrode is not connected to the patient;
- the transition resistance from the stimulation electrode to the skin is too
high;
- the stimulation electrode is defective (cable broken), and a stimulus is trig-
gered.
F Examination programs
Patient selection
The patient selection window is displayed automatically when the program is started.
Otherwise, you can switch to patient selection with:
Patient selection
Mouse Keyboard
"F2"
F NEUROWERK EMG
Page 54 Method selection
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
Method selection
Mouse Keyboard
"F3"
Selecting a method
The method selection window is displayed. Here, you must first select the required
method.
Method selection
Mouse Keyboard
Click on the "plus" sign in Arrow Right
front of the method
then
Arrow Up/Down
or
first letter
then
Arrow Right and Arrow
Up/Down
Examination definition
Mouse Keyboard
Double-click on the examina- Arrow Up/Down
tion definition
or
Motor
first letter
then ENTER
NEUROWERK EMG F
Method selection Page 55
Preparation
The first examination sequence starts with the "Preparation" function.
This offers you the following possibilities:
a. Selection of the trace point (muscle) and entering of an additional text
b. Specification of the stimulation point (nerve)
c. Electrode positioning map (if activated in the setup)
Preparation
After the last preparation step, the "Continue" button takes you to the first examina-
tion sequence. The first examination sequence is started in the "Trigger" mode.
Check the stimulator parameters (current, pulse width) and modify, if necessary, in the
control bar:
F NEUROWERK EMG
Page 56 Method selection
Example: Current
Mouse Keyboard
Position the mouse pointer Select parameter
over the "Current" parameter
Arrow Left/Right
Over the upper half, the
Change parameter
mouse pointer shows a
Arrow Up/Down
" + " symbol
Use the left mouse button to
increase the current.
Over the lower half, the
mouse pointer shows a
" – " symbol
Use the left mouse button to
reduce the current.
Note: If the transition resistance at the stimulation electrode is too high, the
yellow lamp STIM1 or STIM2 on the central unit will light permanently in the
case of electrical stimulation and, in addition, the
symbol is displayed on the screen in the title bar. This may have the follow-
ing reasons:
- the stimulation electrode is not connected to the patient;
- the transition resistance from the stimulation electrode to the skin is too
high;
- the stimulation electrode is defective (cable broken), and a stimulus is trig-
gered.
Next sequence
Next sequence
Before you end the examination, you will always be asked whether or not the traces
are to be saved.
Print
You can print out the results of an examination with:
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
Method selection
Mouse Keyboard
"F3"
Selecting a method
The method selection window is displayed. First select the required method.
Method selection
Mouse Keyboard
Click on the "plus" sign in Arrow Right
front of the method
then
Arrow Up/Down
NCV or
first letter
then
Arrow Right and Arrow
Up/Down
F NEUROWERK EMG
Page 60 Method selection
Examination definition
Mouse Keyboard
Double-click on the method Arrow Up/Down
Sensory or
first letter
then ENTER
Examination definition
Mouse Keyboard
Double-click on the examina- Arrow Up/Down
tion definition
or
first letter
then ENTER
Preparation
The first examination sequence starts with the "Preparation" function.
This offers you the following possibilities:
a. Selection of the trace point (muscle) and entering of an addi-
tional text
b. Specification of the stimulation point (nerve)
c. Electrode positioning map (if activated in the setup)
Preparation
After the last preparation step, the "Continue" button takes you to the first examina-
tion sequence.
The first examination sequence is started in the "Trigger" mode.
Check the stimulator parameters (current, pulse width)
Example: Current
Mouse Keyboard
Position the mouse pointer Select parameter
over the "Current" parameter
Arrow Left/Right
Over the upper half, the
Change parameter
mouse pointer shows a
Arrow Up/Down
" + " symbol
Use the left mouse button to
increase the current.
Over the lower half, the
mouse pointer shows a
" – " symbol
Use the left mouse button to
reduce the current.
Mouse Keyboard
Click on "F7"
Next track
Next track
Before you end the examination, you will always be asked whether or not the traces
are to be saved.
Print
You can print out the results of an examination with:
F NEUROWERK EMG
Page 64 Method selection
Patient selection
The patient selection window is displayed automatically when the program is started.
Patient selection
Mouse Keyboard
"F2"
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
NEUROWERK EMG F
Method selection Page 65
Method selection
Mouse Keyboard
"F3"
Selecting a method
The method selection window is displayed. First select the required method.
Selecting a method
Mouse Keyboard
Click on the "plus" sign in Arrow Up/Down
front of the method
then
Arrow Right
NCV or
first letter
then
Arrow Right
Examination definition
Mouse Keyboard
Double-click on the examina- Arrow Up/Down
tion definition
or
Myasthenia test first letter, then ENTER
Preparation
The first examination sequence starts with the "Preparation" function.
F NEUROWERK EMG
Page 66 Method selection
Preparation
After the last preparation step, the "Continue" button takes you to the first examina-
tion sequence.
The first examination sequence is started in the "Trigger" mode.
Check the stimulator parameters (current, pulse width)
Example: Current
Mouse Keyboard
Position the mouse pointer Select parameter
over the "Current" parameter
Arrow Left/Right
Over the upper half, the
Change parameter
mouse pointer shows a
Arrow Up/Down
" + " symbol
Use the left mouse button to
increase the current.
Over the lower half, the
mouse pointer shows a
" – " symbol
Use the left mouse button to
reduce the current.
Start measurement
Mouse Keyboard
"F7"
Next track
Next track
The decrement calculation is performed automatically and shown in the results win-
dow.
The loss of amplitude between the first and fifth muscle responses is calculated and
displayed in graphic form.
Note: If you select a track with the left mouse button and then press the right
L mouse button, a context-sensitive menu appears and you can modify the
display.
Cascade vertical
Cascade overlaid
Cascade overlapping
NEUROWERK EMG F
Method selection Page 69
Print
You can print out the results of an examination with:
F.1.4 F response
Patient selection
The patient selection window is displayed automatically when the program is started.
Otherwise, you can switch to patient selection with:
Patient selection
Mouse Keyboard
"F2"
F NEUROWERK EMG
Page 70 Method selection
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
Method selection
Mouse Keyboard
"F3"
Selecting a method
The method selection window is displayed. Here, you must first select the required
method.
NEUROWERK EMG F
Method selection Page 71
Method selection
Mouse Keyboard
Click on the "plus" sign in Arrow Up/Down
front of the method
then
Arrow Right
NCV or
first letter
then
Arrow Right
Mouse Keyboard
Double-click on the examina- Arrow Up/Down
tion definition
or
F wave
first letter
then ENTER
Preparation
The first examination sequence starts with the "Preparation" function.
This offers you the following possibilities:
a. Selection of the trace point (muscle) and entering of an additional text
b. Specification of the stimulation point (nerve)
c. Electrode positioning map (if activated in the setup)
Preparation
After the last preparation step, the "Continue" button takes you to the first examina-
tion sequence. The first examination sequence is started in the "Trigger" mode.
Check the stimulator parameters (current, pulse width) and modify, if necessary, in the
control bar:
F NEUROWERK EMG
Page 72 Method selection
Example: Current
Mouse Keyboard
Position the mouse pointer Select parameter
over the "Current" parameter
Arrow Left/Right
Over the upper half, the
Change parameter
mouse pointer shows a
Arrow Up/Down
" + " symbol
Use the left mouse button to
increase the current.
Over the lower half, the
mouse pointer shows a
" – " symbol
Use the left mouse button to
reduce the current.
Start measurement
Print
You can print out the results of an examination with:
Patient selection
Mouse Keyboard
"F2"
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
Method selection
Mouse Keyboard
"F3"
Selecting a method
The method selection window is displayed. Here, you must first select the required
method.
NEUROWERK EMG F
Method selection Page 77
Selecting a method
Mouse Keyboard
Click on the "plus" sign in Arrow Up/Down
front of the method
then
Arrow Right
MEP or
first letter
then
Arrow Right
Mouse Keyboard
Double-click on the examina- Arrow Up/Down
tion definition
or
Lower extremities
first letter
then ENTER
All definitions are configured for a trigger signal from the Neurowerk EMG system to
trigger the magnet stimulator.
Order of measurement:
Cortex left
Nerve root stimulation
Cortex right
Nerve root stimulation
Monitor
In monitor mode, you can assess the quality of the incoming signal and the patient can
practice the tensing of the muscles before the actual measurement is started. The EMG
signal can be muted or turned back on using the spacebar.
You are now in the trigger mode. This mode serves to determine the stimu-
lation intensity and the stimulation position.
Trigger
Next track
Next track
Print
You can print out the results of an examination with:
F.3.1 SEP
This examination program is preconfigured to record the SEPs at cortical
trace points on the left and right following electrical stimulation of the N.
medianus, first with stimulation on the right.
A trace for a medianus SEP is taken as an example.
Patient selection
The patient selection window is displayed automatically when the program
is started.
Patient selection
Mouse Keyboard
"F2"
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
Selecting a method
Mouse Keyboard
"F3"
F NEUROWERK EMG
Page 82 Method selection
Selecting a method
The method selection window is displayed. Here, you must first select the required
method.
Selecting a method
Mouse Keyboard
Click on the "plus" sign in Arrow Up/Down
front of the method
then
Arrow Right
SEP or
first letter
then
Arrow Right
Mouse Keyboard
Double-click on the examina- Arrow Up/Down
tion definition
or
Medianus 1 channel
first letter
then ENTER
NEUROWERK EMG F
Method selection Page 83
Preparation
Electrode map
A The electrode map is displayed automatically (if activated in the setup).
Place the tracing electrodes such that tracing is first possible on the left hemisphere af-
ter stimulation on the right-hand side.
Mouse Keyboard
ENTER
Note: Correct the electrodes if the contact resistance exceeds 10 kOhm and
L the graph block is displayed red. If the impedance is less than 10 kOhm, the
graph block is green. The absolute impedance values should be as low as
possible. Ensure also that the values of the active (A) and reference elec-
trodes (R) are approximately the same.
Low impedance values are an important prerequisite for a good signal qual-
ity.
F NEUROWERK EMG
Page 84 Method selection
Mouse Keyboard
ENTER
The first examination sequence is started in the "Trigger" mode. Check the signal qual-
ity and position the stimulation electrode.
Mouse Keyboard
Position the mouse pointer Arrow Up/Down
over the "Current" parameter
Over the upper half, the
mouse pointer shows a
" + " symbol
Use the left mouse button to
increase the current.
NEUROWERK EMG F
Method selection Page 85
Start averaging
Start averaging
Impedance measurement
Determine the stimulation point and stimulation intensity
Start averaging
Manual operation
Next sequence
Next sequence
Start measurement
Start measurement
Save
Save
Reset averaging
To restart averaging from the beginning, press the button shown below during aver-
aging:
Reset averaging
Continue averaging
If you have interrupted the averaging and you are in the STOP mode, you can continue
the averaging by pressing:
NEUROWERK EMG F
Method selection Page 87
Continue averaging
Writing a report
You can enter a text with report pertaining to the current measurement.
Writing a report
Print
You can print out the results of an examination with:
F.4.1 FAEP
This examination program is preconfigured to record the early, acoustically evoked
potentials at cortical trace points on the right and left following acoustic stimulation of
the corresponding ear, first with stimulation on the right.
Patient selection
The patient selection window is displayed automatically when the program is started.
Patient selection
Mouse Keyboard
"F2"
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
Selecting a method
Mouse Keyboard
"F3"
NEUROWERK EMG F
Method selection Page 91
Selecting a method
The method selection window is displayed. First select the required method.
Selecting a method
Mouse Keyboard
Click on the "plus" sign in Arrow Up/Down
front of the method
then
Arrow Right
AEP or
first letter
then
Arrow Right
Mouse Keyboard
Double-click on the examina- Arrow Up/Down
tion definition
or
FAEP alternating
first letter, then ENTER
Electrode map
The electrode map is displayed automatically (if activated in the setup). Connect the
electrodes to the bus cable as shown on the electrode map and plug the bus cable
into the circular connector on:
• channel 1 in the case of 2-channel devices (NEUROWERK EMG2)
• channel 2 in the case of 4-channel devices (NEUROWERK EMG4)
F NEUROWERK EMG
Page 92 Method selection
Mouse Keyboard
ENTER
Note: Correct the electrodes if the contact resistance exceeds 10 kOhm and
L the graph block is displayed red. If the impedance is less than 10 kOhm, the
graph block is green. The absolute impedance values should be as low as
possible. Ensure also that the values of the active (A) and reference elec-
trodes (R) are approximately the same.
Low impedance values are an important prerequisite for a good signal qual-
ity.
NEUROWERK EMG F
Method selection Page 93
Hearing thresholds
You are taken automatically to the determination of the hearing threshold.
With the aid of the threshold determination function, you can determine the patient's
individual subjective hearing threshold for the left and right ears.
Mouse Keyboard
Hold left mouse button Tab switches between
pressed and shift the slide bar
left and right ear
Arrow Up/Down changes the
volume
The current volume is displayed in dB SPL under the headings "Left" and "Right".
Gradually increase the acoustic stimulation until the patient is able to hear the stimu-
lus. In this way, you can determine the individual hearing threshold for the patient –
separately for left and right.
F NEUROWERK EMG
Page 94 Method selection
Mouse Keyboard
ENTER
Note: The individual values from the hearing threshold determination are
L
saved automatically. The standard examination definition for AEP contains
the specification that the stimulation is to be performed at 70 dB above
threshold for the actual examination. In other words, the stimulation vol-
ume during the actual measurement is 70 dB above the individually deter-
mined hearing threshold.
Since the maximum volume is limited to 100 dB SPL for safety reasons, the
device reduces the stimulation volume automatically once the total of hear-
ing threshold and stimulation volume exceeds this value.
Start measurement to commence the actual examination.
Start averaging
Start averaging
Mouse Keyboard
"F7"
After completion of 2,000 accepted cycles, the active track changes to tracing on the
left; the stimulation side of the headphones and the amplifier inputs are switched ac-
cordingly.
The procedure repeats itself until measurement is stopped automatically at the end of
the left channel after similarly 2,000 accepted cycles.
Manual operation
Next sequence
Next sequence
Start measurement
Start measurement
Save
Save
Reset averaging
To restart averaging from the beginning, press the button shown below during aver-
aging:
Reset averaging
Continue averaging
If you have interrupted the averaging and you are in the STOP mode, you can continue
the averaging by pressing:
Continue averaging
Writing a report
You can enter a text with report pertaining to the current measurement.
Writing a report
Print
You can print out the results of an examination with:
F.4.2 P300
This examination program is preconfigured for the recording of an event-related AEP.
The patient will hear a rare target stimulus (high tone) and a frequent standard stimu-
lus (low tone). The patient will be asked to concentrate his attention on the high tone
(e.g. by counting, pressing the placebo button).
Patient selection
The patient selection window is displayed automatically when the program is started.
Patient selection
Mouse Keyboard
"F2"
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
Method selection
Mouse Keyboard
"F3"
Selecting a method
The method selection window is displayed. Here, you must first select the required
method.
NEUROWERK EMG F
Method selection Page 99
Method selection
Mouse Keyboard
Click on the "plus" sign in Arrow Up/Down
front of the method
then
Arrow Right
AEP or
first letter
then
Arrow Right
Mouse Keyboard
Double-click on the examina- Arrow Up/Down
tion definition
or
P300
first letter
then ENTER
Placing the electrodes: The ear electrodes are connected to each other to
serve as the reference electrodes.
Position the active electrode on "Pz".
Connect the electrodes to channel 1. The software will detect whether the standard or
the target stimulus has been offered and will assign the result to the corresponding
track.
Note: Correct the electrodes if the contact resistance exceeds 10 kOhm and
L the graph block is displayed red. If the impedance is less than 10 kOhm, the
graph block is green. The absolute impedance values should be as low as
possible. Ensure also that the values of the active (A) and reference elec-
trodes (R) are approximately the same.
Low impedance values are an important prerequisite for a good signal qual-
ity.
Start averaging
Mouse Keyboard
"F7"
The averaged signal is built up until the target stimulus reaches the preset number of
cycles (default: 40 cycles).
Manual operation
Next sequence
Next sequence
Start measurement
Start measurement
Save
Reset averaging
To return averaging to the beginning again, restart measurement with
Reset averaging
Report
You can enter a text with report pertaining to the current measurement.
Report
Print
You can print out the results of an examination with:
F.5.1 VEP
This examination program is preconfigured to record the visually evoked potentials at
a cortical trace point following visual stimulation of the right and left eyes, first with
stimulation on the right and with the left eye covered.
Position the patient or the stimulation monitor such that the distance between the pa-
tient's eyes and the monitor is approx. 1.40 m (VEP 60’, 19" TFT display).
The distance depends both on the dimensions of the monitor and on the desired view-
ing angle when viewing the pattern. The distance must be calculated using the for-
mula below depending on the monitor type used,
Edge _ length _ of _ box _(cm)
Dist _(cm) =
tan α
where the following applies for VEP 60’ with a 19" TFT display:
tanα=tan60’=tan1°=0.017455
Edge length=2.4cm
For the examination, the patient is to focus on the fixation point on the monitor. The
examination room must be darkened. The examination should be performed with any
visual aids. It is recommended to determine the patient's acuteness of vision (visus).
Patient selection
The patient selection window is displayed automatically when the program is started.
Patient selection
Mouse Keyboard
"F2"
F NEUROWERK EMG
Page 106 Method selection
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
Selecting a method
Mouse Keyboard
"F3"
Selecting a method
The method selection window is displayed. First select the required method.
NEUROWERK EMG F
Method selection Page 107
Selecting a method
Mouse Keyboard
Double-click on the method Arrow Up/Down
VEP then
Arrow Right
or
first letter
then
Arrow Right
Mouse Keyboard
Double-click on the method Arrow Up/Down
VEP60’ or
first letter, then ENTER
Electrode map
The electrode map is displayed automatically (if activated in the setup).
Position the electrodes so that, after stimulation, the trace is taken on the left.
Mouse Keyboard
ENTER
Note: Correct the electrodes if the contact resistance exceeds 10 kOhm and
L the graph block is displayed red. If the impedance is less than 10 kOhm, the
graph block is green. The absolute impedance values should be as low as
possible. Ensure also that the values of the active (A) and reference elec-
trodes (R) are approximately the same.
Low impedance values are an important prerequisite for a good signal qual-
ity.
Mouse Keyboard
ENTER
Start averaging
Start averaging
Mouse Keyboard
"F7"
After completion of 100 accepted cycles, the active track changes to tracing on the left.
Now switch the eye cover to the right eye.
Starting measurement
Mouse Keyboard
"F7"
The procedure repeats itself until measurement is stopped automatically at the end of
the left reference track after similarly 100 accepted cycles.
Manual operation
Next sequence
Next sequence
Start measurement
Start measurement
Reset averaging
To restart averaging from the beginning, press the button shown below during aver-
aging:
Reset averaging
Continue averaging
If you have interrupted the averaging and you are in the STOP mode, you can continue
the averaging by pressing:
Continue averaging
Report
You can enter a text with report pertaining to the current measurement.
Report
Print
You can print out the results of an examination with:
Patient selection
Mouse Keyboard
"F2"
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
Method selection
Mouse Keyboard
"F3"
F NEUROWERK EMG
Page 114 Method selection
Selecting a method
The method selection window is displayed. First select the required method.
Method selection
Mouse Keyboard
Click on the "plus" sign in Arrow Up/Down
front of the method
then
EMG
Arrow Right
or
first letter
then
Arrow Right
Mouse Keyboard
Double-click on the method Arrow Up/Down
Needle EMG trigger or
first letter, then ENTER
1. Spontaneous activity
The first examination sequence (spontaneous activity) is started in the "MONITOR"
mode. In monitor mode, the EMG is traced continuously.
In the info bar, you can see the letters A, R and Z. The info bar also displays the selected
examination sequence (spontaneous, voluntary, maximal) under "R". This should not
be modified.
NEUROWERK EMG F
Method selection Page 115
Spontaneous activity
Mouse Keyboard
"A"
Mouse Keyboard
Left-click on the muscle name. Arrow Up/Down
or
first letter
then
Arrow Up/Down
Mouse Keyboard
Left-click on Key "F11"
or
key "F12"
Now you can perform the tracing for the spontaneous activity. It is possible to save
several measurements (sequences). To this end, the measurement must first be
stopped:
Now you can decide whether you wish to save the displayed measurement or discard
your changes without saving.
Several measurements (sequences) can be saved by repeating the steps "Stop meas-
urement" and "Save measurement" as often as necessary.
The curves are displayed as a cascade (5 tracks one beneath the other) when tracing
for the spontaneous activity is performed. This can be switched using the key combi-
nation [CTRL]+X while the measurement is running.
NEUROWERK EMG F
Method selection Page 117
Switch
2. Voluntary activity
The examination sequence is started in the "Monitor" mode. The EMG is recorded con-
tinuously.
Trigger
Thanks to an adjustable amplitude trigger, it is possible to display individual muscle
action potentials in the centre of the screen and to accept them for an analysis.
Trigger
Mouse Keyboard
Position the mouse pointer Select parameter
over the "Trigger" field.
Arrow Left/Right
Over the upper half, the
Change parameter
mouse pointer shows a
Arrow Up/Down
" + " symbol
The trigger window is dis-
abled when the left mouse
button is pressed.
Operator controls
Mouse Keyboard
Position the mouse pointer Select parameter
over the "Trigger range" field.
Arrow Left/Right
In the upper half, the mouse
Change parameter
pointer shows a
Arrow Up/Down
"+" symbol.
The trigger window is maxi-
mised when the left mouse
button is pressed.
In the lower half, the mouse
pointer shows a
" – " symbol.
The trigger window is mini-
mised when the left mouse
button is pressed.
As already described with the spontaneous activity, you have the following options:
NEUROWERK EMG F
Method selection Page 119
Several measurements (sequences) can be saved by repeating the steps "Stop meas-
urement" and "Save measurement" as often as necessary.
Switching
3. Maximal innervation
The examination sequence is started in the "Monitor" mode. The EMG is recorded con-
tinuously.
Contrary to the measurement of the spontaneous and light voluntary activity, the time
base is set to 100ms/Div, and the sensitivity to 1mV/Div.
F NEUROWERK EMG
Page 120 Method selection
The trace screen now displays 1 second of maximal innervation. In the results table,
the numerical values of the turns-amplitude analysis are listed to accompany the
graphic representation of the analysis.
Several measurements (sequences) can be saved by repeating the steps "Stop meas-
urement" and "Save measurement" as often as necessary.
Reports
Mouse Keyboard
---
The properties are highlighted when clicked. A new notebook is created for each mus-
cle examined. Complete your input with OK.
Switching
End measurement
Important!!
To be able to complete and - if desired - print the measurement, it is IMPERATIVE to
save the examination.
F NEUROWERK EMG
Page 122 Method selection
End measurement
When you quit the EMG examination, whether by selecting a different examination or
a different patient, the window shown below is displayed automatically.
Print
You can print out the results of an examination with:
After you have selected both the muscle and the appropriate mode, you will see the
EMG curves.
With the voluntary activity, you can view the corresponding raw data by left-clicking
on
A) sequence;
B) the results line in the table;
C) the motoric unit.
Cursors
Check the automatically set cursors for start, end, maximum and minimum in the "MU"
window and change them as necessary with:
Set cursors
Left-click on the appropriate MUAP in the left MUAP window; then click on the right-
hand mouse button.
Now you can select with a mouse click:
• Delete MUAP (removes the active MUAP from the MU);
• Move MUAP to MU (moves the MUAP to another existing MU);
• MUAP into new MU (moves the MUAP to a new MU generated by the system auto-
matically)
Save
Upon completion of your changes or after a manual analysis, save your examinations
with:
Save
F.6.2.1 Tremor
The patient selection window is displayed automatically when the program is started.
Otherwise, you can switch to patient selection with:
Patient selection
Mouse Keyboard
"F2"
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
Patient selection
Mouse Keyboard
"F3"
NEUROWERK EMG F
Method selection Page 127
Selecting a method
The method selection window is displayed. First select the required method.
Method selection
Mouse Keyboard
Click on the "plus" sign in Arrow Up/Down
front of the method
then
EMG Arrow Right
or
first letter
then
Arrow Right
Mouse Keyboard
Double-click on the method Arrow Up/Down
Tremor or
first letter, then ENTER
Mouse Keyboard
"A"
F NEUROWERK EMG
Page 128 Method selection
Then select the 2nd track and also enter a muscle name here.
Start measurement
Start measurement
Print
You can print out the results of an examination with:
F.7.1 OOR
Patient selection
The patient selection window is displayed automatically when the program is started.
Otherwise, you can switch to patient selection with:
Patient selection
Mouse Keyboard
"F2"
After you have selected a patient (see Section B "Patient Database"), activate the
method selection if it is still hidden.
Method selection
Mouse Keyboard
"F3"
Selecting a method
The method selection window is displayed. Here, you must first select the required
method.
NEUROWERK EMG F
Method selection Page 131
Method selection
Mouse Keyboard
Click on the "plus" sign in Arrow Up/Down
front of the method
then
Arrow Right
Reflexes or
first letter
then
Arrow Right
Mouse Keyboard
Double-click on the examina- Arrow Up/Down
tion definition
or
OOR
first letter
then ENTER
Trigger
The first examination sequence is started in the "Trigger" mode.
Check the stimulator parameters (current, pulse width) and modify, if necessary, in the
control bar:
F NEUROWERK EMG
Page 132 Method selection
Example: Current
Mouse Keyboard
Position the mouse pointer Select parameter
over the "Current" parameter
Arrow Left/Right
Over the upper half, the
Change parameter
mouse pointer shows a
Arrow Up/Down
" + " symbol
Use the left mouse button to
increase the current.
Over the lower half, the
mouse pointer shows a
" – " symbol
Use the left mouse button to
reduce the current.
Start measurement
Next track
Next track
Now repeat the measuring process for the next tracks to stimulate the left-hand N.
trigeminus.
Print
You can print out the results of an examination with:
G Technical Data
Acoustic stimulator:
Stimulation mode: Click, click-clack; stimulation duration can be set from
50 µs to 1 ms
H Annex
Neurowerk EMG is intended for use in environments as specified below. The customer or the user of Neu-
rowerk EMG should ensure that the device is only used in such an environment.
Neurowerk EMG is intended for use in environments as specified below. The customer or the user of Neurowerk
EMG should ensure that the device is only used in such an environment.
Emitted HF interference +- 2 kV for mains lines +- 2 kV for mains lines The quality of the supplying volt-
to CISPR 11 age should comply with the stan-
+- 1 kV for I/O lines +- 1 kV for I/O lines dards for a typical business or
hospital environment.
Surge voltages to IEC +- 1 kV for normal-mode +- 1 kV for normal-mode The quality of the supplying volt-
61000-4-5 voltages voltages age should comply with the stan-
dards for a typical business or
+- 2 kV for common- +- 2 kV for common- hospital environment.
mode voltages mode voltages
Voltage dips, short-time <5 % for 5s <5 % for 5s The quality of the supplying volt-
interruptions and fluctua- age should comply with the stan-
tions of the supplying <5 % for 0.5 periods <5 % for 0.5 periods dards for a typical business or
voltage to IEC 61000-4-11 <40 % for 5 periods <40 % for 5 periods hospital environment. If the Neu-
rowerk EMG user also requires
<70 % for 25 periods <70 % for 25 periods continued functioning if the power
supply is interrupted, it is recom-
mended to power the Neurowerk
EMG from an uninterruptible
power supply or a battery.
Magnetic field at the 3 A/m 3 A/m Magnetic fields at the mains fre-
supply frequency quency should comply with the
(50/60Hz) to IEC 61000-4- typical values as to be found in
8 business or hospital environments.
NEUROWERK EMG H
Annex Page 141
Neurowerk EMG is intended for use in environments as specified below. The customer or the user of Neurowerk
EMG should ensure that the device is only used in such an environment.
REMARK 1: The higher frequency range applies for 80 MHz and 800MHz.
REMARK 2: These guidelines may not apply in all cases. The propagation of electro-
magnetic interference is influenced by the absorption and reflexion of buildings, ob-
jects and humans.
The field strength of stationary transmitters, such as base stations of radio telephones
and mobile farming radio systems, amateur radio transmitters, AM and FM radio and
TV transmitters cannot be predetermined exactly theoretically.
To be able to determine the electromagnetic environment in respect of the stationary
transmitters, the user is recommended to make a study of the location. If the measured
field strength at the location at which Neurowerk EMG is used, exceeds the compli-
ance levels specified above, the Neurowerk EMG should be observed to verify its use as
intended. If abnormal functioning is observed, additional measures may be required,
such as a changed alignment or a different location of the Neurowerk EMG
The field strength in the frequency range between 150kHz and 80 MHz should be less
than 3 V/m.
H NEUROWERK EMG
Page 142 Annex
Neurowerk EMG is intended for use in an electromagnetic environment with controlled HF interference. The
customer or the user of the Neurowerk EMG may contribute to the avoidance of electromagnetic interference by
observing a minimum distance between portable and mobile HF telecommunication devices (transmitters) and
the Neurowerk EMG, depending on the output power of the communication device, as specified below.
Nominal power 80 MHz ... 800 800 MHz ... 2.5 150 kHz ... 80
of the transmit- MHz GHz MHz
ter P in W
d= 1.17 * root(P/W) d= 2.33 * root(P/W) d= 1.17 * root(P/W)
For transmitters whose maximum nominal power is not specified in the table above,
the recommended safe distance d in metres (m) can be determined using the equation
given in the appropriate column, where P is the maximum nominal power of the
transmitter manufacturer.
REMARK 1: The higher frequency range applies for 80 MHz and 800MHz.
REMARK 2: These guidelines may not apply in all cases. The propagation of electro-
magnetic interference is influenced by the absorption and reflexions of buildings, ob-
jects and humans.
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and
in the frequency range between 80 MHz and 2.5 GHz are intended to reduce the prob-
ability of mobile/portable communication equipment causing interference if it is inad-
vertently brought into the vicinity of the patient. For this reason, the additional factor
10/3 is applied for calculation of the recommended safe distances in these frequency
ranges.
These guidelines may not apply in all cases. The propagation of electromagnetic inter-
ference is influenced by the absorption and reflexion of buildings, objects and hu-
mans.
NEUROWERK EMG H
Annex Page 143
H.2 Cables/lines/electrodes
Central unit connection cables
PC connection cables