Professional Documents
Culture Documents
www.msf.org
Twitter: @MSF_HIV
www.msfaccess.org/makingviralloadroutine
2016
CONTENTS
Abbreviations 3
Foreword 5
Executive Summary 6
Introduction 8
Sample Types 11
Infrastructure requirements 24
Human resources 24
Waste management 26
HR human resources
QA quality assurance
TB tuberculosis
VL viral load
Abbott Molecular (Des Plaines, Illinois, USA): Instruments: m2000sp and m2000rt; Assay: Abbott
Abbott VL
RealTime HIV-1
Biocentric (Bandol, France): Instruments: Nordiag Arrow and Fluorocycler (Bio-Rad realtime
Biocentric VL
thermocycler CFX96); Assay: Generic HIV Charge Virale
BioMérieux (Marcy L'Étoile, France): Instruments: NucliSENS EasyMAG and NucliSENS EasyQ;
BioMérieux VL
Assay: NucliSENS EasyQ HIV-1 V2.0
Diagnostics for the Real World (Little Chesterford, UK): Instruments: SAMBAprep and
SAMBA I
SAMBAamp; Assay: SAMBA HIV-1 Semi Q Test
Xpert HIV-1 VL Cepheid (Sunnyvale, California, USA): Instrument: Genexpert IV; Assay: Xpert HIV-1 Viral Load
it may be possible to use our GeneXpert machine to test for VL. If it's
possible here it's possible anywhere.
Dr Vicente Descalzo Jorro, Medical Doctor. Yambio, South Sudan
*
LESSONS LEARNED
• Scale up of HIV viral load testing is feasible in • Strengthening of laboratory data quality through
resource-limited settings laboratory information systems (LIS) coupled with
innovative result delivery systems is key in scaling up
• The choice of platform must be context-specific. It
HIV VL testing
should take into account the sample type, volume of
samples expected, testing needs beyond HIV-VL (e.g. • In country maintenance and servicing support must
EID, TB) and the clinical urgency of the test be improved to reduce down-time of testing. Rental
agreement plans may facilitate improved provision of
• If plasma is used in decentralised settings, significant
maintenance
investment in sample transport and/or infrastructure
to centrifuge and store samples at clinic level is • Backup planning (e.g. outsourcing testing to another
needed. In settings where this is not possible, using laboratory) is essential and should be triggered by
DBS or near POC technologies (where throughput, clear criteria (e.g. specified target intra-laboratory
human resources and infrastructure allow) significantly turnaround time exceeded)
facilitated the scale up of viral load testing
• Leasing agreements allow flexibility, may encourage
• Near POC technologies allow task shifting to lower better maintenance and facilitate the future upgrading
cadres to perform sample preparation and testing and of platforms
provided same day results for 80% of patients
• Solutions for waste management have not yet been
• Outsourcing VL testing to established private addressed
laboratories allows VL testing in challenging contexts
• Pooling of procurement across countries supports
where no in-country capacity existed, and is important
price reductions. This should be coordinated by
for back-up planning
international donors
CHIRADZULU, District Hospital Laboratory Diagnostics for the Real World SAMBA HIV-1
MALAWI and 4 primary care clinics
Semi Q Test 26,106
(1 SAMBAprep, 3 SAMBAamps)
Plasma DBS
AND VL PLATFORM
*
LESSONS LEARNED
• Using plasma samples and a centralised platform is
feasible but requires substantial investment in sample
transport and/or additional investment in infrastructure
(cold chain) and human resources (HR) at clinic level to
centrifuge and store samples.
• Is phlebotomy feasible? If phlebotomy is not feasible, At the start of this project, only one platform was approved
finger-prick DBS may be used and a platform for DBS samples (BioMérieux VL) and no near or true
accommodating DBS selected POC technologies were available on the commercial
market. In Thyolo, Malawi, and all sites in Zimbabwe and
• Is sample transport for plasma feasible, without making
Mozambique, collection of plasma samples from all sites
patients attend on additional days for sample collection
was not feasible. Hence DBS was selected as the sample
(is increased transport frequency or clinic centrifugation
type, limiting the choice of VL platform to BioMérieux until
and refrigeration possible?) If not, consider DBS or
the end of 2015.
plasma with near POC
• How many sites send VL and what is their forecasted In Chiradzulu, Malawi, and Arua, Uganda, SAMBA I was
weekly throughput by site and as a collective? Which selected to demonstrate feasibility of this new VL near
platforms or combination of platforms (knowing that POC technology, tested with close cooperation with the
more than one near POC machine can be aligned to manufacturers to demonstrate sensitivity, specificity and
increase throughput) may meet demand? ease of use in field conditions. Biocentric was chosen
• Considering both centralised and near POC, what in Swaziland because collection of plasma samples was
infrastructure/space and HR requirements will be needed feasible, the platform could be used at a district laboratory
to meet demand? and it provided experience in using this “open source”
platform.
• What and how many other sample types are sent
from the area served that may also be run on a given
Abbott has since developed an improved technique with
platform? For example, EID and TB samples
better sensitivities and specificities around the 1000
• Are all samples of equal clinical priority? copies/ml threshold for tests processed from DBS samples,
» VL in a clinically well versus VL in a clinically and was recently CE marked. In 2016, with increasing
failing patient – a near POC platform using plasma choice of platforms, the strategic use of near POC and
samples positioned at the hospital level may centralised high-throughput options requires further
be the best choice for clinically unwell patients feasibility studies and may benefit from modelling analysis.
» EID in a child whose mother has a suppressed However, context is key and each setting must assess the
VL versus EID in a child whose mother is not above elements. One solution will not fit all.
on ART – near POC technologies for EID use
Guilio Donini
DIAGNOSTIC FOR THE REAL WORLD, SAMBA I BIOMÉRIEUX, NUCLISENS EASYMAG & EASYQ
BioMérieux NucliSENS was the first platform approved for
We designed a floor plan for a mini lab in each clinic that
DBS testing using a 1000 copies/ml threshold, so we have
could be copied for all sites. We had enough room for one
quite a lot of experience using this platform. Space wise, it
SAMBAprep and three SAMBAamps. After renovations were
requires a relatively small bench-top size and we did not need
carried out, installation of the equipment just took half a day.
to have double doors fitted into the laboratories. With two
Training lasted 3 days and most of the operators reported that
easyMAGs and one easyQ, the throughput is higher than with
they felt confident after processing 4–5 runs. What was great
the other platforms but could be better if the easyMAG could
is that we managed to task shift using SAMBA I to lay workers
extract more than 24 samples. Also we can store the eluates
recruited from the local village; it hopefully will help with staff
longer (>24 hours), which is convenient. The QC system is
retention, which is difficult with the highly trained laboratory
also very convenient as each specimen has an internal control.
technicians. The other good thing is that we managed to give
However if the negative and/or positive control fail, the whole
80% of the patients their result on the same day, and that
batch of tests must be repeated.
most patients were willing to wait for their results.
However, there are a lot of manual steps, meaning there
We did have some technical challenges along the way, which is more workload for the laboratory staff. Also, there is no
were addressed by the manufacturer (changing the shape self-contained decontamination system; this would be really
of the adaptor to reduce interruptions in processing and the beneficial. One other thing that takes additional time is having
problem of cartridge sealing). At present there is also limited to manually calculate the correction factor for the haematocrit;
manufacturer capacity with an official lead time for equipment why can’t this be built within the system?
and test kits of 4 months (although this can be halved on
The maintenance procedures for the user (daily, weekly,
occasions) and internal challenges for importation. However,
monthly) are also quite time consuming. Maintenance has
ordering is very simple because all consumables are supplied
been a challenge but has definitely improved over time as the
with the kits, which can be stored up to 37°C.
local engineers have gained experience.
At the end of the day though, the laboratory is very reliant on (Laboratory technician, National Medical Reference Laboratory,
how well the patient flow is organised. If phlebotomy is done Harare, Zimbabwe)
late or the clinician is absent in the afternoon and cannot
release the results on the same day, the benefit of POC testing
disappears.
YES
SCENARIO 01
• There are 2 hospitals and 7 primary care clinics (PHC). 1 hospital and 1 PHC (total cohort 4000)
are near to each other and have difficult transport access for three months of the year due to
flooding. Daily VL requirements could be met for these two sites with a near point of
care technology.
Centralised Near POC
• Throughput at remaining sites is high, and there is no additional funding available to enhance
high throughput
30% sample transport to more than once per week. DBS will be used for routine viral load testing from
the remaining sites.
70%
• At the hospital a near point of care platform is available that could be used for VL testing for
inpatients suspected of treatment failure.
SCENARIO 03
• District population 220,000, HIV prevalence 6%.
SCENARIO 04
• This hospital ART clinic is in a low prevalence (2%) urban setting with an ART cohort of
2,100 patients.
Near POC • Staff can perform phlebotomy.
• Many complicated patients are referred to this site and the daily throughput required for VL testing
100% is 8 VL tests per day. TB testing is also performed on site.
• Near POC is chosen due to the clinical urgency of testing for the majority of patients in this site.
Further modules may be added if demand increases.
Laboratory setting Specifications of platform Sample transport system Setting up the laboratory Quality assurance Maintenance Testing capacity and performance
Approx time from New building or In-country Maximum capacity Annual
Setting of Platform Regulatory Sample Cadre able to Size of the Maximum Sample Annual capacity in Capacity used Error rate
Date Number of ART Level of Frequency of platform order adaptation of technical Most common Backup in 8-hour shift with capacity Hours worked VL performed
Site viral load VL platform conformation Test approved type used perform the machine throughput: Polyvalency Training transport Power supply Water Storage HR LIS Result delivery format Internal QC External QC Response time specified working in working hours (%) Yr
established sites served automation sample transport to functioning existing laboratory support malfunction and its cause planning current number of (240 work /day Yr 1-3
platform in projects sample type in project test (WxHxD cm) 8-hour day system hours 2015 (%) 1-3
laboratory space available platforms ing days)
1.99; 0.54;
Daily testing Failure to reach tem 0.44
All consumables are 2 Laboratory technicians,
2 extraction on site; twice Generator as a back perature (1 SAMBAprep Online assis higher error
Uganda, Arua
Regional hospital stored in a dedicated 1 laboratory clerk, Research Electronic Data DRW
Provincial +5 MSF car/ weekly sample Adaptation of up for power cuts, Capture (REDCap University CDC PT replaced after 2.5 years), tance within 24 2061; 6466; rate in Yr
Sep-13 and 2 nearby Plasma 6-12 mths Water tank container in the hospital 1 phlebotomist, DRW QC panels every 2 months. technician is 80 19 200 7 16 800 41%
hospital amplification driver transport for existing building plus dedicated UPS of Vanderbilt, USA). REDCap programme electrical contact defect hours (site visit 6864 1 related
health centres compound, plus stock in 1 data entry clerk Continuous QC of 5% of samples Ugandan
modules SAMBA Minimal peripheral for each instrument is a web-based open source (1 SAMBAamp replaced 7-14 days) to sealing
Plasma MSF Kampala pharmacy tested with a conventional
I HIV-1 training Training clinics software allowing minor after 2 years) defect on
(WHO Using 1 Results are released as hard platform. Plasma storage: -80°C cartridge
SemiQ for lab SAMBAprep VL provided modifcations by users. It has
pre-quali SAMBAprep copies and REDCap can pro freezer. IC band checked with
Plasma Semi-auto technicians; 60x60x50 semi-quanti and included Daily for sample been used off-line in Uganda.
SAMBA 1 fication in and 3 duce regular result summary every test. Checks carried out: District
Test. Gives mated lay workers SAMBAamp tative (1000 in initial transport of REDCap is provided free but
District process; Test SAMBAamps: lists. High VLs are manually reagent expiry date, individual hospital: 1100;
District more/less possible 40x32x14.5 copies), EID equipment targeted viral Frequent power cuts; installation and maintenance Sent to District District hospital District hospital:
CE-marked 48 tests/day flagged on the result form checking of cartridges for buffer
Malawi, Chiradzulu
hospital and than 1000 after 10 day cost Storage space limited at require IT engineer. SAMBA is 5218; 6027
hospital and District hospital March 2016) load samples Independant UPS 8 Water tanks District hospital 2 lab levels and cartridge integrity. The high staff turnover Online assis neighbouring hospital 8 hrs; 11520; 52%
each clinic: copies/ml training <6 mths in Adaptation of district hospital, capacity not integrated to REDCap with DRW PHC 1: 491;
4 primary and 4 primary care MSF car/ from the hours often not suffi installed for techs and 1 clerk; Dedicated QC manager and SOP CDC PT significantly contributed tance within 24 clinic with PHC 1 5hrs: PHC1 7200 PHC1: 40% 0.44; 0.21;
Sep-13 1 extraction Plasma primary care existing clinic for 250 tests so regular all data entered manually. 580 technician is 48 in 1 site 11 520 2040; 2884
care clinics clinics in rural driver peripheral clinics cient; plans to install all primary Each PHC 1 lab tech and 1 for QC processes programme to frequent SAMBAprep hours (site visit SAMBA; only PHC 2 6hrs; PHC2 8640 PHC2: 32% 0.26
with 3 clinics space restock from central results are released per month Ugandan PHC 2:
in rural district not equipped UPS with capacity for care clinics clerk; none had experience processing delays 7-14 days) one breakdown PHC 3 6hrs; PHC3 8640 PHC3: 21%
amplification pharmacy required in Uganda 1478; 2782
district with SAMBA 48 hours in 2015 PHC 4 5.5hrs PHC4 7920 PHC4: 25%
modules PHC 3: 190; 1835
to the district
PHC 4: 124; 1937
laboratory
District
hospital lab
Plasma
oratory with Low prevalence
Congo, Kinshasha
DBS: WHO
MSF HIV cohort at one extractor Training setting - device
Referral Abbott pre-quali Sp EID, HCV
central CHK device (Abbott Semi-auto Laboratory provided in is used once/
hospital in Jun-15 Abbott VL realtime fication in Plasma 179x187x124; 93 VL, HCV On site Not applicable Yes 2015: 2364
Hospital in m2000sp) mated technician country by week for HIV VL
capital HIV-1 process; Test Rt 34x49x45 genotyping
Kinshasa and one manufacturer and once/week
CE marked
amplification for EID
2016
device (Abbott
m2000rt)
1 data clerk is responsible for
Adaptation:
Challenge Laboratory Information Manage data entry. Results are issued
Malawi, Thyolo
BioMérieux VL Motorcycle BioMérieux VL Staff redirected to VL testing: Humidity checked for DBS
Thyolo district Frequent power cuts; during set up; Limited: additional ment Systems (LIMS CHAI). only as hard copy. LIMS 8 hours
Rural district (changed to 1 easyMAG; provided by machines are 3 trained for BM; a further 2 based on the dessicants and the CDC pro- Damage to amplifier 9619; 15,416;
Mid-2011 (28 sites); Nsanje FP DBS Once weekly 12-18 mths end 2014 very bad power cuts storage is provided at VL results from the equipment prints VL >1000 copies/ Yes 2-3 days 140 33 600 (1 easyMAG: 33 600 44%
hospital Abbott VL end 1 easyQ Riders for accommodated for Abbott; none with humidity indicator card. QA SOPs gramme cable (m2000rt) 14,761
District (14 sites) due to heavy rainfall affect water MSF office are uploaded into LIMS using ml in red but availability of 1 easyQ)
2015) Health in a very small experience in place
pump a USB port colour cartridges has proved
space
a challenge
Individual results are
delivered by hard copy
Samples rejected due to poor
twice weekly from national
quality or sample identification or
laboratory to district sites.
Mozambique, Maputo
(28 clinics), Gutu samples and MSF) or identically to the format of (indicated on the VL request Pump failure, liquid response time 8 or 16 hours
tained using a hygrometer. NMRL Send samples
National district (19), uses existing Riders for Occasional power Challenge in the VL request form. Samples form) an SMS is sent to the Yes, but need level sensor failure, lamp from local rep 67 200; 2 according to 10,055; 35,439;
2 easyMAGs; Limited: 2 weeks of 6 Laboratory scientists; 4 aiming for ISO 15189 accredita CDC pro- to private 80% of one
Laboratory Sep-13 Chikomba district Roche plat EDTA DBS Health from Once weekly 6 mths Adaptation cuts but UPS 8 hours initial set up; are identified using bar codes patient with a message saying capacity failure, calibration failure. resentative 3 280 shifts run demand 67 200 or 134 400 53,686 + 4577 9; 15; 3
1 easyQ reagents data clerks tion through SLIMPTA. There is a gramme laboratory in shift
Harare (13); targeted VL form for EID clinic to insufficient now constant placed on specimens when everything is fine or advising building Average 3 break-downs days; generally, 134 400 (2 easyMAGs: (outsourced)
dedicated QC manager. EQA VL South Africa
Nyanga district and plasma district; MSF DBS are spotted. VLIS requires to attend the clinic as soon as in 2015 local rep has 1 easyQ)
samples sent to CDC proficiency
(19), Harare VL where car from Microsoft Access. Four data possible. High VL results are no spare parts
testing scheme
central hospitals feasible district to clerks are employed to encode SMSed to the clinic
capital; plan (approx 4474 results/mth). In
to transfer to pilot site has been double data
DHL entry for research purposes
Installed Manual
mid 2015; sample The GeneXpert
in-country Plasma prepara The district lab is enrolled in pro GX4 has an
Zimbabwe, Gutu
validation Routine VL in Gutu (undergo tion with Minimal Four module TB and rif ficiency testing with the NMRL. average optimal
2 District 4 module Four modular Training No additional staff employed: Feasiblity phase
phase (1 district hospital, Xpert HIV-1 Xpert HIV-1 ing WHO automated laboratory device 24 resistance, Each Xpert HIV-1 Quant reagent daily capacity of
hospitals and device in each Plasma system 27.94 x included in On site Not applicable < 6 mths existing microscopists perform started early
completed; 1 rural hospital VL Quantitative pre-qual testing. training (e.g. (90 mins/ EID, CT/NG, cartridge is a self-contained test 16 samples in an
1 rural clinic setting 30.48 x 29.72 the price tests 2016
feasibility and 5 clinics) ification; <5 mins microscopist) test) HCV, HPV device with internal controls and 8-hour day with
studies CE-marked) hands-on PCR reagents results accessible
starting time during within 90 minutes
2016 sample prep
Manual
MSF-facilitat At hospital: 4 laboratory tech are generated with no flagging extracted and amplified in each
method. Only
Nhlangano Generic ed transporta nologists, 2 phlebotomists, 1 system. Monthly summary run. Regular calibration of the Failures in standard and
Shiselweni district automated Laboratory HIV-2VL, TB, On-site train Biocentric Database. VL plat CDC pro- Not available 17,179; 21,487;
rural District 2012 Biocentric VL HIV charge Plasma Plasma 40x45x46 40 tion/car from Twice weekly Adaptation data clerk; at each PHC: lay- reports can be generated equipment. Annual revision of controls during extraction. 150 36 000 8 36 000 75%
(25 sites) process technician HBV VL. ing offered form is not directly connected gramme locally 26,967
Hospital virale the clinic to worker to perform phlebotomy for each site. 3000 results SOPs. Frequent training. Annual Arrows extractors failing
is RNA
the laboratory and centrifugation are released monthly with 1 maintenance certificates are
extraction
data clerk provided by the manufacturer
and external companies for
smaller devices. The project has
an assigned QC manager
TABLE 2: MAKING VIRAL LOAD ROUTINE: PROJECTS PERFORMING ROUTINE VL USING
NON-MSF-SUPPORTED LABORATORIES
Specifications of
Laboratory setting Sample transport system
platform
Number
VL Sample Frequency
Setting of of ART Sample transport
Site laboratory VL platform type used of sample
project sites system
used in project transport
served
Mumbai, India
Independent collection
Private by the private labora
Urban 1 Clinic Twice daily
laboratory tory, from the clinic by
car/land transportation
Roche VL Plasma
Nairobi, Kenya
2 Primary
MoH VL Twice
Urban care MSF car/driver
programme weekly
clinics
to plasma separation
MoH VL site; EGPAF transports
Rural 32 clinics Abbott VL Plasma Daily
programme the frozen plasma
samples from the
separation site to the
referral laboratory
Samples
Motorbike courier
sent to:
Roma, Lesotho
sent to:
Global
Mobile MSF plane to Juba.
Clinical and BioMérieux Every 2
Rural outreach EDTA DBS DHL to laboratory in
Viral Labora VL weeks
approach South Africa
tory, Durban
South Africa
SETTING UP A VIRAL LOAD
TESTING LABORATORY
*
LESSONS LEARNED
• Leasing allows for the inclusion of maintenance costs » Flagging of high VL results
within the leasing fees, ensuring that maintenance is » Monthly lists by clinic – flagging high VL results
paid for, which is not guaranteed when platforms are » User-friendly search functions for result queries
purchased or donated
QUALITY ASSURANCE
INFRASTRUCTURE REQUIREMENTS
• Transparency on the timing and duration of the quality
• All sites required adaptations taking 6–12 months to assurance processes has had a significant impact on
place the selected VL platform the decisions to place platforms in specific settings
• Near POC technologies were easier to install but still
required infrastructure investment for power, WASTE MANAGEMENT
air-conditioning and water supply
• Waste management demands more attention as VL
• Finding sufficient space (and for some platforms testing is scaled up
the need for separation between extraction and
• Compounds generated by VL testing include guanidine
amplification) was a major challenge and limited the
thiocyanate, a compound defined as acute toxic level 4.
number and choice of devices that could be installed
• It is estimated that 10 mg of cyanide – coming from
• Power cuts were a problem in all sites. Uninterruptable
guanidine thiocyanate – will be present in every kg
power supply (UPS) systems with longer backup or
of waste from the Roche, BioMérieux and Abbott VL
positioning two UPS in series should be recommended
platforms
to avoid down time related to power cuts
• Clear SOPs for adequate incineration processes should be
• Providing sufficient storage space for consumables was
urgently developed and funded
20a challenge PART 2: THE VIRAL LOAD LABORATORY
DO I NEED TO SET UP A VIRAL LOAD LABORATORY?
6 - 12
MONTHS
21
MSF HAS USED OUTSOURCING OF VL TESTING IN FOUR SCENARIOS:
As the primary strategy for routine VL testing where existing As the first step of a phased implementation plan for
platforms were not validated for DBS and there was no national VL scale up (Zimbabwe).
short- to medium-term solution to introduce new platforms
(e.g. Lesotho). In Zimbabwe, VL testing was initiated by outsourcing
testing to a private laboratory in South Africa in 2011.
In Lesotho, existing VL testing platforms (Roche VL) This allowed time to build capacity and create demand
were not validated for DBS samples. To scale up access at clinic level and establish sample transport whilst the
to VL testing in some of the very remote mountainous infrastructure requirements for in-country VL testing
sites, DBS sampling was felt to be more appropriate to were made. Once in-country laboratory capacity was
allow everyday sampling and simplify sample transport. established at the national reference laboratory, a phased
Therefore, DBS samples were outsourced to a high shift of samples was made to in-house testing. (Fig 1).
throughput private laboratory with a validated platform The option for outsourcing remained as a backup.
(BioMérieux) for DBS in South Africa. Results were
received by email within 2-3 weeks, with the local team
printing results for distribution. Tests were charged at VL TESTING IN ZIMBABWE
In Yambio, a rural district in South Sudan recently Fig 1: In country and outsourced VL testing in Zimbabwe
affected by increasing conflict, mobile outreach teams are
providing ART to the population. Through an outsourcing
approach, VL monitoring has also been provided. EDTA
DBS samples are prepared by nurses in the field and
sent via MSF transport to Juba. Samples are then sent
SCENARIO 04: BREAKDOWNS OR STOCKOUTS
This all-inclusive approach also ensures that the local Linked to the ROI, currently the leasing option is only
distributor is well incentivised to have the machine offered (by the manufacturers) in high HIV prevalence
working (they are responsible for the maintenance, and countries/sites with large ART cohorts. In the low
any downtime means less tests sold) prevalence country where a leasing contract was
discussed, throughput was too low to make a leasing deal
feasible either for MSF or at country level
A target price for leasing was set for $15 per test plus All sites required investment in water supply during
$1.2 for the rental and maintenance. initial set-up and most experienced frequent power
Annual tests performed in each site were estimated as cuts. UPS systems installed were reported as often
follows: DRC 8,000; Malawi 30,000; Mozambique 25,000 inadequate to ensure continuous processing of samples
across all sites. Storage for reagents and consumables
Due to the low testing volumes in DRC, a rental agreement
was also raised as an issue with the cold chain storage
plan (RAP) was not considered feasible by the manufacturer.
area needing full rehabilitation in some sites such as
A discounted price was negotiated and a separate
Zimbabwe. Where VL testing is further decentralised,
maintenance agreement was signed with Abbott.
storage volume capacity for cartridges and other
In Malawi and Mozambique, a RAP was negotiated. The consumables should be considered. Table 2 describes
price included: the infrastructure investments made at each site.
1. Reagents
2. Consumables HUMAN RESOURCES
3. Instrument leasing
With the introduction of VL testing, each site must
4. Service and maintenance
allocate sufficient human resources to ensure
5. Shipment to the country throughput. Additional human resources for data
6. Distributor commission entry have also been essential to ensure efficient
7. Custom clearance paid flow of samples and results. All sites had difficulties
identifying and retaining qualified laboratory staff.
8. DBS reagents for the open mode protocol
Infrastructure challenges such as frequent power cuts
9. Delivery and installation to the MSF laboratory and stock-outs of consumables also contributed to staff
demotivation. Centralised platforms with increasing
levels of automation reduce the staffing requirements,
but where DBS is used the elution processes still must
be performed by skilled laboratory technicians.
LABORATORY RUNNING
*
LESSONS LEARNED
POWER CUTS
MAINTENANCE
PROCUREMENT
PROCURING CONSUMABLES
• Increase the available equipment
Procurement and forecasting depends on reliable In Mozambique and Zimbabwe, two easyMAG extraction
consumption and ART programme data. The links between devices were used with one easyQ amplifier doubling the
the laboratory and real-time ART cohort and VL cascade number of samples run in an 8-hour shift. This was not
data (See Part 1) must be strengthened to improve possible in Thyolo, Malawi, due to space limitations. With
forecasting both to avoid stock-outs and wastage. Each near POC technologies, additional modules can be added
reagent should be recorded in terms of the number of tests where space allows. Space again was a limitation for
that can be performed rather than quantities of reagents. adding modules in Chiradzulu district hospital.
The number of staff adequately trained to carry out orders
or conduct stock counts was also noted as a weakness • Increase the hours the machine is running by introduction
(a minimum of two should be fully trained) along with of a shift system
Pooling resulted in a reduction of 29% in HIV VL tests • Strict adherence to manufacturers’ instructions and
performed, to cover the same number of samples tested, laboratory testing SOPs
translating to significant cost savings ($47,110/year)(6). • Equipment maintenance and regular stock audits
(expiries)
Pooling does, however, involve additional hands-on
• Regular proficiency assessment of personnel with
workload, and with decreasing costs of testing and
continuous revisions of laboratory workflow
increasing automation of devices there is debate as to
whether this approach will be efficient in the future. • Use of appropriate controls (internal and external)
Efficiency will also depend on the prevalence of virological
The laboratory should keep a record of all errors and
failure as each positive pool requires re-testing of
document the corrective actions taken. Fig 3 illustrates
individual samples.
error rates for the VL platforms in Mozambique, Zimbabwe
and the two SAMBA 1 sites. Most errors in Zimbabwe
Figure 2 (number of HIV tests performed in the VL
(BioMérieux VL) occurred during a period of delayed
laboratory in Zimbabwe) illustrates three points:
scheduled platform maintenance. The lamp in the platform
1. The importance of having a monthly target for VL testing had ceased functioning and was replaced by a used lamp,
whereas in Mozambique (BioMérieux VL), the high error
2. The impact of introducing a second shift as demand rate was due to power cuts. Error rates for SAMBA 1 were
increased low, including when performed by lay workers. Xpert HIV-1
VL testing has so far been conducted during a diagnostic
3. The ability to trigger outsourcing when, due to
performance evaluation of the index instrument under
downtime related to machine breakdowns, the monthly
ideal laboratory conditions and had a total of error rate of
target was not met
4.3%. However, field performance evaluation in district
laboratories is ongoing.
7000
J
J F J
6000 S O
MONTHLY TARGET N
5000
2ND SHIFT M D
MONTHLY TARGET INTRODUCED J
4000 A
A A
O J M J S
3000 F O N D M
A
N D M
2000 J S
A
1000
Figure 2: Monthly VL tests performed at NMRL Zimbabwe demonstrating introduction of shift system and outsourcing
15 15
% Error rates
3 2 2 2
0.4 0.2 0.3 0.5 0.4
70
60
50
DAYS
40
30
20
10
0
11-2014
12-2014
01-2015
02-2015
03-2015
01-2014
02-2014
03-2014
04-2014
05-2014
06-2014
07-2014
08-2014
09-2014
10-2014
04-2015
05-2015
06-2015
07-2015
08-2015
09-2015
10-2015
PERIOD
10
70
40
20
50
30
60
10
70
40
20
50
30
60
01-2014
01-2014
02-2014
02-2014
TRANSPORT TO LAB
TRANSPORT TO LAB
03-2014
03-2014
04-2014
04-2014
05-2014
05-2014
06-2014
AWAITING TESTING
AWAITING TESTING
06-2014
07-2014
03-2015 04-2015
04-2015 05-2015
05-2015 06-2015
06-2015 07-2015
07-2015 08-2015
08-2015 09-2015
09-2015 10-2015
10-2015 11-2015
11-2015 12-2015
12-2015
PERIOD PERIOD
33
ENSURING QUALITY
FACTORS AFFECTING INTRA-LABORATORY TAT The maintenance of a quality management system
IN NMRL, ZIMBABWE (QMS) is crucial for any laboratory. Important elements
of a QMS include documentation, standard operating
procedures (SOPs) and QC samples. Table 2 outlines the
• Increasing volume of samples received each month QC measures undertaken across the sites. Part of the QA
for a laboratory lies not only in the accuracy of the test
• Machine breakdown with prolonged downtime,
result, but also in the intra-laboratory turnaround time
especially when spare parts are not locally available
of the result, from the time the laboratory receives the
in the country
sample to when the result becomes available for clinical
• Reagents stock-out dispatch to the clinician/clinic. A budget for internal and
external quality assurance programmes must be assured
• Competing priorities for HIV VL testing laboratory
as part of the laboratory scale-up plan. An on-site
technicians. Laboratory managers should be able
dedicated QC manager was found to impact the correct
to determine the ‘man hours’ that each laboratory
implementation of QC processes positively.
technician should spend per day processing HIV VL
samples
An external quality assessment scheme (EQAS) is a
• Organisational and behavioural culture within the programme in which the accuracy of testing within a
laboratory laboratory is ensured through external review. A blinded
• Highly manual transcription/writing/calculations testing panel is sent to the laboratory, the panel is run
which could be done by LIS on the device and the test results are submitted to
the external provider. The external provider compares
the results and generates an accuracy report for the
laboratory. The laboratory uses this process to ensure a
high quality of testing on an ongoing basis. This process
is commonly known as ‘proficiency testing’. Funding for
EQAS must be included in budgets and is required if
laboratories wish to obtain any form of accreditation.
Viral load testing is recognised by the major donors as one As it is unlikely that POC technologies will be available for
of the essential programmatic components to reach the all tests at all sites how do we advocate for a coordinated
third “90” target. Although there is technical agreement approach to sample transport within a national VL testing
to scale-up VL testing, allocating funds to VL testing over scale up plan?
other priorities requires ongoing advocacy.
Ensuring adequate in country maintenance and a
systematic approach to back up planning must be a priority
Within a 3-year period, VL testing capacity for MSF-
to ensure uninterrupted VL testing services.
supported ART cohorts has been significantly increased,
hence demonstrating that VL testing is feasible in such What about the waste? This has not been adequately
contexts. However, this experience has demonstrated the considered, and in the absence of government regulatory
significant resources that were invested to reach where we frameworks potentially may be left unaddressed. Advocacy
are today. Basic infrastructure challenges (power and water is needed towards both manufacturers and donors to
supplies) continue to affect the ability of laboratories to ensure that information and funding are provided to
maintain targets and retain motivated staff. develop improved technological solutions to mitigate
the health risks posed to both laboratory staff and the
WHAT DO WE NEED TO FOCUS ON GOING FORWARD? neighbouring populations.
Due to the delayed availability of near and true POC VL
devices, operational strategies to assess how best to Finally, we must recognise that no one solution will fit all
combine POC technologies with high-throughput platforms settings. Patient’s needs and the realities faced at health
have not been tested. Although placement of platforms will centres providing ART must be central to any VL testing
be context-specific, development of a decision framework scale-up plan. These needs should encourage us to call for
that can be used to guide Ministries of Health on how best further adaptations to VL testing devices in order to truly
to use these technologies is urgently needed. make VL routine in such contexts.