ULTRAPURE WATER SYSTEMS

Crystal E series

The Crystal E ultrapure systems All Crystal E systems produce two


are economy class, multi-purpose, types of water: Ultrapure (ISO 3696
water purification systems. Grade 1) and Pure (ISO 3696 Grade
2). Ultrapure water produced by
Crystal E ultrapure water systems the Crystal E systems has resistiv-
are available in the following ity 18.2 MegaOhm*cm (conduc-
configurations: tivity 0.055 μS/cm). This exceeds
• Crystal E Trace System (P/N CE- requirements of all the relevant
1001) produces water for inor- standards (ISO 3696 Grade 1,
ganic trace analysis. This water ASTM Type I, CLSI Type I). Puri-
is recommended for atomic fied water is collected in a storage
absorption spectrometry (with tank. The recirculation system en-
graphite furnace atomizer), ICP- sures a consistent quality of water,
OES analysis, ICP-MS and other and a low level of organic carbon
inorganic analytical methods. content (TOC). TOC is <2ppb for
• Crystal E HPLC System (P/N CE- “HPLC” and “Bio” configurations; be done by the operator and no
1101) produces water with very and 5-10 ppb for the “Trace” con- tools are required. The Crystal E
low organic carbon content figuration. systems include important safety
(TOC) to comply with the re- The dispensing rate of high-quali- functions, such as:
quirements of liquid chromatog- ty, Ultrapure water is 2 L/min. ttank filling control
raphy methods. Crystal E HPLC ttap water pressure control
water can also be used for some Pure water produced by Crystal E tprotection from a tank sensor
microbiological and molecular systems can be used for labware failure
biology applications. washing, wet chemistry methods,
• Crystal E Bio System (P/N CE- flame spectrophotometers, etc.. Running costs can be decreased
1201) produces water with very Pure water is dispensed directly even more by installing an 8L
low organic and RNase/DNase from the storage tank. The dis- deionization module option (P/N
content, intended for molecular pensing flow rate of Pure water is 10101).
biology, including RNase-sensi- 4 L/min.
tive applications. All systems have a graphic LCD Crystal E systems include:
display that provides a clear water- tBoost pump
quality readout. The display also tPre-filter
provides information about the tRO module
system status. This includes cur- tDeionization module
rent resistivity and remaining pre- tPolishing module
filter service life. The smart deion- t25L storage tank with integrated
ization (DI) module performance Grade 2 dispensing pump
monitoring system provides a tRecirculation system
reduction in running costs. All car- tPoint-of-use Final Filter:
tridge and filter replacement can t0.22 μm microfilter for Crystal
E Trace and HPLC systems
Ordering Information tultrafilter for Crystal E Bio
systems
Model Part number tPhoto-oxidation module for
Crystal E Trace CE-1001 reduction of organics (for HPLC
Crystal E HPLC CE-1101 and Bio configurations)
tUV sterilization module (for the
Crystal E Bio CE-1201
Bio configuration)
12
Flow diagrams

Deionization
Input module Recirculation
valve
Boost Shunting
pump Pure water
Reverse quality restrictor Ultrapure water
Strainer osmosis sensor dispense valve
module
Pressure (RO module) Recirculation
switch pump Final point
of-use filter

Input Checkvalve Ultrapure water


valve Flow Polishing
restrictor module dispense

Ultrapure
water quality
Activated Pure water sensor
Protective
carbon/5um dispense filter
pre-filter
Drain Storage
tank

Pure (Grade 2)
water dispense pump Crystal E Trace
Deionization
Input module Recirculation
valve
Boost Shunting
pump Pure water
Reverse quality restrictor Ultrapure water
Strainer osmosis sensor dispense valve
module
Pressure (RO module) Recirculation
switch pump Polishing Final point
module of-use filter

Input Checkvalve Ultrapure water


valve Flow
restrictor dispense

Ultrapure
water quality
Activated Pure water sensor
Protective
carbon/5um dispense filter
pre-filter Dual wavelength
Drain Storage UV photooxidation
tank module (185/254 nm)

Pure (Grade 2)
water dispense pump Crystal E HPLC
Pure water
Deionization quality
module sensor Recirculation
Input
valve
Boost Shunting
pump restrictor
Reverse Ultrapure water
Strainer osmosis dispense valve
module
Pressure (RO module) Recirculation
switch pump Polishing Final point
module of-use filter
UV sterilization
Input Checkvalve module Ultrapure water
valve Flow (254 nm)
restrictor dispense

Ultrapure
water quality
Activated Pure water sensor
Protective
carbon/5um dispense filter
pre-filter Dual wavelength
Drain Storage UV photooxidation
tank module (185/254 nm)

Pure (Grade 2)
water dispense pump
Crystal E Bio
13
Components Applications
Crystal E Crystal E Crystal E Crystal E Crystal E Crystal E
Part Accessory Application Trace HPLC Bio
Trace HPLC Bio
10009
Pre-filter for chlorine and
organics reduction
+ + + Reagent preparation t t t
Boost pump + + + Ion chromatography t t t
Reverse osmosis module + + + ICP-MS t t t
10310
Deionization module,
standard + + + Atomic absorption t t t
High capacity ICP-OES t t t
t t
10101 deionization module, option option option
(replaces standard HPLC –

10029
module)
Polishing module + + +
Gas chromatography – t t
10105 Photo-oxidation module – + + Total Organic Carbon – t t
10012 Point-of-use microfilter + + – Flow cytometry – – t
10109 Point-of-use ultrafiltration
module
– – + Cell and tissue culture – – t
10102 UV sterilization module – – + Molecular biology – – t
Consumables
Part Description Replacement criteria Comments
number
Replacement pre-filter,
10009 Filter life counter is zero or the filter is clogged
Crystal E
Replacement deionization „DI Err” message is shown, or water conductivity
10310 module is consistently > 0.5 μS/cm
Replacement high-capacity „DI Err” message is shown, or water conductivity
10113 Systems with 10101 option only
deionization module is consistently > 0.5 μS/cm
10029 Replacement polishing Every 1–2 years, depending on operation
module
Replacement sterilization
10011 UV bulb
As required (on average every 3 years) ”Bio” systems only

10018 Replacement photo- As required (on average every 3 years) ”HPLC” and “Bio” systems only
oxidation UV bulb
Replacement 0.22 μm
10012 dispense microfilter Every 6–12 months “Trace” and “HPLC” systems

10120 Replacement ultrafilter Every 6–12 months “Bio” systems only

Specifications
Purified water specifications Crystal E Crystal E Crystal E
Trace HPLC Bio
Grade 2 water resistivity >10 MΩ x cm >10 MΩ x cm >10 MΩ x cm
Grade 2 water conductivity <0.1 μS/cm <0.1 μS/cm <0.1 μS/cm
Grade 1 water resistivity 18.2 MΩ x cm 18.2 MΩ x cm 18.2 MΩ x cm
Grade 1 water conductivity 0.055 μS/cm 0.055 μS/cm 0.055 μS/cm
TOC 5 – 10 ppb <2 ppb <2 ppb
RNase N/A N/A <0.01 ng/mL
DNase N/A N/A <4 pg/μL
Bacteria < 1 cfu/mL < 1 cfu/mL < 1 cfu/mL
Endotoxins <0.15 EU/mL <0.15 EU/mL < 0.001 EU/mL
Particles > 0.22 μm <1 per mL <1 per mL <1 per mL
Nominal flow, pure water (to storage tank) 10 L/h 10 L/h 10 L/h
Nominal dispense flow, ultrapure water 2 L/min 2 L/min 2 L/min
Nominal dispense flow, pure water 4 L/min 4 L/min 4 L/min
Deionization module life (standard module) 1 m3 1 m3 1 m3
Deionization module life (high capacity module) 3 m3 3 m3 3 m3
Recovery >30 % >30 % >30 %
Dimensions (WxDxH), cm 40x35x55 40x35x55 40x35x55
Storage tank 25 L 25 L 25 L
Tank dimensions (WxDxH), cm 30x25x50 30x25x50 30x25x50
Feed water pressure 0.5 – 5 bar 0.5 – 5 bar 0.5 – 5 bar
14 Feed water conductivity < 900 μS/cm < 900 μS/cm < 900 μS/cm