You are on page 1of 6

application note Real-Time PCR: Understanding CT

Real-Time PCR: Understanding CT 

4.500 3.500 1000e + 001

4.000 3.000

1000e + 000
3.500 2.500
Threshold
3.000 2.000
1000e - 001
∆Rn

∆Rn
Rn

2500 1500

1000e - 002
2000 1000

1500 0500 Threshold Exponential phase


Baseline 1000e - 003

1000 0000

0500 -0500 1000e - 004


1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40

Cycle Number Cycle Number Cycle Number

Figure 1. A: Rn is the fluorescence of the reporter dye divided by the fluorescence of a passive reference dye. In other words, Rn is the reporter signal normalized
to the fluorescence signal of ROX ™. In this view, Rn is graphed versus cycle. B: ΔRn is Rn minus the baseline, graphed here versus the cycle of PCR. C: Amplification
plot shows the Log (ΔRn) graphed versus cycle.

Introduction Figure 1 shows several parameters two different master mixes. Note that
Real-time PCR, also called quantitative of the real-time reaction amplification the fluorescence intensity is higher in
PCR or qPCR, can provide a simple plot. The exponential phase in Figure Master Mix A even though the target,
and elegant method for determining 1B corresponds to the linear phase in probe and ROX™ concentrations are the
the amount of a target sequence or Figure 1C. The threshold must be set same in both cases.
gene that is present in a sample. Its in the linear phase of the amplification
very simplicity can sometimes lead to plot in Figure 1C. The CT value increases 6.00 E + 3

problems of overlooking some of the with a decreasing amount of template. Master Mix A
5.00 E + 3

critical factors that make it work. This However, anything from the reaction Master Mix B

review will highlight these factors that mix or instrument that changes the 4.00 E + 3

must be considered when setting up and fluorescent measurements associated


Amplitude

3.00 E + 3

evaluating a real-time PCR reaction. with the CT calculation will result in VIC® ROX™

template-independent changes to the CT 2.00 E + 3

Factors that can Influence CT


value. Therefore, the CT values from PCR
CT (threshold cycle) is the intersection 1.00 E + 3

reactions run under different conditions


between an amplification curve and a
or with different reagents cannot be 0.00

threshold line (Figure 1B). It is a relative 0 5 10 15 20 25 30

compared directly. Bin


measure of the concentration of target
in the PCR reaction. Many factors impact The Effect of Master Mix Components Figure 2. Raw fluorescence data obtained with one
assay and two master mixes with the same ROX ™
the absolute value of CT besides the The fluorescent emission of any
level. The difference in signal is due to the master
concentration of the target. We will molecule is dependent on environmental mix composition. Reaction was performed on an
discuss the most common template- factors such as the pH of a solution Applied Biosystems 7900HT Fast Real-Time PCR
independent factors that can influence and salt concentration. Figure 2 System with a VIC ® MGB probe. The X axis shows
the emission wavelength of the fluorophore and
CT and describe how to evaluate the shows the raw fluorescence data of a the Y axis shows the intensity of the emission.
performance of a real-time PCR reaction. TaqMan® probe in the background of
The resulting ΔRn value will, therefore, Rn value assuming FAM fluorescence Efficiency of a PCR Reaction
vary as shown in Figure 3. Note that signal stays the same. This would The efficiency of a PCR reaction can
the baseline fluorescence signals, in lead to an increase in baseline Rn and also affect CT. A dilution series amplified
a template-independent factor, are subsequently a smaller ∆Rn as well as a under low efficiency conditions could
different for the two master mixes different CT value. The different CT value yield a standard curve with a different
(Figure 3A). Variations in CT value do not obtained by lowering the ROX level has slope than one amplified under high
reflect the overall performance of the no bearing on the true sensitivity of the efficiency conditions. In Figure 5, two
reaction system (Figure 3B). Master reaction, but can have other unintended samples (X and Y) amplified under
mixes with equivalent sensitivity may consequences. Low ROX concentration low and high efficiency conditions
have different absolute CT values. can result in increased standard show different CT values for the same
deviation of the CT value, as shown target concentration. In this example,
ROX™ Passive Reference Dye
in Figure 4. The greater the standard although the high efficiency condition
The Rn value is calculated as the ratio
deviation, the lower the confidence in (the blue curve in Figure 5) gives a later
of FAM™ fluorescence divided by the
distinguishing between small differences CT at high concentration, it gives better
ROX fluorescence. Therefore, a lower
in target concentration (see the precision sensitivity at low target concentration.
amount of ROX would produce a higher
section on the next page).

A B
0.866 1.000

0.766
Master Mix A

0.100 Master Mix B


0.666 Baseline A
∆Rn
Rn

Master Mix A
0.566
CTB
0.010

Master Mix B
0.466
CTA
Baseline B

0.366 0.001

1 2 3 3 3 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 1 2 3 3 3 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40

Cycle Number Cycle Number

Figure 3. Master Mix A and Master Mix B were used to amplify RNase P in equal amounts of human gDNA using the Applied Biosystems 7500 Real-Time PCR
System. Figure 3A shows the Rn versus cycle number and the baselines for both reactions. Figure 3B shows the Log (∆Rn) versus cycle number. The threshold
(green) is set at the same level for both master mixes. The CT value of Master Mix B (CT B) is earlier than that of Master Mix A (CTA) for identical concentrations of
target, reflecting the lower baseline of Master Mix B.

14.5
A B
0.30
14.0

0.20
CT (VIC®)

13.5
Std Dev

13.0 0.10

12.5
0.00

0.6x 0.75x 1x 0.6x 0.75x 1x

Rox Concentration

Rox Concentration

Figure 4. Master mixes containing 3 different concentrations of ROX ™ were used to amplify the TGF beta assay on the Applied Biosystems 7900HT Fast
Real-Time PCR System using the 96-well block. Figure 4A shows the CT value and Figure 4B shows the standard deviation with variable ROX concentrations.
Decreasing ROX concentration gives an earlier CT but increases the standard deviation.
40
1 Log
Blue Green

35 Slope -3.3x + 39.8 -4x + 42

Efficiency 100% 78%


168%

30

100%
CT

CT
25

70%

20

5 Logs 108%
X Y
100%
15 92%
1.E + 00 1.E + 01 1.E + 02 1.E + 03 1.E + 04 1.E + 05
Quantity Dilution

Figure 5. The blue standard curve has an efficiency of 100% (slope is -3.3). The green Figure 6. Accurate calculation of PCR efficiency depends
standard curve has an efficiency of 78% (slope is -4). Amplification of the Y quantity gives on the range of template amount used for the serial dilution
an earlier CT with low efficiency condition (green) compared to the high efficiency condition used to calculate the efficiency. With a 2-fold dilution with
(blue). With a lower quantity (X) there is an inversion and the low efficiency condition (green) 5 points (orange), the potential artifact is higher than for the
gives a later CT compared to the high efficiency condition (blue). 10-fold dilution with 5 points (blue).

The PCR efficiency is dependent on suggested level of rigor is illustrated Precision


the assay, the master mix performance in Figure 6, which demonstrates the The standard deviation (square root
and sample quality. Generally speaking, possible mathematical variation of of the variance) is the most common
an efficiency between 90-110% is slope/efficiency one gets when testing measure of precision. If many data
considered acceptable. dilutions over 1 log vs. 5 logs. Thus, points are close to the mean, the
even if the assay is 100% efficient, one standard deviation is small; if many
The observation that the CT value
can get a range from 70-170% when data points are far from the mean,
produced from one sample is higher than
testing a dilution series of a single log, the standard deviation is large.
that of the other, could be valuable in
due to the standard deviation in one
concluding that the amount of template In practice, a data set with a sufficient
dilution. Doing the same number of
is less in the first sample, assuming number of replicates forms an
dilutions/replicates on a 5-log range, the
all other factors such as instruments, approximately normal distribution. This is
potential artifact is only +/- 8 %. That
reagents and assays are equal. However frequently justified by the classic central
means that if we find 94% efficiency
this is not true if different instruments, limit theorem which states that sums of
on a 5-log range, the assay would have
reagents, primers and probes or reaction many independent, identically-distributed
a range of 88% to 100% efficiency. To
volumes are involved in producing the random variables tend towards the
accurately determine the efficiency of
two CTs. Therefore, the absolute CT value normal distribution as a limit. As shown
a PCR reaction, a 5-log dilution series
comparison is only meaningful when in Figure 8A, about 68% of the values are
must be performed. A slope of -3.3
comparing experiments using the same within 1 standard deviation of the mean,
+/- 10% reflects an efficiency of 100%
reaction conditions as defined above. about 95% of the values are within two
+/- 10%. A PCR reaction with lower
standard deviations, and about 99.7% lie
How to Evaluate the Performance efficiency will have lower sensitivity.
within 3 standard deviations.
of a Real-Time PCR Reaction
R2 Value
In order to compare two reactions If a PCR is 100% efficient, there is one
Another critical parameter to evaluating
where a condition is changed (for CT between the mean of a 2-fold dilution
PCR efficiency is R2, which is a statistical
example two different master mixes or (Figure 8B). To be able to quantify a
term that says how good one value is at
two different instruments), the following 2-fold dilution in more than 99.7% of
predicting another. If R2 is 1 then you
parameters must be evaluated. cases, the standard deviation has to
can perfectly predict the value of
be ≤ 0.167. The greater the standard
Dynamic Range X (quantity) with the value of Y (CT)
deviation, the lower the ability to
To properly evaluate PCR efficiency, a (Figure 7A). If R2 is 0, then you cannot
distinguish between 2-fold dilutions.
minimum of 3 replicates and a minimum predict the value of X with the value of Y
To be able to discriminate between a
of 5 logs of template concentration (Figure 7B). An R2 value >0.99 provides
2-fold dilution in more than 95% of
are necessary. The reason for this good confidence in correlating two values.
cases, the standard deviation has to of template would not be expected Conclusion
be ≤ 0.250 (Figure 8C). to be normal. Instead, it would follow These factors – efficiency, R2,
a Poisson distribution which predicts precision, sensitivity – are used to
Sensitivity
that in a large number of replicates determine performance of a PCR
Any system capable of effectively
containing an average of one copy of reaction when comparing different
amplifying and detecting one copy of
starting template, about 37% should reaction conditions. For a rigorous
starting template has achieved the
actually have no copies, only about evaluation, all factors listed in Table 1
ultimate level of sensitivity, regardless
37% should contain one copy, and must be evaluated together.
of the absolute value of the CT.
about 18% should contain two copies
In addition to these factors, proper
As described earlier, efficiency is a key (see Figure 9). Thus, for a reliable low
experimental controls (such as no
factor in determining the sensitivity of copy detection, a large number of
template control, no RT control) and
a reaction (Figure 5). Another important replicates are necessary to provide
template quality must be evaluated
consideration with detecting very low statistical significance and to overcome
and validated.
copy numbers is that the distribution the Poisson distribution limitation.

A 7
B 7
R2=1 R2=0
6 6

5 5
Sample 2

Sample 2
4 4

3 3

2 2

1 1

0 0
0 1 2 3 4 5 6 0 1 2 3 4 5 6 7

Sample 1 Sample 1

Figure 7. Example of R 2 value calculated for 2 straight lines. A: There is a direct relation between x and y values. B: There is no relation between x and y values.

x y
A 99.6% B
95.4%

68.2%

1S 1S 1S 1S 1S 1S 1S

1CT

x y
C

1S 1S 1S 1S

3S 2S 1S 1S 2S 3S 1CT

Figure 8. Normal distribution and standard deviation. (A) shows a normal distribution of data. If the PCR efficiency is 100% there is one CT between the mean of
2-fold dilution samples (sample X and sample Y). To be able to quantify both samples in 99.7% of cases, the standard deviation has to be less than 1 CT divided by
6 standard deviations (1/6=0.167), shown in (B). To be able to quantify both samples in 95% of the case, the standard deviation has to be less than 1 CT divided by
4 standard deviations (1/4=0.25), shown in (C).
TABLE 1. Performance Evaluation of Real-Time PCR
Frequency among 64 Replicates

20
18
16
14
Factors Recommendations Criteria
12
3.3 pg
10
6.6 pg Serial dilution with 5-log Slope~ -3.3
8
Efficiency
6 dilutions R2 > 0.99
4
2
0
Standard deviation
Precision Minimum of 3 replicates
34 35 36 37 38 39 40
< 0.167
CT
High replicate number
of reactions for low copy
Figure 9. Poisson distribution for low copy number. The blue curve represents Sensitivity Statistical test analysis
Poisson distribution for 3.3 pg of DNA (1 copy of DNA). The pink curve
number sample input due to
represents Poisson distribution for 6.6 pg of DNA (1 cell, 2 copies of DNA). Poisson distribution

APPENDIX

Amplification Plot the reporter dye signal can be amplification as well. Specificity
An amplification plot is the plot of normalized during data analysis. should be checked with a melt curve
fluorescence signal versus cycle Normalization is necessary to correct (Power SYBR® Green PCR Master Mix
number. Reactions are characterized for fluorescence fluctuations caused and RT-PCR Protocol, P/N 4367218)
by the point in time during cycling by change in concentration, volume or gel analysis of the PCR product.
when amplification of a PCR product is or sample effects.
Rn
first detected. The higher the starting
PCR Efficiency Normalized reporter is the ratio of the
copy number of the nucleic acid target,
The equations below describe the fluorescence emission intensity of
the sooner a significant increase in
exponential amplification of PCR. the reporter dye to the fluorescence
fluorescence is observed.
emission intensity of the passive
Cn= Ci * (1 + E)n
Baseline reference dye.
Ci = initial copy number
In the initial cycles of PCR there is
Cn = copy number at cycle n Threshold
little change in fluorescence signal.
n = number of cycles A level of ΔRn used for the CT
This defines the baseline for the
E = efficiency of target amplification determination in real-time assays. The
amplification plot. In these cycles we
level is set to be above the baseline
see the fluorescence background of If efficiency is maximum (=1) the
and sufficiently low to be within the
the reaction. This will be subtracted equation is: Cn=Ci * 2n and it means
exponential growth region of the
from the results when setting the that the fold increase will be 2 at each
amplification curve. The threshold
baseline. (For information of how cycle. If the efficiency decreases, the
is the line whose intersection with
to set up the baseline, download quantity of PCR product generated
the amplification plot defines the CT
the document “Applied Biosystems at each cycle will decrease and the
(threshold cycle.) For information on
7300/7500 Real-Time PCR System” amplification plot will be delayed. The
how to set up the threshold, download
PN 4347825 from the Applied recommended efficiency is between
the document “Applied Biosystems
Biosystems website 90 to 110%.
7300/7500 Real-Time PCR System”
www.appliedbiosystems.com) Reporter Dye PN 4347825 from the Applied
Delta Rn Reporter dye is the dye attached to Biosystems website
ΔRn is the normalization of the Rn the 5’ end of the TaqMan® probe. The www.appliedbiosystems.com
obtained by subtracting the baseline dye provides a fluorescence signal
Threshold Cycle (CT)
(ΔRn = Rn - baseline). that indicates specific amplification. If
The fractional cycle number at
SYBR® Green I is used, this dye binds
Passive Reference which the fluorescence passes
double-stranded DNA and the increase
A dye that provides an internal the threshold.
of fluorescence signal indicates
fluorescence reference to which
For Research Use Only. Not for use in diagnostic procedures.

Applera, Applied Biosystems, AB (Design) and VIC are registered trademarks and FAM and ROX are trademarks of Applera Corporation or its subsidiaries in the US and/or certain
other countries. TaqMan is a registered trademark of Roche Molecular Systems, Inc. SYBR is a registered trademark of Molecular Probes, Inc.

© 2008 Applied Biosystems. All rights reserved. Printed in the USA. 05/2008 Publication 136AP01-01

Headquarters International Sales


850 Lincoln Centre Drive | Foster City, CA 94404 USA For our office locations please call the division
Phone 650.638.5800 | Toll Free 800.345.5224 headquarters or refer to our Web site at
www.appliedbiosystems.com www.appliedbiosystems.com/about/offices.cfm

You might also like