Professional Documents
Culture Documents
Abstract
Keywords: Spatially resolved measurement; Electric field; Ion flux; Charging conditions; Point-plane
arrangement
Corresponding author. Tel.: +49 721 608 6567; fax: +49 721 608 6563.
E-mail address: joerg.meyer@mvm.uka.de (J. Meyer).
0304-3886/$ - see front matter r 2005 Elsevier B.V. All rights reserved.
doi:10.1016/j.elstat.2005.03.044
ARTICLE IN PRESS
776 J. Meyer et al. / Journal of Electrostatics 63 (2005) 775–780
collection electrode
1.5 m shielded
probe support shielded
sphere current
0.3 m
measurement
HV Iprobe
2.0 m
1.6 m
Icorona
point 0.8 m
electrode
(-60 kV) current probe faraday cage HV
on xy-stack (grounded)
2.3 m
Fig. 1. Experimental set-up (left: top view, right: front view): point-plate corona arrangement with
measurement device for local electrical data (all housed in Faraday cage).
ARTICLE IN PRESS
J. Meyer et al. / Journal of Electrostatics 63 (2005) 775–780 777
0 0 20 20 20
collection plate /2 /2 0/ 0/ 0/ y
30 90 15 21 27
0/20
0/80 90/80
0/140 150/150
0/200 210/210
0/240
0/270 270/270
probe x
position:
x/y corona needle
Fig. 2. Coordinate system and individual measuring points (coordinates in mm from plate centre).
unipolar
L
0
probe current I p
ΦR
probe potential ΦP
Fig. 3. Calculated IP–FP characteristic: spherical probe in hom. E-field and unipolar ion atmosphere.
~R j and thus to r:
and local field strength jE
dI P j~
jj
¼ 4prP ¼ 4prP rm. (2)
dFP ~R j
jE
~R j can generally be calculated from the dimension of the
The local field strength jE
quadratic section, which is difficult and susceptible to errors, if measured data scatter
significantly. Therefore, a least-squares fit was used to determine j~ ~R j; and FR
jj; jE
from measured IP–FP characteristics.
2. Experimental results
The influence of the measurement probe on the corona discharge was examined by
determining the corona current while varying FP around FR for different probe
ARTICLE IN PRESS
778 J. Meyer et al. / Journal of Electrostatics 63 (2005) 775–780
positions. The corona current IC normalised with the corona current at FP ¼ FR vs.
the deviation of the probe potential from the local space potential FP–FR is shown
for the line from point to plate (Fig. 4a), whilst Fig. 4b shows the data for a line
parallel to the plate and for the diagonal line (xEy), respectively. Only in the vicinity
(x ¼ 270 mm) of the corona needle is IC significantly influenced by the probe
potential (over 75%).
In the vicinity (x ¼ 20 mm) of the collection plate, the measured probe
characteristics match the predicted shape almost perfectly, as illustrated by the
1.05 1.02
y = 0 mm x = 20 mm
rel. corona current IC / IC,Φ [-]
R
1.02
x≈y
x = 20 mm
x = 80 mm
x = 140 mm 1
x = 200 mm
x = 240 mm x = 20 mm x = 150 mm x = 270 mm
x = 270 mm y = 0 mm y = 150 mm y = 270 mm
0.95 0.98
-2 -1 0 1 2 3 -2 -1 0 1 2 3
(a) Probe potential - local space potential ΦP -ΦR [kV] (b) Probe potential - local space potential ΦP -ΦR [kV]
Fig. 4. Influence of probe on discharge: normalised corona current vs. FP–FR (a) y ¼ 0 mm; (b) x ¼
20 mm and xEy, respectively.
10 2
probe current Ip [nA]
200 5
probe current I p [nA]
Fig. 5. Measured IP–FP characteristic and corresponding least square fit for different probe positions.
ARTICLE IN PRESS
J. Meyer et al. / Journal of Electrostatics 63 (2005) 775–780 779
250 5
x = 20 mm
200 4
j / E [10-9C/(s·V·m)]
j [A/m2]
E [kV/m]
150 3
-Φ [kV]
100 2
50 1
0 0
0 100 200 300
(a) y / mm
350 7000
y = 0 mm
300 6000
j / E [10-9C/(s·V·m)]
250 5000
j [A/m2]
200 4000
E [kV/m]
-Φ [kV]
150 3000
100 2000
5 1000
0 0
0 100 200 300
(b) x / mm
300 0.9
x≈y
j / E [10-9C/(s·V·m)]
200 0.6
E [kV/m]-
j [A/m2]
-Φ [kV]
100 0.3
0 0.0
0 100 200 300
(c) x / mm
Fig. 6. Measured local electrical data FP (triangles, m), E (squares, &), j (diamonds, }) and j=Er
(bullets, ) for (a) x ¼ 20 mm (parallel to collection plate), (b) y ¼ 0 mm (from corona needle to plate), (c)
xEy (diagonal line towards stray field).
corresponding IP–FP plots for two individual measurements (Fig. 5a and b).
However, near the corona needle (x ¼ 270 mm; y ¼ 0 mm; Fig. 5c) and at the
outer regions of the stray field of the set-up towards the surrounding Faraday
cage (x ¼ 270 mm; y ¼ 270 mm; Fig. 5d), the deviation between measured data and
ARTICLE IN PRESS
780 J. Meyer et al. / Journal of Electrostatics 63 (2005) 775–780
least-squares fit increases. While the determination of FR and the slope of the linear
section N (and from that j~ jj=jE~R jr) is still possible with acceptable precision, the
dimension of the quadratic section becomes harder to detect leading to an increasing
error for jE ~R j and thus for j~
jj:
Average values for FR, j~ ~R j and j~
jj; jE ~R j (Fig. 6) were determined from at least
jj=jE
four individual measurements, with error bars indicating the standard deviation s
for all measurements at one location. As expected, s is small for the area near the
collection plate (Fig. 6a), and increases both towards the corona tip (Fig. 6b) and the
stray field (Fig. 6c). Also, errors for j~jj and jE~R j are generally larger than for FR and
~ ~
jjj=jE R j: The E-field is rather flat on the axis from corona needle to the collection
plane with only a slight increase near the corona tip (as can be expected qualitatively
from the effect of the ion space charge), while ion flux and ion concentration increase
drastically towards the needle electrode. In spite of a significant drop along the
diagonal line xEy (Fig. 6c), field strength and ion concentration stay rather high in
the stray field, being only halved compared to the centre position near the collection
plate (x ¼ 20 mm; y ¼ 0 mm).
As the gradient of the electric field is (due to the space charge effects) rather low,
even at positions near the corona needle (x ¼ 2002270 mm; y ¼ 0 mm), the
assumption of a quasi-homogenous E-field—as a prerequisite for the applicability
of the measurement technique—holds true. Indeed, in the vicinity of the corona
needle, at x ¼ 240 mm; y ¼ 0 mm; the E-field changes only 1.2% across the probe
diameter, when approximating the local gradient of the electric field strength
between position x ¼ 200 and 270 mm (at y ¼ 0 mm) by a linear correlation.
3. Conclusions
The determination of local electrical data was successfully completed for the
largest region of a point-plate corona arrangement. However there are some
concerns on the accuracy of the data in the very close vicinity of the corona
electrode. The quantitative data can now be used to validate simulations of FR and r
for the same set-up [4]. These simulations in turn can be used to calculate
geometrically more complex charging devices, which are not measurable by the
technique shown in this paper without extensive further work.
References