GALILEO

Ventilation without limitations


The full-functioned GALILEO can ventilate all patients from 2 to 200 kg. It offers a wide choice of ventilation modes, including five conventional modes and two unique pressure-based modes. Up to 26 real-time monitoring parameters can be displayed as waveforms, loops, or trends, for use in pulmonary function analysis and evaluation of therapy. With GALILEOs intuitive user interface, supported by a large color screen and two press-and-turn knobs, GALILEOs setup is quick and easy. The GALILEOs alarm system maximizes patients safety. A universal power supply, integrated battery, communications interface, and trolleys of different heights make the GALILEO a flexible and powerful addition to the ICU.

GALILEO ordering information


To order your GALILEO, you must select a ventilator unit, a country kit, a mounting option, expiratory valve covers and membranes, Flow Sensors, and an oxygen cell. Complete your order by selecting from the parts at the bottom of the page as desired.

Select from each category:


1

Ventilator unit
See p. 6 for a description of options. 155053 GALILEO ventilator, with nebulization and communications interface options 155052 GALILEO ventilator, with nebulization option 155051 GALILEO ventilator, with communications interface option 155050 GALILEO ventilator, without options
1 2 9

Country kit
Includes medication tray, DISS gas connectors, hose clamps, power cord with Schuko (continental Europe) plug, language-specific documentation, drilling template for shelf mounting, and screw to secure GALILEO to trolley or shelf. To replace DISS fittings with NIST fittings, also order a NIST kit. 155411 Country kit for GALILEO, English 155412 Country kit for GALILEO, French 155410 Country kit for GALILEO, German 155414 Country kit for GALILEO, Italian 155413 Country kit for GALILEO, Spanish 155330 NIST fittings kit, GALILEO/RAPHAEL
12 11 10 5 6 4

Mounting options
Trolleys include an accessory basket. 155450 Standard trolley with battery backup, GALILEO 155451 Short trolley with battery backup, GALILEO 155305 Shelf mount without battery backup, GALILEO

Expiratory valve cover and membranes


Available individually or included in some HAMILTON MEDICAL patient breathing circuits. 151228 Cover, expiratory valve, reusable 151233 Membrane, expiratory valve, reusable, package of 10

Flow Sensor
A Flow Sensor is required. See p. 33 for details.

Oxygen cell
Either the Catalyst or Teledyne oxygen cell will provide satisfactory results. 396008 Oxygen cell, Catalyst, for GALILEO, RAPHAEL, and ARABELLA 396009 Oxygen cell, Teledyne, for GALILEO, RAPHAEL, and ARABELLA

Complete your order with the following:


7

Compressor (not shown)


You can mount a compressor on the standard or short trolley with no special installation hardware required. See p. 25 for details.

Cylinder holder (not shown)


With this holder, your trolley can accommodate two gas cylinders. 155394 Cylinder holder, GALILEO

10

Patient breathing set


A patient breathing set is required. See p. 27 for details.

11

Humidifier
See p. 37 for details.

Support arm and tubing holders


A support arm is required. See p. 36 for details.

12

Test bag
A test bag is required for preoperational testing. See p. 34 for details.

Field upgrades
HAMILTON MEDICAL offers upgrades allowing you to add options to your ventilator after purchase. Field upgrades must be installed by a qualified technician.

Software upgrade 1
Enhances your GALILEO with neonatal ventilation capability; monitoring improvements, such as 10-hour trending, waveform freeze and cursor measurement, an ASV target graphics window, and programmable main monitoring parameters; plus work of breathing and other software improvements. Includes language-specific documentation. 155431 Software upgrade 1 kit, GALILEO, English 155432 Software upgrade 1 kit, GALILEO, French 155430 Software upgrade 1 kit, GALILEO, German 155434 Software upgrade 1 kit, GALILEO, Italian 155433 Software upgrade 1 kit, GALILEO, Spanish

Nebulizer
Adds the capability to power a nebulizer with 6 l/min flow at 100 kPa for 30 minute periods. Requires software upgrade 1. Nebulizer jar and accessories are not included. 155452 Nebulizer upgrade kit, GALILEO

Battery backup
Adds battery backup to your GALILEO with trolley to permit the ventilator to operate without ac power for approximately 1 hour for intrahospital transport purposes. For other languages, contact your HAMILTON MEDICAL representative. 155379 Battery backup upgrade kit, GALILEO with standard trolley, English 155386 Battery backup upgrade kit, GALILEO with standard trolley, German 155378 Battery backup upgrade kit, GALILEO with short trolley, English 155385 Battery backup upgrade kit, GALILEO with short trolley, German

Communications interface
Adds communications capabilities to your GALILEO, allowing its interface with a patient monitor or computer, plus a remote alarm (nurses call) device, external nebulizer, or other device. Through the RS-232 connector (9-pin female), the GALILEO can output monitored data, waveforms, modes, control settings, and alarms to a patient monitor or a computer. Patient monitors compatible with the GALILEO include Hewlett-Packard, SpaceLabs, and Marquette. Through the Special connector (15-pin female), the GALILEO can output signals to a remote alarm (nurses call), or it can output I:E timing signals to a device such as an external nebulizer. Cables and adapters are not included. See p. 43 for ordering information. 155453 Communications interface kit, GALILEO

In addition to the communications interface kit, order the parts below to complete your implementation. Application Hewlett-Packard monitor SpaceLabs monitor Marquette monitor Computer Nurses call device, remote alarm, NO application device, or external nebulizer Parts required Vuelink and cable for HAMIILTON MEDICAL ventilator, available from monitor manufacturer, adapter Flexport and cable for HAMILTON MEDICAL ventilator, available from monitor manufacturer, adapter Octanet and cable for HAMILTON MEDICAL ventilator, available from monitor manufacturer, adapter. Note: Tram-net does not work. RS-232 communications cable (shielded and grounded on one side only), DataLogger software* Cable with 15-pin male connector plus a connector to interface to external device

*Can be downloaded from the HAMILTON MEDICAL Web site (www.hamilton-medical.com). This allows the importing of files into Excel for data management and clinical studies.

Operator replacement parts


155202 Medication tray, GALILEO

279166 Filter, fan, for GALILEO, AMADEUSFT, and VEOLARFT

Short trolley configuration shown

155198 Accessory basket, GALILEO

279676 Microfilter, gas supply, 5 m (GALILEO/RAPHAEL/ARABELLA) 279729 Bowl, gas supply filter assembly, GALILEO/RAPHAEL/ARABELLA 363071 Fuse, T 4.0 A H 250 V (GALILEO/ AMADEUSFT)

Maintenance tips
For complete information, consult the operators manual. Check the fan filter every month. Clean or replace if necessary. Check the gas supply filter assemblies every three months. Clean or replace parts as necessary. Replace the oxygen cell every year or 5000 operating hours, whichever comes first.

RAPHAEL

The ABCs of ventilation: Adaptive, Biphasic, Compact


RAPHAEL is HAMILTON MEDICALs latest development: an intensive care ventilator for pediatric and adult patients, in a compact package. Offering todays sophisticated ventilation modalities, the RAPHAEL is adaptive to the individual condition and needs of each patient. RAPHAELs adaptive controller delivers the target tidal volume with the lowest pressure possible, combining the benefits of pressure-controlled ventilation with a volume guarantee. RAPHAELs biphasic ventilation concept allows the patient to breathe freely in all ventilation modes and respiratory phases. RAPHAEL comes complete with battery backup, nebulizer, and oxygen monitoring. The small package fits into a tight corner of the ICU -- and into a small budget too.

RAPHAEL ordering information


To order your RAPHAEL you must select a ventilator unit, a country kit, expiratory valve covers and membranes, Flow Sensors, and an oxygen cell. Complete your order by selecting from the parts at the bottom of the page as desired.

Select from each category:


1

Ventilator unit
157100 RAPHAEL ventilator basic unit

Country kit
Includes DISS gas connectors, hose clamps, power cord with Schuko (continental Europe) plug, language-specific documentation, drilling template for shelf mounting, and screw to secure RAPHAEL to trolley or shelf. To replace DISS fittings with NIST fittings, also order a NIST kit. 157300 Country kit for RAPHAEL, English 157302 Country kit for RAPHAEL, French 157301 Country kit for RAPHAEL, German 157304 Country kit for RAPHAEL, Italian 157303 Country kit for RAPHAEL, Spanish 155330 NIST fittings kit, GALILEO/RAPHAEL
11 5 3

1 2 4

Expiratory valve cover and membranes


Available individually or included in some HAMILTON MEDICAL patient breathing sets. 151228 Cover, expiratory valve, reusable 151233 Membrane, expiratory valve, reusable, package of 10
7 10 9

Flow Sensor
155362 Flow Sensor, pediatric/adult, reusable, box of 10 279331 Flow Sensor, pediatric/adult, single-patient use, box of 10

Oxygen cell
Either the Catalyst or Teledyne oxygen cell will provide satisfactory results. 396008 Oxygen cell, Catalyst, for GALILEO, RAPHAEL, and ARABELLA 396009 Oxygen cell, Teledyne, for GALILEO, RAPHAEL, and ARABELLA

Complete your order with the following:


6

Trolley
Order a trolley, and add a basket and dual gas cylinder holder if desired. 157240 Trolley, RAPHAEL 157242 Basket, for RAPHAEL trolley 157241 Cylinder holder, RAPHAEL

Compressor
If you intend to install the compressor to the trolley, order an adapter. See p. 25 for compressor ordering information. 157243 Adapter, VENTILAIRII to RAPHAEL trolley

Patient breathing set


A patient breathing set is required. See p. 27 for details.

10

Humidifier
See p. 37 for details.

11

Test bag
A test bag is required for preoperational testing. See p. 34 for details.

Support arm and tubing holders


A support arm is required. See p. 36 for details.

Operator replacement parts

281264 Filter, fan, for RAPHAEL

363077 Fuse, T 1.0 A H 250 V

279676 Microfilter, gas supply, 5 m (GALILEO/RAPHAEL/ARABELLA) 279729 Bowl, gas supply filter assembly, GALILEO/RAPHAEL/ARABELLA

Maintenance tips
For complete information, consult the operators manual. Check the fan filter every month. Clean or replace if necessary. Check the gas supply filter assemblies every three months. Clean or replace parts as necessary. Replace the oxygen cell every year or 5000 operating hours, whichever comes first.

VENTILAIRII

A reliable and powerful source of medical compressed air


With its superior pressure and flow characteristics, the VENTILAIRII assures constant, clean, dry compressed air for medical applications, supporting all the performance demands of advanced ventilators. It can be used either as a primary air source or as an automatic backup air system to your central air pipeline system. You can use the VENTILAIRII as a standalone compressor or integrated with the GALILEO, RAPHAEL, VEOLARFT, and AMADEUSFT ventilators. The VENTILAIRII cart lets you use the compressor as a mobile air compressor system.

VENTILAIRII compressor unit


Can be used with all HAMILTON MEDICAL ventilators. Includes ventilator connecting hoses with DISS fittings and a power cord with Schuko (continental Europe) plug. The compressor unit is available only in the electrical configurations indicated. On the GALILEO: No additional parts required to mount on GALILEO with trolley. On the RAPHAEL: Requires an adapter (see below). On the AMADEUSFT and VEOLARII: Can be mounted to trolley versions using a mounting kit (see below). 155601 VENTILAIRII compressor, 115 V 10%, 60 Hz 155600 VENTILAIRII compressor, 230 V 10%, 50 Hz

GALILEO with VENTILAIRII

RAPHAEL with VENTILAIR II

VENTILAIRII-to-ventilator mounting kits


Lets your adapt the VENTILAIRII to your specific ventilator. No extra parts are required to mount the compressor to the GALILEO. 157243 Adapter, VENTILAIRII to RAPHAEL trolley 155660 Mounting kit, VENTILAIRII to AMADEUSFT or VEOLARFT trolley
AMADEUSFT with VENTILAIRII

VENTILAIRII cart
Lets you use the VENTILAIRII as a mobile air compressor system for the GALILEO shelf mount version and other applications. 155670 Cart, VENTILAIRII

VENTILAIRII cart

User maintenance parts


155637 Filter, inlet, VENTILAIRII 363080 Fuse, T 10 A H 250 V (for 115 V units) 363073 Fuse, T 3.15 A H 250 V (for 230 V units)

Maintenance tips
Consult the operators manual for complete information. Check the inlet filter weekly. Clean or replace as necessary.