DEPARTMENT OF TECHNICAL EDUCATION

ANDHRA PRADESH
Name : P. Srinivas
Designation : Lecturer
Branch : Electronics & Communication Engg.
Institute : GPW, Srikakulam, Srikakulam Dist.
Year/Semester : VI Semester
Subject : Analytical Instrumentation
Subject Code : AEI-603
Topic : Pollution Analysers
Duration : 50 Mts
Sub Topic : Monitoring of Carbon Monoxide
Teaching Aids : Diagrams
AEI 603.48 1
Objectives

On completion of this period, you would be


able to know :

• Monitoring of Carbon Monoxide

AEI 603.48 2
Methods of monitoring of CarbonMonoxide

• Non-dispersive Infrared Gas Analyser

• Gas Chromatography

AEI 603.48 3
Infra red gas Analyser

AEI 603.48 4
Basic principle

• The infrared gas analysers depend for their operation

upon the fact that some gases and vapours absorb

specific wavelength of infrared radiation.

AEI 603.48 5
Description of Infrared Gas Analyser (3/1)
• The analyser contains a pair of matched cells

• One cell is filled with a reference gas like Nitrogen

• The other cell is filled with to be analysed sample gas

• In the arrangement there is one Infrared source


present

• The mirror arrangement make make the light beam


into two paths
AEI 603.48 6
Description of Infrared Gas Analyser (3/2)
• These reflected light beams passing through reference

• And sample chambers through the chopper disc

• The chopper disc is driven by the motor

• The chopper disc makes the continuous beam into a


pulses

AEI 603.48 7
Description of Infrared Gas Analyser (3/3)
• The pulsating light beam is interacted with sample gas

• Presence of CO in the sample gas absorbs the infrared


rays

• The second mirror reflector , reflects the outputs of the


two chambers on to the detector

• The detector develops a signal corresponding to


presence of CO in the sample
AEI 603.48 8
Block Diagram of a Gas Chromatograph

AEI 603.48 9
Description of Gas Chromatography

• An air sample containing CO is passed through a stripper


column

• The heavy hydrocarbons are retained and CO and methane


are passed into a chromatographic column

• And then into a catalytic reducing chamber. The methane


would pass through the reducing chamber unaffected

AEI 603.48 10
Description of Gas Chromatography (2/2)

• While CO is reduced to methane

• By using hydrogen flame ionisation detector, both


methane peaks can be detected

• The first peak is due to methane, while the second peak


would correspond to CO

AEI 603.48 11
Summary

In this period, you have learnt about

• Basic principle: Infrared gas analysers depend for


their operation upon the fact that some gases absorb
specific Wavelengths of infrared radiation.

• Chopper: It is used to chopped the continuous


signal It is driven by a motor

AEI 603.48 12
Quiz

1. The detector is used in Gas Chromatography

a) Hydrogen Flame Ionisation detector

b) Photo Flame Detector

c) CO Detector

AEI 603.48 13
Frequently asked Question

• Explain principle of operation of carbon monoxide


monitoring ?

AEI 603.48 14