DEPARTMENT OF TECHNICAL EDUCATION

ANDHRA PRADESH

Name : J.Srinivasa Rao


Designation : Lecturer
Branch : Electronics & Communications
Institution : Govt Polytechnic Bellampally
Year/semester : VI semester
Subject : Analytical Instrumentation
Subject Code : AEI- 603
Topic : Computer Applications in Medicine
Duration : 50mts
Sub Topic : Earth free patient monitoring and
standards

Teaching aids :Diagrams, PPT

AEI-603.60 1
OBJECTIVES
Upon completion of this period, you would be able to
know

• Earth free patient monitoring

• standards

AEI-603.60 2
RECAP

In the previous class we have studied about


• Patient safety
• Microshock
• Macroshock
• Preventive measures to reduce the shock hazards
• Maintaining the safe patient electrical environment

AEI-603.60 3
EARTH FREE PATIENT MONITORING

• The use of single instrument with earthed patient circuits is


not particularly hazardous.
• The use of a number of devices connected simultaneously to
the same patient ,increases the technical problems of
protection.
• Therefore the use of earth-free equipment is becoming
necessary.

AEI-603.60 4
A BASIC EARTH-FREE INSTRUMENT

AEI-603.60 5
A BASIC EARTH-FREE INSTRUMENT

• The Earth-free instrument has a main transformer


with an earthed inter winding screen to provide the
major internal power supplies.

• Most of the circuits and all the exposed networks


are earthed.

AEI-603.60 6
STANDARDS

• On account of the major importance of electrical


safety factors in the design, production and usage of
medical equipment ,standards have been set for this
field.

AEI-603.60 7
STANDARDS

• One such standard is given by the Bureau of Indian


standards under ‘Medical Electrical Equipment part 1

• The standards state that the “Equipment shall be


designed that the risk of electrical shock in normal
use and in single fault condition is obviated as far as
possible”

AEI-603.60 8
A EARTH-FREE INSTRUMENT ASSEMBLY
FREE FROM MOST RISKS

AEI-603.60 9
Single fault conditions

• Interruption of a protective earth conductor

• Interruption of a one supply conductor

• Appearance of an external voltage on an F-type


applied part

AEI-603.60 10
Single fault conditions

• Appearance of an external voltage on a signal input


or on a signal output part

• Failure of an electrical component which might


cause a safety hazard

AEI-603.60 11
STANDARDS

Current type Normal Single fault


condition(mA) condition(mA)
Earth leakage 0.5 1.0
current(general)
Enclosure 0.1 0.5
leakage current
Patient leakage 0.01 0.05
current
AEI-603.60 12
SUMMARY

We have discussed

• Earth free patient monitoring


• Standards

AEI-603.60 13
IMPORTANT QUESTIONS

• Explain about the Earth free patient monitoring


• Explain the standards of earth free patient
monitoring

AEI-603.60 14