Objectives

Upon completion of this topic the student will be


able to know about

• The principle of operation of the capillary tube viscometer

AEI405.62 to 63 2
Capillary Tube Viscometer
Flow in

Liquid vessel

Constant
head (h)
Liquid
Flow
Capillary tube

Overflow
Measuring jar

Fig1
AEI405.62 to 63 3
Capillary Tube Viscometer

• It mainly consists of a Flow in

• Liquid vessel
Liquid vessel

Constant
head (h)
Liquid
Flow

• Capillary tube Capillary tube

Overflow
L
• Measuring jar Measuring jar

• The measuring jar is used to collect the specified volume


of sample liquid
• The constant pressure head or hydrostatic head of fluid
causes the liquid flow through the capillary tube

AEI405.62 to 63 4
• The discharge rate (Q) can be easily calculated by using
measuring jar and stop watch

Volume of liquid collected in the measuring jar


Q=
Time taken to collect the liquid in the measuring jar

AEI405.62 to 63 5
If the flow is laminar, the discharge rate (Q) is given by

πD4ΔP
Q=
128μL
Where
Q = Volume of the liquid passing through the tube
per second
L = Length of the capillary tube
D = Diameter of the capillary tube
μ = Coefficient of viscosity
ΔP = pressure drop across the ends of the tube

AEI405.62 to 63 6
ΔP = ρ g h

Where ρ = Density of the fluid

g = Acceleration due to gravity

h = constant head

πD4ρg h
Q=
128μL

πD4 ρg h
μ=
128QL

AEI405.62 to 63 7
• In the above equation the diameter of Capillary tube is
raised to fourth power

• Hence it is essential to measure it as accurately as


possible

• By using traveling microscope the diameter of the


capillary tube can be measured accurately

• Capillary tube viscometer can be used as a flow


metering device if the viscosity of the liquid is known

AEI405.62 to 63 8
Advantages

• Simple in construction

• No maintenance required

• Easy to use

AEI405.62 to 63 9
Disadvantage

• The main disadvantage of this viscometer is that it is not

suitable for unclean fluids as the dirt or grit tends to clog

the capillary tube

AEI405.62 to 63 10
Applications

• They can be used as secondary standards for the

calibration of other type of viscometers

• They are used in the refineries to measure the viscosity

of petroleum products

AEI405.62 to 63 11
Fig 2
AEI405.62 to 63 12
Summary

We have discussed about :


• Principal of operation of the capillary tube viscometer

• Advantages & Disadvantages

• Applications

AEI405.62 to 63 13
Quiz

1. In capillary tube viscometer, the coefficient of

viscosity μ =

a) (πD4 ρ gh) / (128QL)

b) (πD4 ρ gh) / (128µL)

c) ρ gh

d) None of the above

AEI405.62 to 63 14
1. The capillary tube viscometer is not suitable for

a) Cleaned fluids

b) Dirty fluids

c) Both a and b

d) None

AEI405.62 to 63 15
Frequently asked questions

1. Explain the principle of operation of the capillary tube


viscometer

AEI405.62 to 63 16