DEPARTMENT OF TECHNICAL EDUCATION

ANDHRA PRADESH
Name of the staff : D. Simhachalam
Designation : Lecturer
Branch : A.E.I.E
Institute : K. D. R .G. P. T, Wanaparthy
Semester : III
Subject : Process Instrumentation-I
Subject code : AEI-305
Topic : Conductivity, Weight & Humidity
measurement
Duration : 100 minutes

Subtopic : Description of Conductivity meters


Teaching Aids : PPT’s, Animations, Diagrams
AEI 305.69 to 70 1
Recap

In the previous period, you have studied about

• The description of conductivity cell

• In theory, a conductivity cell is formed by two 1-cm


square surfaces spaced 1-cm apart

• Now this cell is used for the measurement of


conductance

AEI 305.69 to 70 2
Objectives

Upon completion of this topic, you would be able


to know about Conductivity meters

• Principles

• Operation

AEI 305.69 to 70 3
Principle of operation

• The solution whose conductance is to be measured, is


taken in the conductivity cell

• The apparatus used for conductance measurement is


shown in the figure

• It is a modified form of Wheatstone bridge

AEI 305.69 to 70 4
Conductivity meter

Fig 1
AEI 305.69 to 70 5
Principle of operation…

• The conductance varies with temperature.

• Therefore, the solution temperature should be kept


constant throughout the experiment.

• This is achieved by placing the conductivity cell in the


thermostat.

AEI 305.69 to 70 6
Principle of operation…

• The cell is connected to a resistance box R on one side

• A long thin wire AB stretched along a scale on the other


side

• Now some known resistance ‘R’ is taken out of the


resistance box

AEI 305.69 to 70 7
Principle of operation…

• An alternating current is passed through the solution with


the help of induction coil

• The sliding contact ‘d’ is moved on the wire AB

• So that minimum or no sound is heard in the head


phone, T.

AEI 305.69 to 70 8
Principle of operation…
• At this point, the following relationship holds good

Resistance of solution length Bd


=
Resistance R length Ad

• Thus by knowing R, Ad and Bd, the resistance of the


solution can be calculated.

AEI 305.69 to 70 9
Principle of operation…

• Since conductivity of the solution = 1/Resistance of the


solution ,

• Therefore, by knowing resistance of the solution, the


conductivity of the solution can be calculated

AEI 305.69 to 70 10
Conductivity Cells

• There are two types of Conductivity Cell

• Two electrode cell

• Four electrode cell

AEI 305.69 to 70 11
Two electrode cell

• The two electrode cell is the most commonly used


conductivity cell

• Glass cells have electrodes made of platinum.

• Epoxy cells have easy to clean graphite plates.

AEI 305.69 to 70 12
Two electrode cell…

• These conductivity cells do not have errors caused by


fringe field effects.

• Because the entire measuring field is contained within


the electrode body.

• This kind of probe can be used with almost any


conductivity meter.

AEI 305.69 to 70 13
Conductivity Measurement

Na+

Cl-

Fig 2

AEI 305.69 to 70 14
Two electrode cell

Fig 3
AEI 305.69 to 70 15
Two electrode cell (Epoxy)

Fig 4

AEI 305.69 to 70 16
Two electrode cell (Glass)

Fig 5

AEI 305.69 to 70 17
Four electrode cell

• The electrode design reduces the problem of polarization


error and fouling of the electrode.

• Polarization comes about because of electro-chemical


reaction with the probe surface

• In the four electrode design alternating current is applied


only to the outer pair of rings.

AEI 305.69 to 70 18
Four electrode cell…

• The voltage is measured on the inner rings without

polarization effects

• Because no current flows in the measuring circuit.

• This type of probe is superior in function in almost any

solution

• Covers the whole conductivity range with one probe

AEI 305.69 to 70 19
Four electrode cell

Fig 5
AEI 305.69 to 70 20
Four Electrode Conductivity Cells

Fig 6

AEI 305.69 to 70 21
Four electrode cell (Epoxy)

Fig 7

AEI 305.69 to 70 22
LaMotte 5 Series

• Some practical conductivity meters such as LaMotte 5


Series are given

• These are CON 5 Meters.

AEI 305.69 to 70 23
The LaMotte 5 Series

CON 5 Meter
Fig 8

AEI 305.69 to 70 24
The LaMotte 5 Series

Fig 9 CON 5 Meter


AEI 305.69 to 70 25
The LaMotte 5 Series

Fig 10
CON 5 Meter

AEI 305.69 to 70 26
Summary

In this period, you have learnt about :

• The principle of operation of measurement of


Conductivity meter

AEI 305.69 to 70 27
Quiz

1. The conductivity of the solution depends on

a) Temperature
b) Mobility of ions in solution
c) Both (a) & (b)
d) None

AEI 305.69 to 70 28
Quiz

1. How to achieve the conductivity measurement


insensitive to the temperature changes?

b) By keeping Conductivity cell in thermostat

c) Conductivity cell kept in cooler

d) Both (a) & (b)

e) None

AEI 305.69 to 70 29
Frequently asked questions

2. Explain the principle of operation of conductivity meter


with neat sketch?

4. Draw the diagram of conductivity meter

AEI 305.69 to 70 30